Skip to content

Surgipro II VP-761-XRecalls & reported problems

Covidien LP · Monofilament Polypropylene · Class 2 · Implant

All models of Surgipro II (73)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This is a Covidien Surgipro II VP-761-X polypropylene surgical suture, a non-absorbable stitching thread used to close tissue. Its 52 reports mostly describe the thread snapping or fraying during use, and the needle coming apart from the thread; a few mention a small tip fragment going missing during surgery. Four reports describe injuries and no deaths. There are no recalls naming this device, and reports have dropped from 10 in the earlier 12 months to 5 in the most recent.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
52
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 52 adverse event reports that list this exact device (DI 10884521039148), received between Sep 18, 2017 and Jul 8, 2026 — classified as 4 injuries, 48 malfunctions.

The most frequently reported specific device problem is “Break” (38 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 10 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 2 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202610
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break38
  • Detachment of Device or Device Component14
  • Material Twisted/Bent3
  • Difficult to Open or Remove Packaging Material2
  • Material Deformation2
  • Bent1
  • Dull, Blunt1
  • Insufficient Device Problem Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions28
  • Insufficient Information11
  • No Consequences Or Impact To Patient10
  • No Known Impact Or Consequence To Patient6
  • Foreign Body In Patient2
  • No Code Available2
  • Tissue Damage2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1

Recent adverse event reports

The 25 most recent of 52 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 8, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25751895 ↗
  • MalfunctionMar 26, 2026Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING A CORONARY BYPASS PROCEDURE, DURING THE SUTURING PROCESS, THE NON-ABSORBABLE MONOFILAMENT POLYPROPYLENE SUTURE WAS USED FOR CLOSURE, AND THE NEEDLE AND SUTURE WERE DETACHED. THE SUTURE WAS REPLACED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24702204 ↗
  • MalfunctionFeb 25, 2026Break; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE THREAD BROKE WHILE TYING A KNOT. ADDITIONALLY, THE NEEDLE AND SUTURE SEPARATED AFTER PASSING THROUGH TISSUE. ANOTHER SUTURE WAS USED TO RESOLVE THE ISSUE.

    Patient problems: Insufficient Information

    MAUDE report 24447550 ↗
  • MalfunctionFeb 16, 2026Break

    ACCORDING TO THE REPORTER, THE PATIENT UNDERWENT ALLOGENEIC ORTHOTOPIC LIVER TRANSPLANTATION UNDER GENERAL ANESTHESIA. THE OPERATOR USED UNOPENED NON-ABSORBABLE MONOFILAMENT POLYPROPYLENE SUTURE, WHICH SHOWED NO VISIBLE ABNORMALITIES AND WAS WITHIN THE VALIDITY PERIOD, AND OPERATED ACCORDING TO THE INSTRUCTIONS FOR USE. THE INSTRUMENT NURSE HANDED THE PROPERLY HELD NON-ABSORBABLE MONOFILAMENT POLYPROPYLENE SUTURE TO THE SURGEON FOR VASCULAR SUTURING OF A BLEEDING VESSEL. AFTER SUTURING, WHEN PREPARING TO TIE THE KNOT, THE SUTURE BROKE, RESULTING IN THE INABILITY TO CLOSE THE VESSEL AND CAUSED INTRAOPERATIVE BLEEDING. UPON DISCOVERING THE BLEEDING, THE SURGEON IMMEDIATELY REMOVED THE BROKEN SUTURE, WHILE THE INSTRUMENT NURSE SIMULTANEOUSLY REPLACED IT WITH A NEW NON-ABSORBABLE MONOFILAMENT POLYPROPYLENE SUTURE FOR RE-SUTURING. AFTER THE NEW SUTURE WAS USED INSTEAD TO COMPLETE VASCULAR CLOSURE, THE SURGERY CONTINUED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 24349527 ↗
  • InjuryDec 12, 2025Break; Break

    ACCORDING TO THE REPORTER, DURING A CARDIAC SURGERY, THE SCRUB NURSE DISCOVERED THAT THE TIP OF THE NON-ABSORBABLE MONOFILAMENT POLYPROPYLENE SUTURE WAS MISSING A 1 X 0.5 MM FRAGMENT WHILE COLLECTING NEEDLES. THE SURGICAL TEAM WAS IMMEDIATELY NOTIFIED TO BEGIN SEARCHING, AND THE OPERATING ROOM SUPERVISOR, HEAD NURSE, AND CHIEF OF CARDIOVASCULAR SURGERY WERE INFORMED. ADDITIONAL PERSONNEL WERE DISPATCHED TO ASSIST WITH THE SEARCH. THE SCRUB NURSE AND SURGICAL TEAM, WHILE ENSURING PATIENT SAFETY, CAREFULLY INSPECTED AND USED A NEEDLE FINDER TO EXAMINE THE OPERATING TABLE, INSTRUMENT TABLE, STERILE ABDOMINAL DRAPES, SURGICAL FIELD AND INCISION AREA, AND INSIDE THE INCISION. THE CIRCULATING NURSE USED A NEEDLE FINDER TO THOROUGHLY INSPECT THE OPERATING ROOM FLOOR, UNDER THE SURGICAL BED, SURFACES OF INSTRUMENTS AND EQUIPMENT, THE CLOTHING AND SHOE SOLES OF THE SURGICAL TEAM, AND EVERY CORNER OF THE OPERATING ROOM. ALL GAUZE, SUCTION BOTTLES, AND GARBAGE BAGS WERE SEARCHED WITH NO FINDINGS. C-ARM FLUOROSCOPY OF THE SURGICAL FIELD AND THE BLOOD RECOVERY CONTAINER REVEALED NO SUSPICIOUS NEEDLE TIP IMAGES. AFTER CONSULTING THE SURGEON, THE BREAKAGE MAY HAVE OCCURRED WHEN THE NEEDLE TIP WAS

    Patient problems: Unintended Radiation Exposure

    MAUDE report 23788452 ↗
  • MalfunctionJun 11, 2025Break; Break

    IT WAS REPORTED THAT DURING THE USE OF THE SUTURE, THERE WERE ISSUES WITH THE SUTURE BREAKING AND PILLING ON THE STRAND. THE SUTURE WAS REPLACED WITH THE SAME SUTURE BUT A DIFFERENT LOT. THERE WAS NO PATIENT INJURY.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 22183567 ↗
  • MalfunctionJun 4, 2025Break; Difficult to Open or Remove Packaging Material; Break; Difficult to Open or Remove Packaging Material

    IT WAS REPORTED THAT DURING THE PROCEDURE, THE SUTURES BROKE IN THE MIDDLE WHILE BEING REMOVED FROM THE RETAINER AND PREPARED PRIOR TO PATIENT INCISION. ADDITIONALLY, THERE WAS DIFFICULTY IN REMOVING THE SUTURES FROM THE RETAINER/PACKAGING. A DIFFERENT PRODUCT OF THE SAME TYPE WAS USED TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22146727 ↗
  • MalfunctionJun 4, 2025Break; Difficult to Open or Remove Packaging Material; Break; Difficult to Open or Remove Packaging Material

    IT WAS REPORTED THAT DURING THE PROCEDURE, THE SUTURES BROKE IN THE MIDDLE WHILE BEING REMOVED FROM THE RETAINER AND PREPARED PRIOR TO PATIENT INCISION. ADDITIONALLY, THERE WAS DIFFICULTY IN REMOVING THE SUTURES FROM THE RETAINER/PACKAGING. A DIFFERENT PRODUCT OF THE SAME TYPE WAS USED TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22146184 ↗
  • MalfunctionJun 4, 2025Break; Break

    IT WAS REPORTED THAT DURING THE USE OF THE SUTURE, THERE WERE ISSUES WITH THE SUTURE BREAKING AND PILLING ON THE STRAND.

    Patient problems: Insufficient Information

    MAUDE report 22142319 ↗
  • MalfunctionJun 3, 2025Break; Break

    IT WAS REPORTED THAT DURING THE USE OF THE SUTURE, THERE WERE ISSUES WITH THE SUTURE BREAKING AND PILLING ON THE STRAND.

    Patient problems: Insufficient Information

    MAUDE report 22138242 ↗
  • MalfunctionJun 2, 2025Break; Break

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, UPON OPENING THE PACKAGING, THE SUTURE WAS BROKEN IN THE MIDDLE. THE SAME ISSUE OCCURRED ON THE OTHER FOUR SUTURES FROM THE SAME BATCH. A DIFFERENT, NON-DEFECTIVE PRODUCT OF THE SAME TYPE WAS USED TO COMPLETE THE PROC EDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22125488 ↗
  • MalfunctionJun 2, 2025Break; Break

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, UPON OPENING THE PACKAGING, THE SUTURE WAS BROKEN IN THE MIDDLE. THE SAME ISSUE OCCURRED ON THE OTHER FOUR SUTURES FROM THE SAME BATCH. A DIFFERENT, NON-DEFECTIVE PRODUCT OF THE SAME TYPE WAS USED TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22125173 ↗
  • MalfunctionJun 2, 2025Break; Break

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, UPON OPENING THE PACKAGING, THE SUTURE WAS BROKEN IN THE MIDDLE. THE SAME ISSUE OCCURRED ON THE OTHER FOUR SUTURES FROM THE SAME BATCH. A DIFFERENT, NON-DEFECTIVE PRODUCT OF THE SAME TYPE WAS USED TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22125056 ↗
  • MalfunctionJun 2, 2025Break; Break

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, UPON OPENING THE PACKAGING, THE SUTURE WAS BROKEN IN THE MIDDLE. THE SAME ISSUE OCCURRED ON THE OTHER FOUR SUTURES FROM THE SAME BATCH. A DIFFERENT, NON-DEFECTIVE PRODUCT OF THE SAME TYPE WAS USED TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22125007 ↗
  • MalfunctionJun 2, 2025Break; Break

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, UPON OPENING THE PACKAGING, THE SUTURE WAS BROKEN IN THE MIDDLE. THE SAME ISSUE OCCURRED ON THE OTHER FOUR SUTURES FROM THE SAME BATCH. A DIFFERENT, NON-DEFECTIVE PRODUCT OF THE SAME TYPE WAS USED TO COMPLETE THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22125002 ↗
  • MalfunctionJun 11, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING A PANCREATICOJEJUNOSTOMY, WHILE SUTURING THE NEEDLE AND SUTURE DETACHED. ANOTHER DEVICE WAS USED TO RESOLVE THE ISSUE.

    Patient problems: Insufficient Information

    MAUDE report 19511076 ↗
  • MalfunctionMay 10, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING A PANCREATICOJEJUNOSTOMY, WHILE SUTURING, THE NEEDLE AND SUTURE DETACHED. ANOTHER DEVICE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19290805 ↗
  • MalfunctionMay 10, 2024Detachment of Device or Device Component; Insufficient Device Problem Information; Detachment of Device or Device Component; Insufficient Device Problem Information

    ACCORDING TO THE REPORTER, DURING A PANCREATICOJEJUNOSTOMY, WHILE SUTURING THE NEEDLE AND SUTURE DETACHED. ANOTHER DEVICE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19289986 ↗
  • MalfunctionFeb 5, 2024Break; Break

    ACCORDING TO THE REPORTER, DURING AN OPEN PROCEDURE, THE THREE SUTURES BROKE, THERE WAS TENSION WHEN PULLING THE SUTURE OUT, IT BROKE PRIOR TO BEING HANDED TO THE SURGEON, AND IT ALSO BROKE DURING KNOT TYING. OPEN A NEW SUTURE TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18641612 ↗
  • MalfunctionFeb 5, 2024Break; Break

    ACCORDING TO THE REPORTER, DURING AN OPEN PROCEDURE, THE THREE SUTURES BROKE, THERE WAS TENSION WHEN PULLING THE SUTURE OUT, IT BROKE PRIOR TO BEING HANDED TO THE SURGEON, AND IT ALSO BROKE DURING KNOT TYING. OPEN A NEW SUTURE TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18641511 ↗
  • MalfunctionJan 9, 2024Break; Break

    ACCORDING TO THE REPORTER, DURING AN OPEN PROCEDURE, THE THREE SUTURES BROKE, THERE WAS TENSION WHEN PULLING THE SUTURE OUT, IT BROKE PRIOR TO BEING HANDED TO THE SURGEON, AND IT ALSO BROKE DURING KNOT TYING. OPEN A NEW SUTURE TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18472912 ↗
  • MalfunctionJan 9, 2024Break; Break

    ACCORDING TO THE REPORTER, DURING AN OPEN PROCEDURE, THE THREE SUTURES BROKE, THERE WAS TENSION WHEN PULLING THE SUTURE OUT, IT BROKE PRIOR TO BEING HANDED TO THE SURGEON, AND IT ALSO BROKE DURING KNOT TYING. OPEN A NEW SUTURE TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18472901 ↗
  • MalfunctionSep 6, 2023Break; Break

    ACCORDING TO THE REPORTER, DURING AN OPEN CORONARY ARTERY BYPASS GRAFT (CABG) PROCEDURE, IN THE ATRIUM, THE THREAD OF TWO SUTURES BROKE. TO RESOLVE THE ISSUE, A NEW SUTURE WAS USED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17690654 ↗
  • MalfunctionApr 28, 2023Break; Break

    ACCORDING TO THE REPORTER, DURING AN OPEN CORONARY ARTERY BYPASS GRAFT (CABG), THE SUTURE BROKE WHEN THE SURGEON WAS ATTACHING THE CANNULA. A COMPETITOR'S SUTURE WAS OPENED AND USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16832728 ↗
  • MalfunctionApr 27, 2023Break; Break

    ACCORDING TO THE REPORTER, DURING AN OPEN CORONARY ARTERY BYPASS GRAFT WITH MITRAL VALVE REPAIR, THE SUTURE THREAD BROKE WHEN THE SURGEON WAS ATTACHING THE CANNULA. A NEW SUTURE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16828541 ↗

Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices

All SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices →