Skip to content

Polysorb CL-885Recalls & reported problems

Covidien LP · Braided Absorbable Suture · Class 2 · Implant

All models of Polysorb (291)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10884521042797), received between Apr 9, 2018 and Oct 18, 2024 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (K963253, K981935); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Adverse Event Without Identified Device or Use Problem1
  • Detachment of Device or Device Component1

Patient problems

  • Insufficient Information1
  • Local Reaction1
  • No Known Impact Or Consequence To Patient1
  • Unintended Radiation Exposure1
  • Wound Dehiscence1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 18, 2024Break; Break

    ACCORDING TO THE REPORTER, IN A LAPAROSCOPIC ABDOMINAL HYSTERECTOMY, WHILE CLOSING THE VAGINAL VAULT, THE TIP OF TWO NEEDLES BROKE WHILE SUTURING. ONE TIP WAS FOUND IN THE STERILE SHARPS TRAY ON THE TROLLEY AND THE OTHER WENT MISSING. AN X-RAY WAS DONE INTRAOPERATIVELY TO SCAN THE PELVIC AREA TO LOCATE THE NEEDLE. A THOROUGH INSPECTION OF THE OPEN CAVITY WAS PERFORMED.

    Patient problems: Insufficient Information

    MAUDE report 20488849 ↗
  • InjurySep 26, 2024Break; Break

    ACCORDING TO THE REPORTER, IN A LAPAROSCOPIC ABDOMINAL HYSTERECTOMY, WHILE CLOSING THE VAGINAL VAULT, THE TIP OF THE NEEDLE BROKE WHILE SUTURING. THE NEEDLE FELL INTO THE CAVITY AND WENT MISSING. AN X-RAY WAS DONE INTRAOPERATIVELY TO SCAN THE PELVIC AREA TO LOCATE THE NEEDLE. A THOROUGH INSPECTION OF THE OPEN CAVITY WAS PERFORMED. THE SURGICAL TIME WAS EXTENDED FOR APPROXIMATELY ONE HOUR.

    Patient problems: Unintended Radiation Exposure

    MAUDE report 20313795 ↗
  • InjuryMay 5, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, FOLLOWING THE OPEN CERVICAL FUSION, THE PATIENT EXPERIENCE WOUND DEHISCENCE AND SENSITIVITY USING THE TWO SUTURES.

    Patient problems: Local Reaction; Wound Dehiscence

    MAUDE report 16879433 ↗
  • MalfunctionApr 9, 2018Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, DURING SUTURING, THE NEEDLE DETACHED. NO PATIENT HARM REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7409441 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K963253, K981935. One submission often covers many models, so check the product and lot details in each record.

  • Z-2601-2018 · Jun 15, 2018 · Terminated
    Polysorb Braided Absorbable Suture 1 36" 90 cm Undyed GS-21, Product Number CL-955
  • Z-2541-2017 · May 3, 2017 · Terminated
    Sutures. POLYSORB: POLSORB 0 18 U/D PRECUT 126R, Item Number: L4; POLSORB 0 18 VIO PRECUT 226T; Item Number: L14; POLSORB 0 30 VIO PRECUT 227C, Item Number: L404; POLSORB 1 18 VI

Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices

All SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page