Surgipro II VP-521XRecalls & reported problems

Covidien LP · Monofilament Polypropylene · Class 2 · Implant

All models of Surgipro II (73)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Covidien Surgipro II polypropylene suture is used to close tissue during surgery. The 42 reports mostly describe sutures detaching from needles or breaking during procedures, sometimes requiring replacement or extending surgery; some mention bleeding. There were no deaths and 4 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
41
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 41 adverse event reports that list this exact device (DI 10884521037786), received between Oct 23, 2017 and Sep 18, 2026 — classified as 4 injuries, 37 malfunctions.

The most frequently reported specific device problem is “Break” (24 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 5 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 3 reportsAug 2024: 0 reportsSep 2024: 8 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break24
  • Detachment of Device or Device Component22
  • Adverse Event Without Identified Device or Use Problem2
  • Material Frayed2
  • Difficult to Remove1
  • Unraveled Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions19
  • Insufficient Information10
  • No Consequences Or Impact To Patient4
  • Hemorrhage/Blood Loss/Bleeding3
  • No Known Impact Or Consequence To Patient3
  • Failure to Anastomose2
  • No Code Available2
  • Tissue Damage1

Recent adverse event reports

The 25 most recent of 41 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 18, 2026Break

    It was reported that during a procedure, the suture thread broke. The was no patient injury.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50819181 ↗
  • MalfunctionJul 15, 2025Break; Break

    ACCORDING TO THE REPORTER DURING A PROCEDURE, TWO SUTURES EXPERIENCED ISSUES. IT WAS NOTED THAT ONE SUTURE BROKE IN THE MIDDLE WHILE SUTURING, AND THE OTHER FAILED TO HOLD THE SUTURE LINE DURING THE PROCEDURE. AS A RESULT, THE SURGICAL TIME WAS EXTENDED TO ALLOW FOR RE-ANASTOMOSIS AND THE HEALTHCARE TEAM RESOLVED THE ISSUE BY SWITCHING TO SUTURES FROM ANOTHER MANUFACTURER.

    Patient problems: Failure to Anastomose

    MAUDE report 22509646 ↗
  • MalfunctionJul 15, 2025Unraveled Material; Unraveled Material

    ACCORDING TO THE REPORTER DURING A PROCEDURE, TWO SUTURES EXPERIENCED ISSUES. IT WAS NOTED THAT ONE SUTURE BROKE IN THE MIDDLE WHILE SUTURING, AND THE OTHER FAILED TO HOLD THE SUTURE LINE DURING THE PROCEDURE. AS A RESULT, THE SURGICAL TIME WAS EXTENDED TO ALLOW FOR RE-ANASTOMOSIS AND THE HEALTHCARE TEAM RESOLVED THE ISSUE BY SWITCHING TO SUTURES FROM ANOTHER MANUFACTURER.

    Patient problems: Failure to Anastomose

    MAUDE report 22509595 ↗
  • MalfunctionJan 9, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING THE LUNG REPAIR SURGERY, WHEN THE DOCTOR USED THE NON-ABSORBABLE SINGLE-STRAND POLYPROPYLENE SUTURE TO REPAIR THE LUNG IN THE PATIENT'S CHEST CAVITY, THE SUTURE AT ONE END AUTOMATICALLY DETACHED FROM THE NEEDLE. THE NEEDLE WAS INITIALLY LOST, AND AFTER CAREFUL SEARCHING, THE NEEDLE WAS FOUND IN THE PATIENT'S CHEST CAVITY. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21112211 ↗
  • MalfunctionSep 16, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE EIGHT SUTURES HAD NEEDLE DETACHMENT.

    Patient problems: Insufficient Information

    MAUDE report 20226674 ↗
  • MalfunctionSep 16, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE EIGHT SUTURES HAD NEEDLE DETACHMENT.

    Patient problems: Insufficient Information

    MAUDE report 20226625 ↗
  • MalfunctionSep 16, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE EIGHT SUTURES HAD NEEDLE DETACHMENT.

    Patient problems: Insufficient Information

    MAUDE report 20226606 ↗
  • MalfunctionSep 16, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE EIGHT SUTURES HAD NEEDLE DETACHMENT.

    Patient problems: Insufficient Information

    MAUDE report 20226534 ↗
  • MalfunctionSep 16, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE EIGHT SUTURES HAD NEEDLE DETACHMENT.

    Patient problems: Insufficient Information

    MAUDE report 20226525 ↗
  • MalfunctionSep 16, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE EIGHT SUTURES HAD NEEDLE DETACHMENT.

    Patient problems: Insufficient Information

    MAUDE report 20226508 ↗
  • MalfunctionSep 16, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE EIGHT SUTURES HAD NEEDLE DETACHMENT.

    Patient problems: Insufficient Information

    MAUDE report 20226499 ↗
  • MalfunctionSep 16, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE EIGHT SUTURES HAD NEEDLE DETACHMENT.

    Patient problems: Insufficient Information

    MAUDE report 20225562 ↗
  • MalfunctionJul 24, 2024Break; Break

    ACCORDING TO THE REPORTER, PRE-OPERATIVELY, THREE SUTURES WERE BROKEN IN A ROW. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19818135 ↗
  • MalfunctionJul 24, 2024Break; Break

    ACCORDING TO THE REPORTER, PRE-OPERATIVELY, THREE SUTURES WERE BROKEN IN A ROW. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19817924 ↗
  • MalfunctionJul 24, 2024Break; Break

    ACCORDING TO THE REPORTER, PRE-OPERATIVELY, THREE SUTURES WERE BROKEN IN A ROW. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19817901 ↗
  • MalfunctionApr 11, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING SUBCOMMISURAL APPLICATION OF AORTIC VALVE, SUPRACORONARY IMPLANTATION OF GRAFT INTERPOSITUM INTERGARD 36MM, AND IMPLANTATION OF PROSTHESIS 40MM, UPON GRAFT SUTURING, THE 10 SUTURES HAD DETACHED FROM THE NEEDLE AND THE SUTURE BROKE AT ONE POINT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. SURGICAL TIME WAS EXTENDED BY MORE THAN 30 MINUTES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19090911 ↗
  • MalfunctionApr 11, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING SUBCOMMISURAL APPLICATION OF AORTIC VALVE, SUPRACORONARY IMPLANTATION OF GRAFT INTERPOSITUM INTERGARD 36MM, AND IMPLANTATION OF PROSTHESIS 40MM, UPON GRAFT SUTURING, THE 10 SUTURES HAD DETACHED FROM THE NEEDLE AND THE SUTURE BROKE AT ONE POINT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. SURGICAL TIME WAS EXTENDED BY MORE THAN 30 MINUTES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19090626 ↗
  • MalfunctionApr 11, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING SUBCOMMISURAL APPLICATION OF AORTIC VALVE, SUPRACORONARY IMPLANTATION OF GRAFT INTERPOSITUM INTERGARD 36MM, AND IMPLANTATION OF PROSTHESIS 40MM, UPON GRAFT SUTURING, THE 10 SUTURES HAD DETACHED FROM THE NEEDLE AND THE SUTURE BROKE AT ONE POINT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. SURGICAL TIME WAS EXTENDED BY MORE THAN 30 MINUTES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19089993 ↗
  • MalfunctionApr 11, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING SUBCOMMISURAL APPLICATION OF AORTIC VALVE, SUPRACORONARY IMPLANTATION OF GRAFT INTERPOSITUM INTERGARD 36MM, AND IMPLANTATION OF PROSTHESIS 40MM, UPON GRAFT SUTURING, THE 10 SUTURES HAD DETACHED FROM THE NEEDLE AND THE SUTURE BROKE AT ONE POINT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. SURGICAL TIME WAS EXTENDED BY MORE THAN 30 MINUTES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19089802 ↗
  • MalfunctionApr 11, 2024Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING SUBCOMMISURAL APPLICATION OF AORTIC VALVE, SUPRACORONARY IMPLANTATION OF GRAFT INTERPOSITUM INTERGARD 36MM, AND IMPLANTATION OF PROSTHESIS 40MM, UPON GRAFT SUTURING, THE 10 SUTURES HAD DETACHED FROM THE NEEDLE AND THE SUTURE BROKE AT ONE POINT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. SURGICAL TIME WAS EXTENDED BY MORE THAN 30 MINUTES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19089391 ↗
  • MalfunctionSep 6, 2023Break; Break

    ACCORDING TO THE REPORTER, DURING AN OPEN CORONARY ARTERY BYPASS GRAFT (CABG) PROCEDURE, IN THE ATRIUM, THE THREAD OF TWO SUTURES BROKE. TO RESOLVE THE ISSUE, A NEW SUTURE WAS USED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17690593 ↗
  • MalfunctionApr 27, 2023Break; Break

    ACCORDING TO THE REPORTER, DURING AN OPEN LIVER TRANSPLANTATION, WHEN COPYING SUTURE LIGATION TO STOP BLEEDING VIA INTERRUPTED SUTURE, AFTER SUTURING, THE THREAD OF THE TWO SUTURES BROKE WHEN THE KNOT WAS TIED. TO RESOLVE THE ISSUE, THE SUTURE WAS REPLACED WITH ANOTHER MODEL FROM ANOTHER COMPANY.

    Patient problems: Insufficient Information

    MAUDE report 16826088 ↗
  • InjuryMar 28, 2023Break; Break

    ACCORDING TO THE REPORTER, POST-OPERATIVELY OF AN OPEN PROCEDURE FOR THE INSERTION OF A PATCH AT THE ARTERIA PROFUNDA FEMORIS, THE THREAD BROKE. THERE WAS AN EMERGENCY SURGERY DUE TO THE RUPTURE OF THE PATCH SUTURE. THERE WAS A BLOOD LOSS OF MORE THAN 500CC DUE TO THE PRODUCT PROBLEM.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 16630817 ↗
  • MalfunctionFeb 22, 2023Break; Adverse Event Without Identified Device or Use Problem; Break; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, DURING OPEN THROMBECTOMY, AT PATCH SUTURING, THE THREAD OF THE TWO SUTURES BROKE. THE OPERATION TIME WAS EXTENDED BY MORE THAN 30 MINUTES. THE SURGERY WAS COMPLETED WITH A SUTURE FROM THE SAME LOT NUMBER. THERE WAS NO PATIENT INJURY.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16419026 ↗
  • MalfunctionMar 29, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, WHILE SURGEON WAS CLOSING THE LEFT ATRIUM DURING AN OPEN PROCEDURE, THE NEEDLE DISENGAGED FROM THE THREAD. SURGEON REPLACED THE DEVICE TO RESOLVE THE ISSUE. NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13939188 ↗

Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices

All SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices →