Caprosyn SC-691Recalls & reported problems
Covidien LP · Monofilament Absorbable Suture · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10884521034990), received between Jan 6, 2023 and Dec 25, 2025 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component10
- Break1
- Patient-Device Incompatibility1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
- Insufficient Information4
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 25, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A CAT OVARIECTOMY, AT THE TIME OF SUTURING, THE SUTURE STRAND WAS TAKEN BUT THE NEEDLE DETACHED IMMEDIATELY FROM THE RETAINER. THE NEEDLE DID NOT FALL INTO THE PATIENT CAVITY. ANOTHER STRAND WAS TAKEN TO FINISH THE PROCEDURE WITHOUT ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23892160 ↗MalfunctionDec 24, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING THE VETERINARY CAT OVARIECTOMY, AT THE TIME OF SUTURING, THE SUTURE STRAND WAS TAKEN AND THE NEEDLE DETACHED RIGHT FROM THE RETAINER. THE NEEDLE DID NOT FALL INTO THE PATIENT'S CAVITY. ANOTHER SUTURE WAS USED TO COMPLETE THE CLOSURE AND RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23891476 ↗MalfunctionSep 2, 2025Patient-Device Incompatibility; Patient-Device Incompatibility
IT WAS REPORTED THAT THE SUTURE HAD PREMATURE ABSORPTION. THE SUTURE HAD CRUMBLED WHEN IT CAME OUT OF THE PACKAGE. NO INTERVENTION WAS REQUIRED. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22948499 ↗MalfunctionSep 5, 2024Break; Break
ACCORDING TO THE REPORTER, THE SUTURE BROKE WHEN THROWING A KNOT. A NEW SUTURE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20151111 ↗MalfunctionJul 19, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, PRIOR TO USE OF THE DEVICE, BEFORE THE DEVICE WAS REMOVED FROM THE PACKAGING, WHILE THE NEEDLE WAS STILL IN THE RETAINER, THE NEEDLE WAS DETACHED FROM THE THREAD. ANOTHER DEVICE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19785989 ↗MalfunctionFeb 28, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING AN OPEN HERNIOPLASTY PROCEDURE, AT THE TIME OF SKIN CLOSURE, THE USER OPENED THE SUTURE AND PLACED IT IN THE NEEDLE HOLDER, AND WHEN GRASPED, IT WAS EASILY DETACHED FROM THE STRAND. ANOTHER SUTURE WAS OPENED, AND THE SAME ISSUE HAPPENED. ANOTHER PRODUCT FROM A COMPETITOR'S WAS OPENED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18798261 ↗MalfunctionFeb 28, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING AN OPEN HERNIOPLASTY PROCEDURE, AT THE TIME OF SKIN CLOSURE, THE USER OPENED THE SUTURE AND PLACED IT IN THE NEEDLE HOLDER, AND WHEN GRASPED, IT WAS EASILY DETACHED FROM THE STRAND. ANOTHER SUTURE WAS OPENED, AND THE SAME ISSUE HAPPENED. ANOTHER PRODUCT FROM A COMPETITOR'S WAS OPENED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18798253 ↗MalfunctionOct 13, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, INTRAOPERATIVELY, WHEN THE USER OPENED THE SUTURE PACKAGE, IT NOTICE THAT THE NEEDLE IS LOOSE FROM THE THREAD.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 17934756 ↗MalfunctionOct 13, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, INTRAOPERATIVELY, CHOLECYSTECTOMY, WHEN THE USER OPENED THE SUTURE PACKAGE, IT NOTICE THAT THE NEEDLE WAS LOOSE FROM THE THREAD. THERE WAS NO PATIENT INJURY.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 17934707 ↗MalfunctionMar 16, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, INTRAOPERATIVELY, THE THREE THREAD OF THE DEVICE BROKE AT THE LEVEL OF THE NEEDLE. PRODUCT WAS RECEIVED WITHOUT ACCOMPANYING INFORMATION.
Patient problems: Insufficient Information
MAUDE report 16553283 ↗MalfunctionJan 10, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, INTRAOPERATIVELY, THE THREE THREADS OF THE DEVICE BROKE AT THE LEVEL OF THE NEEDLE.
Patient problems: Insufficient Information
MAUDE report 16131043 ↗MalfunctionJan 6, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING AN INTRACUTANEOUS CLOSURE, BEFORE USING ON THE PATIENT, FIVE OF THE NEEDLES CAME OFF THE THREAD WHEN THE NURSE TOOK THE THREAD OUT OF THE PACKAGE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16112311 ↗
Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices
- Coated VICRYL VCP433HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death1,093 reports
- Polysorb CL-915Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID503 reports
- Coated VICRYL VCP493HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID435 reports
- V-Loc 180 VLOCL0316Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall376 reports
- Coated VICRYL VCP345HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death405 reports
- V-Loc 180 VLOCL0604Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall328 reports
- V-Loc 180 VLOCA008LCovidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall294 reports
- V-Loc 180 VLOCL0326Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID330 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page