Polysorb GL-303Recalls & reported problems
Covidien LP · Braided Absorbable Suture · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10884521045224), received between Apr 26, 2017 and Jun 11, 2025 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (4 reports).
2 other recalls cover devices cleared under the same FDA submission (K963253, K981935); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component4
- Break3
- Component Missing1
- Difficult to Open or Remove Packaging Material1
- Packaging Problem1
- Unraveled Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Insufficient Information3
- No Known Impact Or Consequence To Patient2
- Tissue Breakdown1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 11, 2025Unraveled Material; Unraveled Material
IT WAS REPORTED THAT THE SUTURES WERE USED TO CLOSE THE ABDOMINAL WALL DURING CASTRATION PROCEDURE PERFORMED UNDER GENERAL ANESTHESIA. POST-OPERATIVELY, THE SUTURE LINE DID NOT HOLD AND THE SEAMS OPENED UP AGAIN. FOLLOW-UP SURGERY WAS REQUIRED UNDER GENERAL ANESTHESIA, WITH REPEAT ABDOMINAL SUTURES.
Patient problems: Tissue Breakdown
MAUDE report 22185395 ↗MalfunctionJan 22, 2025Break; Break
ACCORDING TO THE REPORTER, IN A LAPAROSCOPIC INGUINAL HERNIA REPAIR, DURING PERITONEAL SUTURING, THE THREAD OF THREE SUTURES BROKE IN THE CENTER. SEVERAL NEW SUTURES FROM ANOTHER LOT WERE USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21205082 ↗MalfunctionJan 22, 2025Break; Break
ACCORDING TO THE REPORTER, IN A LAPAROSCOPIC INGUINAL HERNIA REPAIR, DURING PERITONEAL SUTURING, THE THREAD OF THREE SUTURES BROKE IN THE CENTER. SEVERAL NEW SUTURES FROM ANOTHER LOT WERE USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21205076 ↗MalfunctionJan 22, 2025Break; Break
ACCORDING TO THE REPORTER, IN A LAPAROSCOPIC INGUINAL HERNIA REPAIR, DURING PERITONEAL SUTURING, THE THREAD OF THREE SUTURES BROKE IN THE CENTER. SEVERAL NEW SUTURES FROM ANOTHER LOT WERE USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21205071 ↗MalfunctionJul 24, 2023Detachment of Device or Device Component; Difficult to Open or Remove Packaging Material; Detachment of Device or Device Component; Difficult to Open or Remove Packaging Material
ACCORDING TO THE REPORTER, DURING A PROCEDURE, WHEN THE PACKAGE WAS OPENED, THE NEEDLE HAD FALLEN INTO THE PACKAGE AND COULD NOT BE TAKEN OUT BECAUSE IT COULD NOT BE CLAMPED BY THE NEEDLE HOLDER. THE SAME THING HAPPENED IN THREE QUANTITIES CONSECUTIVELY, AND WITH THE FOURTH QUANTITY THE PRODUCT IN THE PACKAGE APPEARED.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 17382072 ↗MalfunctionJul 5, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, IN A LAPAROSCOPIC COELIO RIGHT INGUINAL HERNIA PROCEDURE, THE NEEDLE WAS LOST WHEN THE THREAD WAS REMOVED THROUGH THE TROCAR. THE NEEDLE DETACHED FROM THE THREAD WHEN IT WAS WITHDRAWN THROUGH THE TROCAR. THE TEAM SEARCHED FOR THE NEEDLE DURING AND AFTER THE OPERATION BUT COULD NOT FIND IT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. IT WAS STATED THAT ANOTHER NEEDLE DETACHED FROM THE THREAD BEFOREHAND DURING THE OPERATION, BUT THE NEEDLE WAS NOT LOST. THE NEEDLE ALSO DID NOT FALL INTO THE PATIENT'S CAVITY. THERE WAS NO PATIENT INJURY.
Patient problems: Insufficient Information
MAUDE report 17259726 ↗MalfunctionJul 5, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, IN A LAPAROSCOPIC COELIO RIGHT INGUINAL HERNIA PROCEDURE, THE NEEDLE WAS LOST WHEN THE THREAD WAS REMOVED THROUGH THE TROCAR. THE NEEDLE DETACHED FROM THE THREAD WHEN IT WAS WITHDRAWN THROUGH THE TROCAR. THE TEAM SEARCHED FOR THE NEEDLE DURING AND AFTER THE OPERATION BUT COULD NOT FIND IT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. IT WAS STATED THAT ANOTHER NEEDLE DETACHED FROM THE THREAD BEFOREHAND DURING THE OPERATION, BUT THE NEEDLE WAS NOT LOST. THERE WAS NO PATIENT INJURY.
Patient problems: Insufficient Information
MAUDE report 17259645 ↗MalfunctionSep 28, 2017Component Missing; Packaging Problem; Component Missing; Packaging Problem
ACCORDING TO THE REPORTER, THE PRODUCT'S NEEDLE WAS NOT IN THE NEEDLE HOLDER AS IT SHOULD BE. THE NEEDLE COULD NOT BE TAKEN OUT IMMEDIATELY IT WAS HIDDEN IN A PLASTIC ENCLOSURE. TO TAKE THE NEEDLE OUT, ONE HAS TO OPEN THE PLASTIC ENCLOSURE. ACCORDINGLY, IT WAS DISTURBING FOR A COMPLICATED SURGERY. PATIENT INVOLVEMENT WAS UNKNOWN.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6900239 ↗MalfunctionApr 26, 2017Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER: OCCURRED DURING THE PROCEDURE, BUT WAS NOT USED ON THE PATIENT. WHEN THE SURGEON GRASPED THE NEEDLE AND ATTEMPTED TO REMOVE THE SUTURE FROM THE PACKAGING; HE NOTICED IT WAS SEPARATED FROM THE SUTURE THREAD. TO COMPLETE THE PROCEDURE, ANOTHER DEVICE WAS USED. NO PATIENT INJURY OR EXTENSION IN SURGICAL TIME. PATIENT STATUS IS NO PROBLEM.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6523824 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K963253, K981935. One submission often covers many models, so check the product and lot details in each record.
- Z-2601-2018 · Jun 15, 2018 · TerminatedPolysorb Braided Absorbable Suture 1 36" 90 cm Undyed GS-21, Product Number CL-955
- Z-2541-2017 · May 3, 2017 · TerminatedSutures. POLYSORB: POLSORB 0 18 U/D PRECUT 126R, Item Number: L4; POLSORB 0 18 VIO PRECUT 226T; Item Number: L14; POLSORB 0 30 VIO PRECUT 227C, Item Number: L404; POLSORB 1 18 VI
Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices
- Coated VICRYL VCP433HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death1,093 reports
- Polysorb CL-915Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID503 reports
- Coated VICRYL VCP493HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID435 reports
- V-Loc 180 VLOCL0316Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall376 reports
- Coated VICRYL VCP345HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death405 reports
- V-Loc 180 VLOCL0604Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall328 reports
- V-Loc 180 VLOCA008LCovidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall294 reports
- V-Loc 180 VLOCL0326Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID330 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.