Skip to content

Palindrome 8888145043CRecalls & reported problems

Mozarc Medical US LLC · H Chronic Catheter Kit Symmetrical Tip, Heparin Coating and Tal VenaTrac Stylet · Class 2 · Implant

All models of Palindrome (133)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10884521056305), received between Jun 12, 2017 and Jun 12, 2017 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

64 other recalls cover devices cleared under the same FDA submission (K112477); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jun 8, 2022Class IOpen, Classified

    There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.

    Z-1317-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 12, 2017Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED TO COVIDIEN ON (B)(6) 2017 THAT A CUSTOMER HAD AN ISSUE WITH A DIALYSIS CATHETER. THE CUSTOMER STATES DURING IMPLANTATION OF THE CATHETER THE DOCTOR FELT THE CATHETER WAS LEAKING AND REMOVED AND REPLACED IT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6633328 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K112477. One submission often covers many models, so check the product and lot details in each record.

  • Z-1319-2022 · Jun 8, 2022 · Open, Classified
    Palindrome HSI Chronic Catheter, Sterile, Single Use, Rx Only. For acute and chronic hemodialysis, apheresis, and infusion.
  • Z-1318-2022 · Jun 8, 2022 · Open, Classified
    Palindrome SI Chronic Catheter, Sterile, Single Use, Rx Only. For acute and chronic hemodialysis, apheresis, and infusion.
  • Z-1867-2015 · Apr 23, 2015 · Terminated
    Palindrome SI Chronic Catheter Kit 14.5 Fr x 55cm Item Code: 8888145066C
  • Z-1907-2015 · Apr 23, 2015 · Terminated
    Palindrome HSI Chronic Catheter Sport Pack w/VenaTrac Stylet 14.5 Fr x 33cm Item Code: 8888133509
  • Z-1909-2015 · Apr 23, 2015 · Terminated
    Palindrome Precision HSI Chronic Catheter Sport Pack w/VenaTrac Stylet 14.5 Fr x 23cm Item Code: 8888123409P
  • Z-1857-2015 · Apr 23, 2015 · Terminated
    Palindrome SI Chronic Catheter Sport Pack 14.5 Fr x 33 cm Item Code: 8888233508
  • Z-1882-2015 · Apr 23, 2015 · Terminated
    Palindrome SI Chronic Catheter Sport Pack w/VenaTrac Stylet 14.5 Fr x 28cm Item Code: 8888128458
  • Z-1852-2015 · Apr 23, 2015 · Terminated
    Palindrome SI Chronic Catheter Kit 14.5 Fr x 28cm Item Code: 8888245064
  • Z-1898-2015 · Apr 23, 2015 · Terminated
    Palindrome Precision HSI Chronic Catheter Kit w/VenaTrac Stylet 14.5 Fr x 23cm Item Code: 8888145048P
  • Z-1871-2015 · Apr 23, 2015 · Terminated
    Palindrome Precision SI Chronic Catheter Kit w/VenaTrac Stylet 14.5 Fr x 23cm Item Code: 8888145063CP
  • Z-1875-2015 · Apr 23, 2015 · Terminated
    Palindrome Precision SI Chronic Catheter Kit w/VenaTrac Stylet 14.5 Fr x 33cmItem Code: 8888145065CP
  • Z-1863-2015 · Apr 23, 2015 · Terminated
    Palindrome SI Chronic Catheter Kit w/VenaTrac Stylet 14.5 Fr x 28cm Item Code: 8888145064C
  • Z-1881-2015 · Apr 23, 2015 · Terminated
    Palindrome SI Chronic Catheter Sport Pack w/VenaTrac Stylet 14.5 Fr x 23cm Item Code: 8888123408
  • Z-1885-2015 · Apr 23, 2015 · Terminated
    Palindrome Precision SI Chronic Catheter Sport Pack w/VenaTrac Stylet 14.5 Fr x 23cm Item Code: 8888123408P
  • Z1856-2015 · Apr 23, 2015 · Terminated
    Palindrome SI Chronic Catheter Sport Pack 14.5 Fr x 28 cm Item Code:8888228458

…and 49 more.

Other Catheter, hemodialysis, implanted devices

All Catheter, hemodialysis, implanted devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page