Skip to content

Curity 6939Recalls & reported problems

Cardinal Health, Inc. · Gauze Sponges,12 Ply · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10884521059320), received between Dec 12, 2016 and Jun 11, 2024 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Material Frayed” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Frayed3

Patient problems

  • No Consequences Or Impact To Patient2
  • No Clinical Signs, Symptoms or Conditions1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 11, 2024

    FOREIGN MATERIAL INSIDE STERILE PACKAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19509869 ↗
  • MalfunctionMar 17, 2020Material Frayed; Material Frayed

    THE CUSTOMER REPORTED THAT THE GAUZE WAS FRAYING. ADDITIONAL INFORMATION PROVIDED BY THE CUSTOMER STATED THAT THEY USE THE GAUZE TO WIPE MULTIPLE WIRES/CATHETERS THAT ARE ENTERING THE PATIENT THROUGH A SHEATH DURING A CEREBRAL ANGIOGRAM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9843317 ↗
  • MalfunctionMar 17, 2020Material Frayed; Material Frayed

    THE CUSTOMER REPORTED THAT THE GAUZE WAS FRAYING. ADDITIONAL INFORMATION PROVIDED BY THE CUSTOMER STATED THAT THEY USE THE GAUZE TO WIPE MULTIPLE WIRES/CATHETERS THAT ARE ENTERING THE PATIENT THROUGH A SHEATH DURING A CEREBRAL ANGIOGRAM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9843296 ↗
  • MalfunctionDec 12, 2016Material Frayed

    THE CUSTOMER REPORTS THERE IS A LOT OF LINT ON THE GAUZE AND IN THE PACKAGE. THEY CAN NOT HAVE LINT IN THEIR OR AREAS. THE PRODUCT WAS NOT UNFOLDED OR UNROLLED. THE CUSTOMER NEVER USED THE PRODUCT BECAUSE WHEN THEY OPENED IT AND EXAMINED IT THERE WAS A LOT OF LINT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6165893 ↗

Other FIBER, MEDICAL, ABSORBENT devices

All FIBER, MEDICAL, ABSORBENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page