Biosyn SM-5690Recalls & reported problems
Covidien LP · Monofilament Absorbable Suture,P-13,1x Undyed Strand · Class 2 · Implant
Quick summary
This Covidien absorbable suture is a synthetic monofilament stitch used to close tissue during surgery. The 30 reports mostly describe the needle pulling away from the thread or the connection breaking, sometimes when the package is opened and sometimes mid-procedure; in a few cases a needle was left behind in soft tissue and had to be located. Two reports involve injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 30
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 10884521033153), received between Jul 12, 2017 and Dec 8, 2025 — classified as 2 injuries, 28 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (22 reports).
28 other recalls cover devices cleared under the same FDA submission (K000037); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component22
- Break10
- Component Missing1
Patient problems
- No Clinical Signs, Symptoms or Conditions18
- Insufficient Information6
- No Consequences Or Impact To Patient3
- Foreign Body In Patient1
- No Information1
- No Known Impact Or Consequence To Patient1
- No Patient Involvement1
- Unspecified Tissue Injury1
Recent adverse event reports
The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 8, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, THE NEEDLE WAS PULLED APART FROM THE SUTURE. NO COMPONENT FELL INTO THE CAVITY OF THE PATIENT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 23752650 ↗MalfunctionSep 9, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING SUTURING, THE NEEDLE WAS DETACHED. IT WAS NOTED THAT NO COMPONENT FELL INTO THE CAVITY OF THE PATIENT. A NEW SUTURE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23000054 ↗MalfunctionAug 13, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A RADIOLOGY PROCEDURE, THE NEEDLE AND THREAD CAME OFF WHEN THE USER OPENED THE PACKAGE AND TRIED TO GRASP IT WITH A NEEDLE HOLDER WHILE PREPARING FOR SUTURING BEFORE PATIENT USE. A NEW SUTURE WAS USED WITHOUT ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22796689 ↗MalfunctionFeb 21, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A PROCEDURE, WHILE SUTURING, THE CONNECTION OF THE NEEDLE AND THE THREAD BROKE FOR BOTH DEVICE. A THIRD SUTURE WAS USED TO RESOLVE THE ISSUE.
Patient problems: Insufficient Information
MAUDE report 21432247 ↗MalfunctionFeb 21, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A PROCEDURE, WHILE SUTURING, THE CONNECTION OF THE NEEDLE AND THE THREAD BROKE FOR BOTH DEVICE. A THIRD SUTURE WAS USED TO RESOLVE THE ISSUE.
Patient problems: Insufficient Information
MAUDE report 21432232 ↗InjuryFeb 12, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A PROCEDURE IN THE HEMORRHOIDS, THE NEEDLE DETACHED FROM THE FILAMENT AT THE END OF THE PROCEDURE AND REMAINED IN THE PERIANAL SOFT TISSUES. IT WAS NOTED THAT THE SURGICAL TIME WAS EXTENDED. AN ANASCOPY WAS DONE TO THE PATIENT TO LOCATE THE NEEDLE.
Patient problems: Foreign Body In Patient
MAUDE report 21368272 ↗MalfunctionNov 15, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A BREAST RECONSTRUCTION, WHILE SUTURING, THE NEEDLE AND THREAD DETACHED. A NEW SUTURE WAS OPENED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20695620 ↗MalfunctionNov 13, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A PROCEDURE, WHILE SUTURING, THE THREAD DETACHED. A NEW SUTURE WAS OPENED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20679047 ↗MalfunctionNov 13, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A SCAR REVISION, WHILE SUTURING, THE SUTURE BROKE. AN ADDITIONAL NEW SUTURE WAS USED WITHOUT ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20673983 ↗MalfunctionOct 18, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING CLOSURE OF THE WOUND, WHILE SUTURING, THE NEEDLE DETACHED AT THE CONNECTION WHERE THE THREAD WAS. A NEW SUTURE WAS USED TO RESOLVE THE ISSUE. SURGICAL TIME WAS EXTENDED FOR FIVE MINUTES. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20484878 ↗MalfunctionMay 10, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, PREOPERATIVELY, WHEN THE NEEDLE WAS HELD IMMEDIATELY AFTER OPENING, THE NEEDLE AND THREAD GOT DETACHED. THE SAME PRODUCT FROM THE SAME LOT WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19288785 ↗MalfunctionDec 5, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A SKIN LESION REMOVAL, THE NEEDLE WAS DETACHED. THE SURGEON HAD TO TRASH THE SUTURE AND OPENED A NEW ONE TO RESOLVE THE ISSUE.
Patient problems: Insufficient Information
MAUDE report 18270848 ↗MalfunctionNov 13, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A SKIN LESION REMOVAL, THE NEEDLE WAS DETACHED. THE SURGEON HAD TO TRASH THE SUTURE AND OPENED A NEW ONE AND HAD THE SAME ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18124883 ↗MalfunctionNov 13, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A SKIN LESION REMOVAL, THE NEEDLE WAS DETACHED. THE SURGEON HAD TO TRASH THE SUTURE, OPENED A NEW ONE BUT HAD THE SAME ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18123057 ↗MalfunctionApr 13, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING AN OPEN PROCEDURE, THE SUTURE WAS PLACED THROUGH THE TUNICA ALBUGINEA. IT WAS ENSURED THAT THE TESTICLE WAS NOT TORSED AND TESTICLE WAS THEN BROUGHT THROUGH THE TUNNEL AND OUT THE SCROTAL INCISION. THE SUTURE WAS LOADED OVER A KEITH NEEDLE WHICH WAS THEN PASSED THROUGH THE MOST DEPENDENT PORTION OF THE SCROTUM. DURING THIS PROCESS, THE SUTURE NEEDLE POPPED OFF FROM THE SUTURE THREAD. NO DEVICE COMPONENT FELL INTO THE PATIENT'S CAVITY. ANOTHER SAME SUTURE FROM THE SAME LOT WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16732171 ↗MalfunctionMar 22, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE SUTURE POPPED OFF THE NEEDLE WHILE THE SURGEON WAS SUTURING THE PATIENT'S RIGHT ARM. IT WAS OPENED OR REMOVED FROM THE PACKAGING A MINUTE OR TWO BEFORE THE SURGEON USED IT.
Patient problems: Insufficient Information
MAUDE report 16592377 ↗MalfunctionFeb 20, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A ROBOTIC LAPAROSCOPIC UMBILICAL HERNIA REPAIR WITH MESH POSSIBLE AND BILATERAL ROBOTIC INGUINAL HERNIA REPAIR, WHILE THE SURGEON WAS SUTURING (THE SKIN WAS CLOSED WITH RUNNING 4.0 MONOCRYL SUTURE AND STERI-STRIPS), THE SUTURE THREAD COMPLETELY SEPARATED FROM THE SUTURE NEEDLE. THE SURGEON USED THE SAME SUTURE TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16404341 ↗MalfunctionSep 22, 2022Break; Break
ACCORDING TO THE REPORTER, DURING AN OPEN EXCHANGE OF BILATERAL IMPLANTS, CAPSULECTOMY OF LEFT BREAST AND EXCISION OF FAT NECROSIS OF LEFT BREAST PROCEDURE, WHEN CLOSING OF INCISION AT THE SKIN, THE THREAD BROKE. THE SURGEON UTILIZED A SECOND SUTURE THAT BROKE AND THEN USED ANOTHER SUTURE TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15462554 ↗MalfunctionSep 21, 2022Break; Break
ACCORDING TO THE REPORTER, DURING AN OPEN EXCHANGE OF BILATERAL IMPLANTS, CAPSULECTOMY OF LEFT BREAST AND EXCISION OF FAT NECROSIS OF LEFT BREAST PROCEDURE, WHEN CLOSING OF INCISION AT THE SKIN, THE THREAD BROKE. THE SURGEON UTILIZED A SECOND SUTURE THAT BROKE AND THEN USED ANOTHER SUTURE TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15461933 ↗MalfunctionJun 1, 2022Break; Break
ACCORDING TO THE REPORTER, THREAD BROKE. ANOTHER SUTURE WAS USED TO RESOLVE THE ISSUE IN ORDER TO COMPLETE THE CASE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14569302 ↗InjuryMar 30, 2022Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A PROCEDURE, WHILE CLOSING THE SKIN WITH SUTURE, THE NEEDLE BROKE AND FELL INTO THE PATIENT'S CAVITY. THE SURGEON OPENED THE INCISION AND DUG THE NEEDLE BACK OUT. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions; Unspecified Tissue Injury
MAUDE report 13950881 ↗MalfunctionMar 15, 2022Component Missing; Detachment of Device or Device Component; Component Missing; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A PROCEDURE, WHEN THE SCRUB TECHNOLOGIST GRABBED AND PULLED THE SUTURE OUT OF THE PACKAGE, IT WAS NOTED THAT THERE WAS NO THREAD ATTACHED TO THE NEEDLE; AND ON THE SECOND SUTURE, THE THREAD EASILY FELL OFF THE NEEDLE WHEN PULLED OUT OF THE PACKAGE. THE SURGEON USED A COMPETITOR'S SUTURE TO RESOLVE THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13773029 ↗MalfunctionNov 17, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, PRE OPERATIVELY, AT THE TIME OF OPENING, THE 2 NEEDLES AND THREAD WERE FOUND TO BE DISENGAGED ALREADY. A NEW SUTURE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12828047 ↗MalfunctionJun 24, 2021Break; Break
ACCORDING TO THE REPORTER, ON AN OPEN REDUCTION INTERNAL FIXATION, WHILE CLOSING THE SKIN, THE THREAD BROKE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE.
Patient problems: Insufficient Information
MAUDE report 12059612 ↗MalfunctionJun 25, 2020Break; Break
ACCORDING TO THE REPORTER, WHILE STITCHING TO CLOSE AN OPEN PROCEDURE, THE SUTURE LINE BROKE. IT WAS STATED THAT MULTIPLE PACKAGES WERE USED AS THE ISSUE OCCURRED WITH MULTIPLE PACKAGES. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10196648 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K000037. One submission often covers many models, so check the product and lot details in each record.
- Z-1292-2020 · Dec 18, 2019 · TerminatedPolysorb Braided Absorbable Suture 6-0 18" UNDYED SS-24 D/A, Item Code L2752K - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, bu
- Z-1283-2020 · Dec 18, 2019 · TerminatedPolysorb Braided Absorbable Suture 2-0 VIOLET 36" GS-21, Item Code CL923 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not
- Z-1286-2020 · Dec 18, 2019 · TerminatedPolysorb Braided Absorbable Suture 2-0 VIOLET 30" V-20, Item Code GL123 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not
- Z-1296-2020 · Dec 18, 2019 · TerminatedPolysorb Braided Absorbable Suture 4/0 30 VIO SC-2 124K, Item Code SL654 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not
- Z-1275-2020 · Dec 18, 2019 · TerminatedPolysorb Braided Absorbable Suture 0 30" UNDYED HOS-11, Item Code CL537 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not
- Z-1297-2020 · Dec 18, 2019 · TerminatedPolysorb Braided Absorbable Suture 4/0 30 VIOLET CVF23, Item Code UL203 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not
- Z-1293-2020 · Dec 18, 2019 · TerminatedPolysorb Braided Absorbable Suture 6/0 18" UNDYED P-10, Item Code SL1625G - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but no
- Z-1284-2020 · Dec 18, 2019 · TerminatedPolysorb Braided Absorbable Suture 1 36 VIOLET GS-21, Item Code CL925 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not in
- Z-1277-2020 · Dec 18, 2019 · TerminatedPolysorb Braided Absorbable Suture 3/0 30 VIOLET GS-11, Item Code CL815 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not
- Z-1285-2020 · Dec 18, 2019 · TerminatedPolysorb Braided Absorbable Suture 2-0 UNDYED 36" GS-21, Item Code CL953 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not
- Z-1276-2020 · Dec 18, 2019 · TerminatedPolysorb Braided Absorbable Suture 2/0 30 VIOLET GS-21, Item Code CL811 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not
- Z-1287-2020 · Dec 18, 2019 · TerminatedPolysorb Braided Absorbable Suture 0 VIOLET 30" V-20, Item Code GL124 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not in
- Z-1279-2020 · Dec 18, 2019 · TerminatedPolysorb Braided Absorbable Suture 13-0 VIOLET 30" GS-23, Item Code CL892 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but no
- Z-1282-2020 · Dec 18, 2019 · TerminatedPolysorb Braided Absorbable Suture 1 36 VIOLET GS-24, Item Code CL915 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not in
- Z-1281-2020 · Dec 18, 2019 · TerminatedPolysorb Braided Absorbable Suture 0 36 VIOLET GS-24, Item Code CL914 - Product Usage: Indicated for use in soft tissue approximation or ligation and ophthalmic surgery, but not in
…and 13 more.
Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices
- Coated VICRYL VCP433HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death1,093 reports
- Polysorb CL-915Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID503 reports
- Coated VICRYL VCP493HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID435 reports
- V-Loc 180 VLOCL0316Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall376 reports
- Coated VICRYL VCP345HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death405 reports
- V-Loc 180 VLOCL0604Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall328 reports
- V-Loc 180 VLOCA008LCovidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall294 reports
- V-Loc 180 VLOCL0326Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID330 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.