Surgipro II VP-735-XRecalls & reported problems

Covidien LP · Monofilament Polypropylene · Class 2 · Implant

All models of Surgipro II (73)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
11
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 10884521038813), received between Apr 27, 2017 and Aug 8, 2024 — classified as 1 injury, 10 malfunctions.

The most frequently reported specific device problem is “Break” (8 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 5 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 2, 2022Class IIOpen, Classified

    Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact

    Z-1180-2023 · Root cause: Packaging process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break8
  • Detachment of Device or Device Component3

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Known Impact Or Consequence To Patient2
  • No Consequences Or Impact To Patient1
  • Tissue Damage1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 8, 2024Break; Break

    ACCORDING TO THE REPORTER, DURING A CORONARY ARTERY BYPASS GRAFT PROCEDURE, THE NEEDLE BROKE WHILE SUTURING FOR ALL 5 DEVICE. ANOTHER DEVICE WAS USED TO RESOLVE THE ISSUE. THE SURGICAL TIME WAS EXTENDED BY MORE THAN 30 MINUTES DUE TO THE DEVICE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19944599 ↗
  • MalfunctionAug 8, 2024Break; Break

    ACCORDING TO THE REPORTER, DURING A CORONARY ARTERY BYPASS GRAFT PROCEDURE, THE NEEDLE BROKE WHILE SUTURING FOR ALL 5 DEVICE. ANOTHER DEVICE WAS USED TO RESOLVE THE ISSUE. THE SURGICAL TIME WAS EXTENDED BY MORE THAN 30 MINUTES DUE TO THE DEVICE ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19944332 ↗
  • MalfunctionAug 7, 2024Break; Break

    ACCORDING TO THE REPORTER, DURING A CORONARY ARTERY BYPASS GRAFT PROCEDURE, THE NEEDLE BROKE WHILE SUTURING FOR ALL 5 DEVICE. ANOTHER DEVICE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19931619 ↗
  • MalfunctionAug 7, 2024Break; Break

    ACCORDING TO THE REPORTER, DURING A CORONARY ARTERY BYPASS GRAFT PROCEDURE, THE NEEDLE BROKE WHILE SUTURING FOR ALL 5 DEVICE. ANOTHER DEVICE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19931535 ↗
  • MalfunctionAug 7, 2024Break; Break

    ACCORDING TO THE REPORTER, DURING A CORONARY ARTERY BYPASS GRAFT PROCEDURE, THE NEEDLE BROKE WHILE SUTURING FOR ALL FIVE DEVICES. ANOTHER DEVICE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19931403 ↗
  • MalfunctionJul 13, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING AN OPEN MYOCARDIAL REVASCULARIZATION PROCEDURE, AN ENVELOPE OF THE SUTURE WAS OPENED AND WHEN REMOVING THE PRODUCT FROM THE PACKAGE, THE NEEDLE WAS LOOSE FROM THE WIRE. THEN, DURING THE BEGINNING OF THE CORONARY ANASTOMOSIS WITH THE SUTURE, AT THE FIRST PASSAGE OF THE NEEDLE INTO THE CORONARY TISSUE, THE THREAD DETACHED FROM THE NEEDLE. IT WAS NECESSARY TO RESTART THE SUTURE WITH A NEW NEEDLE PASSAGE THROUGH THE CORONARY ARTERY TISSUE. AN ADDITIONAL SUTURE FROM THE SAME BATCH WAS USED WITHOUT ISSUE. NO DEVICE COMPONENT FELL INTO THE PATIENT¿S CAVITY. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17317470 ↗
  • MalfunctionAug 18, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING A CORONARY ARTERY BYPASS GRAFT, THE FIRST TWO SUTURES WERE RUPTURED FROM THE SWAGE AND THE SECOND TWO SUTURES WERE RUPTURED FROM THE MIDDLE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15249991 ↗
  • MalfunctionJul 21, 2020Break; Break

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE THREAD BROKE. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THE EVENT OCCURRED DURING THE PROCEDURE BUT THE PRODUCT WAS NOT USED FOR PATIENT. THERE WAS NO PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10301505 ↗
  • MalfunctionJan 16, 2019Break; Break

    ACCORDING TO THE REPORTER, DURING A BYPASS PROCEDURE, THE THREAD AND THE NEEDLE BROKE. THEY CHANGED SUTURE AND SUTURED AGAIN. THERE WAS NO PATIENT INJURY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8255403 ↗
  • MalfunctionJul 12, 2018Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING A CORONARY ARTERY BYPASS GRAFT ON THE SAPHENOUS VEIN, THE SUTURE DETACHED FROM THE NEEDLE AND STRAND ON VESSEL ANASTOMOSIS. A NEW DEVICE WITH THE SAME LOT NUMBER WAS USED TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7682793 ↗
  • InjuryApr 27, 2017Break; Break

    ACCORDING TO THE REPORTER: DURING CABG TRIPLE GRAFT WITH LAD LIMA CONNECTION (OPEN PROCEDURE), SUTURES THREADS BROKE WHEN THE SURGEON THREW AN KNOT WHILE PERFORMING DISTAL ANASTOMOSIS. SUTURE WAS HYDRATED TO MAKE PASSAGE SMOOTH. THE SURGEON EMPLOYED CONTINUOUS SUTURING TECHNIQUE. THE PATIENT HAS TO BE SHIFTED FROM BEATING HEART TO ON PUMP SURGERY BECAUSE OF CONTINUOUS BREAKING OF DISTAL ANASTOMOSIS SUTURE WHICH TRIGGERED THE BLOOD PRESSURE OF THE PATIENT TO PREVENT PERMANENT IMPAIRMENT OF A FUNCTION SO IT LED TO EXTENSION IN SURGICAL TIME UP TO 2 HOURS. THE DEVICE WAS BEING APPLIED TO SAPHENOUS VEIN. THERE WAS UNANTICIPATED TISSUE LOSS AS A RESULT OF THIS PROBLEM. THERE WAS TISSUE DAMAGE DUE TO CONTINUOUS NEEDLE PRICK AS THE THREAD WAS BREAKING CONTINUOUSLY SO SURGEON HAD TO HARVEST ANOTHER VEIN FOR THE ADDITIONAL GRAFT AS A RESULT OF THIS PROBLEM. USED ANOTHER SUTURE TO RESOLVE THE ISSUE IN ORDER TO COMPLETE THE CASE. PATIENT WAS TEMPORARILY INJURED.

    Patient problems: Tissue Damage

    MAUDE report 6525343 ↗

Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices

All SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices →

Sources and updates