Dover 13718Recalls & reported problems
Cardinal Health, Inc. · Silicone Coated Latex Foley Catheter,Pre-connected Drainage Bag · Class 2
All models of Dover (90)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10884521051539), received between Nov 12, 2019 and Nov 12, 2019 — classified as 1 malfunction.
The most frequently reported specific device problem is “Deflation Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Deflation Problem1
Patient problems
- Pain1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 12, 2019Deflation Problem; Deflation Problem
THE CUSTOMER REPORTED THAT DURING AN ABLATION PROCEDURE, THE CATHETER'S BALLOON WAS DEFLATED AND VERIFIED BY THE PHYSICIAN. DIFFICULTY OF WITHDRAWAL OF THE PROBE WAS NOTED AND WHEN IT WAS REMOVED THEY FOUND THAT THE BALLOON WAS SLIGHTLY INFLATED. A LITTLE BLEEDING WAS NOTED AT THE LEVEL OF THE MEATUS AND PAIN WAS NOTED BY THE PATIENT.
Patient problems: Pain
MAUDE report 9309538 ↗
FDA clearance and classification
The FDA regulates product code KNX (collector, urine, (and accessories) for indwelling catheter) under 21 CFR 876.5250 as a Class II device. 21 CFR 876.5250 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Dover 13718
Has the Dover 13718 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10884521051539), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the Dover 13718?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Nov 12, 2019. That report lists device problem “Deflation Problem” (1); patient problem “Pain” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Dover 13718?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Dover 13718 FDA approved?
Devices of this type (product code KNX) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Cardinal Health, Inc.
Who makes the Dover 13718, and is it still sold?
The FDA's device database lists Cardinal Health, Inc. as the company responsible for it (brand Dover, model 13718). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other COLLECTOR, URINE, (AND ACCESSORIES) FOR INDWELLING CATHETER devices
- Bardia® Closed System Urinary Drainage Bag With Anti-Reflux Chamber 802001C. R. Bard, Inc. · COLLECTOR, URINE, (AND ACCESSORIES) FOR INDWELLING CATHETER117 reports
- Bard® Urine Meter with Bag 153214AC. R. Bard, Inc. · COLLECTOR, URINE, (AND ACCESSORIES) FOR INDWELLING CATHETER109 reports
- Bard® Urinary Drainage Bag with Anti-Reflux Chamber 154002C. R. Bard, Inc. · COLLECTOR, URINE, (AND ACCESSORIES) FOR INDWELLING CATHETER108 reports
- Bard® Urine Meter with Bag 153204C. R. Bard, Inc. · COLLECTOR, URINE, (AND ACCESSORIES) FOR INDWELLING CATHETER108 reports
- SureStep™ Foley Tray System Complete Care® Add-A-Foley Catheter A399400AC. R. Bard, Inc. · COLLECTOR, URINE, (AND ACCESSORIES) FOR INDWELLING CATHETER2 recalls0 reports
- Bard® Urinary Drainage Bag 153504C. R. Bard, Inc. · COLLECTOR, URINE, (AND ACCESSORIES) FOR INDWELLING CATHETER91 reports
- Bard® Urinary Drainage Bag 153509C. R. Bard, Inc. · COLLECTOR, URINE, (AND ACCESSORIES) FOR INDWELLING CATHETER91 reports
- SureStep™ Foley Tray System Complete Care® Add-A-Foley Catheter A304400AC. R. Bard, Inc. · COLLECTOR, URINE, (AND ACCESSORIES) FOR INDWELLING CATHETER1 recall12 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page