Skip to content

Surgipro ll VP-902-XRecalls & reported problems

Covidien LP · Monofilament Polypropylene · Class 2 · Implant

All models of Surgipro ll (70)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Covidien Surgipro nonabsorbable polypropylene suture is a surgical stitching thread. Reports mostly describe the thread breaking or coming loose from the needle mid-procedure, sometimes several times in one surgery, and once separating into pieces. There are 71 reports, no deaths and 1 injury. One open recall, started December 2022, covers a packaging seal breach in the breather pouch that could allow infection or vision loss in eye surgery.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
71
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 71 adverse event reports that list this exact device (DI 10884521039568), received between Oct 25, 2017 and Aug 13, 2026 — classified as 1 injury, 70 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (41 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 20 reportsSep 2024: 8 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 3 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 2 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 2 reportsMar 2026: 0 reportsApr 2026: 4 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 2 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202620
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Dec 2, 2022Class IIOpen, Classified

    Sterile Barrier Breach of Breather Pouch may cause infection or vision loss for ophthalmic procedures in cases where there is direct patient contact

    Z-1180-2023 · Root cause: Packaging process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component41
  • Break32
  • Adverse Event Without Identified Device or Use Problem1
  • Detachment Of Device Component1
  • Dull, Blunt1
  • Manufacturing, Packaging or Shipping Problem1
  • Material Frayed1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions50
  • Insufficient Information15
  • No Consequences Or Impact To Patient8
  • No Information4
  • No Known Impact Or Consequence To Patient3
  • Unintended Radiation Exposure3
  • Foreign Body In Patient2
  • No Code Available2

Recent adverse event reports

The 25 most recent of 71 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 13, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50232033 ↗
  • MalfunctionAug 13, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50232007 ↗
  • MalfunctionApr 20, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING THE LAPAROSCOPIC PROCEDURE, WHEN THE SURGEON USED THE THREAD, IT BECAME LOOSE SEVERAL TIMES (AT LEAST FOUR), AND THE THREAD DETACHED FROM THE NEEDLE IN THE MIDDLE OF THE ANASTOMOSIS. THE NEEDLE DID NOT FALL INTO THE PATIENT¿S CAVITY. RESUMPTION OF A THREAD AND SUTURING WAS MADE IN 2 THREADS INSTEAD OF ONE. A NEW SUTURE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24926759 ↗
  • MalfunctionApr 20, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING THE LAPAROSCOPIC PROCEDURE, WHEN THE SURGEON USED THE THREAD, IT BECAME LOOSE SEVERAL TIMES (AT LEAST FOUR), AND THE THREAD DETACHED FROM THE NEEDLE IN THE MIDDLE OF THE ANASTOMOSIS. THE NEEDLE DID NOT FALL INTO THE PATIENT¿S CAVITY. RESUMPTION OF A THREAD AND SUTURING WAS MADE IN 2 THREADS INSTEAD OF ONE. A NEW SUTURE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24926633 ↗
  • MalfunctionApr 20, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING AN OPEN SURGERY, WHEN THE SURGEON USED THE THREAD, IT BECAME LOOSE SEVERAL TIMES (AT LEAST FOUR), AND THE THREAD DETACHED FROM THE NEEDLE IN THE MIDDLE OF THE ANASTOMOSIS. THE NEEDLE DID NOT FALL INTO THE PATIENT¿S CAVITY. RESUMPTION OF A THREAD AND SUTURING WAS MADE IN 2 THREADS INSTEAD OF ONE. A NEW SUTURE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24926298 ↗
  • MalfunctionApr 17, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING THE PROCEDURE, WHEN THE SURGEON USED THE THREAD, IT BECAME LOOSE SEVERAL TIMES (AT LEAST FOUR), AND THE THREAD DETACHED FROM THE NEEDLE IN THE MIDDLE OF THE ANASTOMOSIS. THE NEEDLE DID NOT FALL INTO THE PATIENT¿S CAVITY. RESUMPTION OF A THREAD AND SUTURING WAS MADE IN TWO THREADS INSTEAD OF ONE. A NEW SUTURE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24911661 ↗
  • MalfunctionFeb 20, 2026Break

    IT WAS REPORTED THAT DURING AN ANASTOMOSIS PROCEDURE, TWO THREADS OF THE SUTURE BROKE CONSECUTIVELY WHILE SUTURING. AN ADDITIONAL NEW SUTURE WAS USED WITHOUT ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24401448 ↗
  • MalfunctionFeb 20, 2026Break

    IT WAS REPORTED THAT DURING AN ANASTOMOSIS PROCEDURE, TWO THREADS OF THE SUTURE BROKE CONSECUTIVELY WHILE SUTURING. AN ADDITIONAL NEW SUTURE WAS USED WITHOUT ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24401329 ↗
  • MalfunctionJan 17, 2026Dull, Blunt; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, DURING CARDIOPULMONARY BYPASS CORONARY ARTERY BYPASS SURGERY, WHEN ANASTOMOSING THE INTERNAL MAMMARY ARTERY, THE VASCULAR SUTURE NEEDLE WAS TOO BLUNT AND COULD NOT PENETRATE THE VESSEL WALL. VASCULAR BLEEDING OF AROUND 1 TO 2 CC OCCURRED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 24105338 ↗
  • MalfunctionAug 5, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, INTRA-OPERATIVELY OF A CORONARY ARTERY BYPASS, WHILE SUTURING, THE CONNECTION OF THE NEEDLE AND THREAD OF TWO SUTURES BROKE SUCCESSIVELY. ANASTOMOSIS WAS REPEATED THREE TIMES WITH A SUTURE FROM ANOTHER LOT NUMBER TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22715819 ↗
  • MalfunctionAug 5, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, INTRA-OPERATIVELY OF A CORONARY ARTERY BYPASS, WHILE SUTURING, THE CONNECTION OF THE NEEDLE AND THREAD OF TWO SUTURES BROKE SUCCESSIVELY. ANASTOMOSIS WAS REPEATED THREE TIMES WITH A SUTURE FROM ANOTHER LOT NUMBER TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22715700 ↗
  • MalfunctionMay 6, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING THE BYPASS SURGERY, WHEN ON VASCULAR ANASTOMOSIS, THE NEEDLE BROKE ON THE SWAGE/CENTRAL PART. IMAGING TEST WAS THEN PERFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Unintended Radiation Exposure

    MAUDE report 21968953 ↗
  • MalfunctionMay 6, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING THE BYPASS SURGERY, WHEN ON VASCULAR ANASTOMOSIS, THE NEEDLE BROKE ON THE SWAGE/CENTRAL PART. IMAGING TEST WAS THEN PERFORMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Unintended Radiation Exposure

    MAUDE report 21968889 ↗
  • MalfunctionMar 28, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING A DIGESTIVE REPAIR, WHEN THE TWO SUTURES WERE REMOVED FROM THE PACKAGING, THE TWO NEEDLES OF THE THREADS BECAME LOOSE AND DETACHED. A NEW SUTURE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21724259 ↗
  • MalfunctionMar 28, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING A DIGESTIVE REPAIR, WHEN THE TWO SUTURES WERE REMOVED FROM THE PACKAGING, THE TWO NEEDLES OF THE THREADS BECAME LOOSE AND DETACHED. A NEW SUTURE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21724217 ↗
  • MalfunctionFeb 26, 2025Break; Break

    ACCORDING TO THE REPORTER, DURING AN OFF-PUMP CORONARY ARTERY BYPASS GRAFTING, THE THREAD BROKE ON ITS OWN DURING THE CRITICAL STEP OF SUTURING AND KNOTTING, REQUIRING THE OPERATION TO BE RE-SUTURED, AFFECTING THE PROGRESS OF THE OPERATION. ANOTHER SUTURE WAS REPLACED TO RESOLVE THE ISSUE.

    Patient problems: Insufficient Information

    MAUDE report 21471151 ↗
  • MalfunctionFeb 5, 2025Break; Break

    IT WAS REPORTED THAT DURING THE LIVER TRANSPLANT INVOLVING THE SUTURE, THE SUTURE BROKE TWICE WHILE SUTURING. THE BREAKAGE OCCURRED IN THE MIDDLE OF THE SUTURE. THE ISSUE WAS RESOLVED BY USING A NEW SUTURE FROM A DIFFERENT BATCH. THERE WERE NO CONSEQUENCES OR IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21314563 ↗
  • MalfunctionFeb 5, 2025Break; Break

    IT WAS REPORTED THAT DURING THE LIVER TRANSPLANT INVOLVING THE SUTURE, THE SUTURE BROKE TWICE WHILE SUTURING. THE BREAKAGE OCCURRED IN THE MIDDLE OF THE SUTURE. THE ISSUE WAS RESOLVED BY USING A NEW SUTURE FROM A DIFFERENT BATCH. THERE WERE NO CONSEQUENCES OR IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21314557 ↗
  • MalfunctionSep 25, 2024Break; Break

    ACCORDING TO THE REPORTER, DURING A CORONARY ARTERY BYPASS GRAFT PROCEDURE, THE NEEDLE BROKE WHILE SUTURING THE ANASTOMOSIS. THIS ISSUE OCCURRED ON SEVEN DEVICES. NO PIECES FALL IN THE PATIENT CAVITY. TO RESOLVE THE ISSUE, A NEW DEVICE WAS USED FROM ANOTHER BATCH. THE PROCEDURE TIME WAS EXTENDED FOR AN HOUR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20305480 ↗
  • MalfunctionSep 25, 2024Break; Break

    ACCORDING TO THE REPORTER, DURING A CORONARY ARTERY BYPASS GRAFT PROCEDURE, THE NEEDLE BROKE WHILE SUTURING THE ANASTOMOSIS. THIS ISSUE OCCURRED ON SEVEN DEVICES. NO PIECES FALL IN THE PATIENT CAVITY. TO RESOLVE THE ISSUE, A NEW DEVICE WAS USED FROM ANOTHER BATCH. THE PROCEDURE TIME WAS EXTENDED FOR AN HOUR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20305351 ↗
  • MalfunctionSep 25, 2024Break; Break

    ACCORDING TO THE REPORTER, DURING A CORONARY ARTERY BYPASS GRAFT PROCEDURE, THE NEEDLE BROKE WHILE SUTURING THE ANASTOMOSIS. THIS ISSUE OCCURRED ON SEVEN DEVICES. NO PIECES FALL IN THE PATIENT CAVITY. TO RESOLVE THE ISSUE, A NEW DEVICE WAS USED FROM ANOTHER BATCH. THE PROCEDURE TIME WAS EXTENDED FOR AN HOUR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20305154 ↗
  • MalfunctionSep 25, 2024Break; Break

    ACCORDING TO THE REPORTER, DURING A CORONARY ARTERY BYPASS GRAFT PROCEDURE, THE NEEDLE BROKE WHILE SUTURING THE ANASTOMOSIS. THIS ISSUE OCCURRED ON SEVEN DEVICES. NO PIECES FALL IN THE PATIENT CAVITY. TO RESOLVE THE ISSUE, A NEW DEVICE WAS USED FROM ANOTHER BATCH. THE PROCEDURE TIME WAS EXTENDED FOR AN HOUR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20305081 ↗
  • MalfunctionSep 25, 2024Break; Break

    ACCORDING TO THE REPORTER, DURING A CORONARY ARTERY BYPASS GRAFT PROCEDURE, THE NEEDLE BROKE WHILE SUTURING THE ANASTOMOSIS. THIS ISSUE OCCURRED ON SEVEN DEVICES. NO PIECES FALL IN THE PATIENT CAVITY. TO RESOLVE THE ISSUE, A NEW DEVICE WAS USED FROM ANOTHER BATCH. THE PROCEDURE TIME WAS EXTENDED FOR AN HOUR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20305006 ↗
  • MalfunctionSep 25, 2024Break; Break

    ACCORDING TO THE REPORTER, DURING A CORONARY ARTERY BYPASS GRAFT PROCEDURE, THE NEEDLE BROKE WHILE SUTURING THE ANASTOMOSIS. THIS ISSUE OCCURRED ON SEVEN DEVICES. NO PIECES FALL IN THE PATIENT CAVITY. TO RESOLVE THE ISSUE, A NEW DEVICE WAS USED FROM ANOTHER BATCH. THE PROCEDURE TIME WAS EXTENDED FOR AN HOUR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20304817 ↗
  • MalfunctionSep 25, 2024Break; Break

    ACCORDING TO THE REPORTER, DURING A CORONARY ARTERY BYPASS GRAFT PROCEDURE, THE NEEDLE BROKE WHILE SUTURING THE ANASTOMOSIS. THIS ISSUE OCCURRED ON SEVEN DEVICES. NO PIECES FALL IN THE PATIENT CAVITY. TO RESOLVE THE ISSUE, A NEW DEVICE WAS USED FROM ANOTHER BATCH. THE PROCEDURE TIME WAS EXTENDED FOR AN HOUR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20304793 ↗

Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices

All SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page