Skip to content

Surgipro II VP-747-XRecalls & reported problems

Covidien LP · Monofilament Polypropylene · Class 2 · Implant

All models of Surgipro II (73)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10884521038943), received between Jun 20, 2025 and Jul 21, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 2, 2022Class IIOpen, Classified

    Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact

    Z-1180-2023 · Root cause: Packaging process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Break2
  • Component Missing1
  • Insufficient Device Problem Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 21, 2025Break; Insufficient Device Problem Information; Break; Insufficient Device Problem Information

    ACCORDING TO THE REPORTER, DURING AN OPEN BYPASS, DURING THE ANASTOMOSIS OF THE DISTAL CORONARY ARTERY, THE NEEDLE BROKE WHILE PERFORMING A CONTINUOUS SUTURING TECHNIQUE. ANOTHER SUTURE FROM THE SAME BOX PRESENTED THE SAME ISSUE. THE COMPONENT DID NOT FALL INTO THE PATIENT¿S CAVITY. A NEW SUTURE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22555387 ↗
  • MalfunctionJul 21, 2025Break; Break

    ACCORDING TO THE REPORTER, DURING AN OPEN BYPASS, DURING THE ANASTOMOSIS OF THE DISTAL CORONARY ARTERY, THE NEEDLE BROKE WHILE PERFORMING A CONTINUOUS SUTURING TECHNIQUE. ANOTHER SUTURE FROM THE SAME BOX PRESENTED THE SAME ISSUE. THE COMPONENT DID NOT FALL INTO THE PATIENT¿S CAVITY. A NEW SUTURE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22555344 ↗
  • MalfunctionJun 20, 2025Component Missing; Component Missing

    IT WAS REPORTED THAT PRIOR TO USE, THE NEEDLES WERE MISSING IN THE PACKAGE. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22296514 ↗

Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices

All SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page