Polysorb SL-5691GRecalls & reported problems
Covidien LP · Braided Absorbable Suture · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10884521048805), received between Apr 19, 2019 and Sep 26, 2024 — classified as 1 injury, 7 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
2 other recalls cover devices cleared under the same FDA submission (K963253, K981935); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
- Detachment of Device or Device Component3
- Dull, Blunt1
- Mechanical Problem1
- Patient-Device Incompatibility1
Patient problems
- No Consequences Or Impact To Patient4
- Insufficient Information3
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
- Swelling/ Edema1
- Unintended Radiation Exposure1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
InjurySep 26, 2024Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING AN OPEN PROCEDURE, WHILE CLOSING THE SCALP AFTER THE INCISION, THE NEEDLE CAME OFF AND FELL INTO THE PATIENT CAVITY. IMAGING WAS REQUIRED TO LOCATE THE NEEDLE BUT THE NEEDLE WAS LOST. THE SURGEON WAS UNABLE TO LOCATE IT. A DIFFERENT SUTURE WAS THEN USED. THE PATIENT WOULD HAVE TO COMEBACK AFTER THE SWELLING GOES DOWN TO REMOVE THE OBJECT. ADDITIONAL SURGERY WAS PERFORMED AS PART OF THE INTERVENTION.
Patient problems: Foreign Body In Patient; Swelling/ Edema; Unintended Radiation Exposure
MAUDE report 20311492 ↗MalfunctionOct 18, 2022Break; Dull, Blunt; Break; Dull, Blunt
ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE NEEDLE SEEMED FLIMSY OR DULL. MADE SUTURING DIFFICULT. THE SUTURE BREAKS AT THE KNOT AFTER TWO THROWS WHEN TYING A DEEP BURIED MATTRESS STITCH AND THE THREAD WILL BREAK AND THE SURGEON WILL RE-USE THE SAME NEEDLE OR THREAD AND IT WILL BREAK AGAIN.
Patient problems: Insufficient Information
MAUDE report 15626253 ↗MalfunctionAug 19, 2022Break; Break
ACCORDING TO THE REPORTER, FIVE OF THE NEEDLES BROKE.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 15259445 ↗MalfunctionDec 2, 2021Mechanical Problem; Patient-Device Incompatibility; Mechanical Problem; Patient-Device Incompatibility
ACCORDING TO THE REPORTER, POST-OPERATIVELY, THE SUTURES WERE NOT ABSORBING IN THE PATIENT. THE DOCTOR HAD TO GO IN AND CUT OUT THE SUTURES. THE SUTURE CAME UP THROUGH THE SKIN AND FORMED A HOLE THAT LOOKED LIKE A SCAB BUT THE SUTURE CAN BE SEEN IN THERE.
Patient problems: Insufficient Information
MAUDE report 12915441 ↗MalfunctionOct 1, 2019Break; Break
ACCORDING TO THE REPORTER, DURING A UMBILICAL HERNIA PROCEDURE, THE NEEDLE PART OF THE DEVICE BROKE. ANOTHER SUTURE WAS OPENED AND USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9141915 ↗MalfunctionJun 26, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC CHOLECYSTECTOMY, WHILE USING A RUNNING STITCH TO CLOSE PORT, THE NEEDLE POPPED OFF THE SUTURE. THEY OPENED ANOTHER DEVICE TO COMPLETE THE CASE. NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8733283 ↗MalfunctionMay 12, 2019Break; Break
ACCORDING TO THE REPORTER, WHEN SUTURING THE SKIN DURING AN OPEN PROCEDURE, A COUPLE OF THREAD BROKE. THEY OPENED ANOTHER SUTURE TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8602415 ↗MalfunctionApr 19, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING AN OPEN PROCEDURE, WHILE CLOSING THE SKIN, THE NEEDLE WAS BREAKING FROM THE MATERIAL CONSISTENTLY FROM THE BOX. ANOTHER SUTURE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8533516 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K963253, K981935. One submission often covers many models, so check the product and lot details in each record.
- Z-2601-2018 · Jun 15, 2018 · TerminatedPolysorb Braided Absorbable Suture 1 36" 90 cm Undyed GS-21, Product Number CL-955
- Z-2541-2017 · May 3, 2017 · TerminatedSutures. POLYSORB: POLSORB 0 18 U/D PRECUT 126R, Item Number: L4; POLSORB 0 18 VIO PRECUT 226T; Item Number: L14; POLSORB 0 30 VIO PRECUT 227C, Item Number: L404; POLSORB 1 18 VI
Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices
- Coated VICRYL VCP433HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death1,093 reports
- Polysorb CL-915Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID503 reports
- Coated VICRYL VCP493HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID435 reports
- V-Loc 180 VLOCL0316Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall376 reports
- Coated VICRYL VCP345HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death405 reports
- V-Loc 180 VLOCL0604Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall328 reports
- V-Loc 180 VLOCA008LCovidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall294 reports
- V-Loc 180 VLOCL0326Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID330 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page