Skip to content

Polysorb L-1897KRecalls & reported problems

Covidien LP · Braided Absorbable Suture · Class 2 · Implant

All models of Polysorb (291)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10884521046368), received between May 11, 2023 and Jan 15, 2026 — classified as 2 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

2 other recalls cover devices cleared under the same FDA submission (K963253, K981935); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Detachment of Device or Device Component3
  • Adverse Event Without Identified Device or Use Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Unintended Radiation Exposure1
  • Unspecified Tissue Injury1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 15, 2026Break

    IT WAS REPORTED THAT DURING A BLEPHAROPLASTY PROCEDURE, THE DISTAL TIP OF THE NEEDLE BROKE OFF WHILE SUTURING AND FELL INTO THE PATIENT CAVITY. THE BROKEN NEEDLE TIP WAS RETRIEVED USING THE LEFTOVER PART OF THE NEEDLE AND A NEEDLE DRIVER. A NEW SUTURE WAS USED WITHOUT ISSUE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24079214 ↗
  • MalfunctionNov 6, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING SUTURING THE NEEDLE BROKE AND FELL INTO THE CAVITY OF THE PATIENT AND IT WAS RETRIEVED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20616225 ↗
  • MalfunctionJun 21, 2024Break; Break

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, WHILE SUTURING, THE THREAD BROKE. ANOTHER SUTURE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19581752 ↗
  • InjuryDec 1, 2023Break; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem; Break; Detachment of Device or Device Component; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC SCROTAL HYPOSPADIAS REPAIR RM OPERATION, DURING THE FIRST SESSION, WHILE THE GRAFT TISSUE WAS BEING SUTURED TO THE VENTRAL PART OF THE PENIS WITH A 7-0 ABSORBABLE SUTURE, IT WAS NOTICED THAT 2/3 OF THE NEEDLE TIP WAS BROKEN. AN X-RAY WAS PERFORMED TO SEARCH FOR THE BROKEN NEEDLE. HOWEVER, NO BROKEN NEEDLE TIP WAS FOUND ON THE PATIENT. THE SURGICAL TIME WAS EXTENDED BY MORE THAN 30 MINUTES. ANOTHER SUTURE WITH THE SAME LOT NUMBER WAS USED TO RESOLVE THE ISSUE.

    Patient problems: Unintended Radiation Exposure

    MAUDE report 18248403 ↗
  • InjuryJun 16, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, POST OPERATIVELY, FIVE DAYS AFTER THE LASER CATARACT SURGERY, THE PATIENT DEVELOPED ABSCESS ON THE EYE. THE PATIENT EXPERIENCED UNANTICIPATED TISSUE LOSS AND TISSUE DAMAGE AS A RESULT. THE SURGEON HAD TO REDO THE SUTURE AFTER ABSCESS REDUCTION.

    Patient problems: Unspecified Tissue Injury

    MAUDE report 17148904 ↗
  • MalfunctionMay 11, 2023Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING AN OPEN VITRECTOMY PROCEDURE, AT THE PART OF THE EYE, THE SURGEON EXPERIENCED A BROKEN NEEDLE WHILE SUTURING. THE SURGEON REMOVED THE BROKEN PIECE FROM THE TISSUE IN ONE PIECE, AND THERE WERE NO REMAINING FRAGMENTS IN THE EYE. A NEW SUTURE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16916320 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K963253, K981935. One submission often covers many models, so check the product and lot details in each record.

  • Z-2601-2018 · Jun 15, 2018 · Terminated
    Polysorb Braided Absorbable Suture 1 36" 90 cm Undyed GS-21, Product Number CL-955
  • Z-2541-2017 · May 3, 2017 · Terminated
    Sutures. POLYSORB: POLSORB 0 18 U/D PRECUT 126R, Item Number: L4; POLSORB 0 18 VIO PRECUT 226T; Item Number: L14; POLSORB 0 30 VIO PRECUT 227C, Item Number: L404; POLSORB 1 18 VI

Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices

All SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page