Skip to content

Surgipro II VP-743-MXRecalls & reported problems

Covidien LP · Monofilament Polypropylene · Class 2 · Implant

All models of Surgipro II (73)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10884521038875), received between Jul 26, 2018 and Aug 11, 2026 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Detachment of Device or Device Component2
  • Adverse Event Without Identified Device or Use Problem1
  • Unraveled Material1

Patient problems

  • No Known Impact Or Consequence To Patient2
  • Device Embedded In Tissue or Plaque1
  • Foreign Body In Patient1
  • Insufficient Information1
  • No Code Available1
  • No Consequences Or Impact To Patient1
  • Unintended Radiation Exposure1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 11, 2026Break

    Patient problems: Insufficient Information

    MAUDE report 50204796 ↗
  • InjuryJul 29, 2022Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, DURING OPEN CORONARY ARTERY BYPASS GRAFTING (CABG), AT BLOOD VESSEL ANASTOMOSIS, WHEN SUTURE WAS BEING USED WITH A BULLDOG FORCEPS, THE NEEDLE DETACHED FROM THE THREAD. THE SURGERY WAS EXTENDED BY AT LEAST THIRTY MINUTES AS IT TOOK TIME TO SEARCH THE NEEDLE BUT WAS UNABLE TO FIND IT. ANOTHER SUTURE WAS USED TO CONTINUE THE CASE.

    Patient problems: Device Embedded In Tissue or Plaque; Foreign Body In Patient; Unintended Radiation Exposure

    MAUDE report 15132597 ↗
  • InjuryApr 11, 2019Unraveled Material; Unraveled Material

    ACCORDING TO THE REPORTER, APPROXIMATELY 2 HOURS DURING AN OPEN-HEART SURGERY, DURING THE SUTURING OF A BIO PATCH USING A RUNNING TECHNIQUE, THE THREAD OF THE SUTURE BROKE. IT WAS NOTED THAT AFTER TESTING THE PATCH WITH ECHOCARDIOGRAM FOR A THIRD TIME, IT WAS DISCOVERED THAT THE 8-0 SUTURE LINE HAD OPENED. THE PROCEDURE WAS EXTENDED TO REDO THE SUTURE LINE. THE SURGICAL TIME WAS EXTENDED BY MORE THAN 30 MINUTES DUE TO THE PRODUCT PROBLEM.

    Patient problems: No Code Available

    MAUDE report 8503468 ↗
  • MalfunctionDec 26, 2018Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING A CORONARY BYPASS PROCEDURE, BOTH OF THE REPORTED DEVICE HAD TROUBLE OF NEEDLE DISCONNECTION AND THREAD CUTTING. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THE SURGICAL TIME WAS EXTENDED BY LESS THAN 30 MIN. THERE WAS NO PATIENT INJURY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8195822 ↗
  • MalfunctionJul 26, 2018Break

    ACCORDING TO THE REPORTER, WHILE USING THE DEVICE DURING A CORONARY BYPASS PROCEDURE, THE SUTURE BROKE VERY EASILY. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THE SURGICAL TIME WAS EXTENDED BY LESS THAN 30 MINUTES. THERE WAS NO PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 7721876 ↗

Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices

All SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices →

Sources and updates