Skip to content

Monosof N-2730-KRecalls & reported problems

Covidien LP · Monofilament Nylon · Class 2 · Implant

All models of Monosof (89)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Covidien Monosof suture is a nylon stitch used in surgery. Reports mostly describe the thread breaking during stitching or coming loose from the needle, requiring a new suture and adding time to the procedure. FDA records show 24 reports naming its identifier, with no deaths and no injuries. No recalls name this device, with recent reports falling from 17 to 0.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
28
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 28 adverse event reports that list this exact device (DI 10884521076266), received between Mar 2, 2020 and Sep 22, 2026 — classified as 28 malfunctions.

The most frequently reported specific device problem is “Break” (20 reports).

2 other recalls cover devices cleared under the same FDA submission (K900531); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 4 reportsFeb 2025: 2 reportsMar 2025: 4 reportsApr 2025: 6 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 4 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break20
  • Detachment of Device or Device Component8
  • Product Quality Problem4
  • Difficult to Open or Remove Packaging Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions21
  • Insufficient Information4
  • No Information4
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 25 most recent of 28 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 22, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50860227 ↗
  • MalfunctionSep 22, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50860225 ↗
  • MalfunctionSep 22, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50860224 ↗
  • MalfunctionSep 14, 2026Break

    Patient problems: Insufficient Information

    MAUDE report 50764177 ↗
  • MalfunctionMay 7, 2025Break; Break

    ACCORDING TO THE REPORTER, DURING A CORNEA REPAIR, THE THREAD BROKE IN THE MIDDLE AFTER TWO OR THREE PASSAGES INTO THE CORNEA. THE THREAD WAS REMOVED, AND THE SURGEON STARTED AGAIN WITH A NEW SUTURE. THE PROCEDURE WAS LONGER DUE TO SUTURE REPAIR, BUT IT DID NOT EXCEED 30 MINUTES. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21978862 ↗
  • MalfunctionApr 29, 2025Break; Break

    ACCORDING TO THE REPORTER, DURING AN OPHTHALMOLOGICAL PROCEDURE, THE SUTURE THREAD BROKE WHILE SUTURING. A NEW SUTURE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21927602 ↗
  • MalfunctionApr 21, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE SUTURE STRAND DETACHED FROM THE NEEDLE, RESULTING IN AN EXTENDED SURGICAL TIME BY HALF AN HOUR. THE HEALTHCARE PROFESSIONAL HAD TO USE FIVE UNITS INSTEAD OF ONE TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21878265 ↗
  • MalfunctionApr 21, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE SUTURE STRAND DETACHED FROM THE NEEDLE, RESULTING IN AN EXTENDED SURGICAL TIME BY HALF AN HOUR. THE HEALTHCARE PROFESSIONAL HAD TO USE FIVE UNITS INSTEAD OF ONE TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21878252 ↗
  • MalfunctionApr 21, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE SUTURE STRAND DETACHED FROM THE NEEDLE, RESULTING IN AN EXTENDED SURGICAL TIME BY HALF AN HOUR. THE HEALTHCARE PROFESSIONAL HAD TO USE FIVE UNITS INSTEAD OF ONE TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21878212 ↗
  • MalfunctionApr 21, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE SUTURE STRAND DETACHED FROM THE NEEDLE, RESULTING IN AN EXTENDED SURGICAL TIME BY HALF AN HOUR. THE HEALTHCARE PROFESSIONAL HAD TO USE FIVE UNITS INSTEAD OF ONE TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21878164 ↗
  • MalfunctionApr 21, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, THE SUTURE STRAND DETACHED FROM THE NEEDLE, RESULTING IN AN EXTENDED SURGICAL TIME BY HALF AN HOUR. THE HEALTHCARE PROFESSIONAL HAD TO USE FIVE UNITS INSTEAD OF ONE TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21878114 ↗
  • MalfunctionMar 24, 2025Break; Break

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, WHEN IT WAS TIME TO SEW, THE THREAD BROKE BECAUSE IT WAS TOO THIN. SEVERAL ATTEMPTS WERE MADE WITH OTHER THREADS FROM DIFFERENT BATCH, ALL OF WHICH ALSO BROKE. A NEW SUTURE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21679027 ↗
  • MalfunctionMar 24, 2025Break; Break

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, WHEN IT WAS TIME TO SEW, THE THREAD BROKE BECAUSE IT WAS TOO THIN. SEVERAL ATTEMPTS WERE MADE WITH OTHER THREADS FROM DIFFERENT BATCH, ALL OF WHICH ALSO BROKE. A NEW SUTURE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21678714 ↗
  • MalfunctionMar 24, 2025Break; Break

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, WHEN IT WAS TIME TO SEW, THE THREAD BROKE BECAUSE IT WAS TOO THIN. SEVERAL ATTEMPTS WERE MADE WITH OTHER THREADS FROM DIFFERENT BATCH, ALL OF WHICH ALSO BROKE. A NEW SUTURE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21678628 ↗
  • MalfunctionMar 20, 2025Break; Product Quality Problem; Break; Product Quality Problem

    ACCORDING TO THE REPORTER, INTRAOPERATIVELY, IT WAS NOTED THAT THE SUTURE APPEARED DIFFERENT. THERE WAS SUTURE DISCOLORATION. THE SUTURE WAS GRAY INSTEAD OF BLACK. THE THICKNESS ALSO RESEMBLED OF A DIFFERENT SUTURE AND THE TENSILE STRENGTH WAS SIGNIFICANTLY LOWER. THE SUTURE BROKE EASILY IN THE MIDDLE WHEN A KNOT WAS MADE AND PULLED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THE SURGICAL TIME WAS EXTENDED FOR 5 MINUTES DUE TO THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21649807 ↗
  • MalfunctionFeb 21, 2025Product Quality Problem; Detachment of Device or Device Component; Difficult to Open or Remove Packaging Material; Product Quality Problem; Detachment of Device or Device Component; Difficult to Open or Remove Packaging Material

    IT WAS REPORTED THAT DURING AN OPHTHALMOLOGY PROCEDURE, THE SUTURE WAS DIFFICULT TO REMOVE FROM THE RETAINER/PACKAGING. A CHANGE IN THE SUTURE'S COLOR TO A BRIGHTER SHADE WAS ALSO OBSERVED. THE THREAD WAS NOTED TO BE MORE BRITTLE AND FRAGILE, BREAKING EASILY AT THE CONNECTION WHEN KNOTS WERE TIED. AN ADDITIONAL NEW SUTURE WAS USED WITHOUT ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21434872 ↗
  • MalfunctionFeb 13, 2025Break; Break

    ACCORDING TO THE REPORTER, DURING A CATARACT SURGERY, WHILE SUTURING, THE THREAD BROKE SPECIFICALLY IN THE MIDDLE. ANOTHER SUTURE FROM OTHER MANUFACTURER WAS USED TO RESOLVE THE ISSUE. THE SURGICAL TIME WAS EXTENDED NOT MORE THAN 30 MINUTES DUE TO THE PRODUCT PROBLEM. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21378658 ↗
  • MalfunctionJan 20, 2025Break; Break

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, WHEN IT WAS TIME TO SEW, THE THREAD BROKE BECAUSE IT WAS TOO THIN. SEVERAL ATTEMPTS WERE MADE WITH OTHER THREADS FROM DIFFERENT BATCH, ALL OF WHICH ALSO BROKE. A NEW SUTURE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21193095 ↗
  • MalfunctionJan 20, 2025Break; Break

    IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE, WHEN IT WAS TIME TO SEW, THE THREAD BROKE BECAUSE IT WAS TOO THIN. SEVERAL ATTEMPTS WERE MADE WITH OTHER THREADS FROM DIFFERENT BATCH, ALL OF WHICH ALSO BROKE. A NEW SUTURE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21191247 ↗
  • MalfunctionJan 16, 2025Break; Product Quality Problem; Break; Product Quality Problem

    IT WAS REPORTED THAT THE SUTURES BROKE, WERE TOO THIN AND BECAME LOOSE DURING SUTURING. AFTER TESTING SEVERAL THREADS, FROM THE SAME BOX, FROM THE SAME BATCH; OTHERS WERE OPENED THAT WERE OF BETTER QUALITY. THE PRODUCT WAS REPLACED BY ANOTHER PRODUCT.

    Patient problems: Insufficient Information

    MAUDE report 21168723 ↗
  • MalfunctionJan 16, 2025Break; Product Quality Problem; Break; Product Quality Problem

    IT WAS REPORTED THAT THE SUTURES BROKE, WERE TOO THIN AND BECAME LOOSE DURING SUTURING. AFTER TESTING SEVERAL THREADS, FROM THE SAME BOX, FROM THE SAME BATCH; OTHERS WERE OPENED THAT WERE OF BETTER QUALITY. THE PRODUCT WAS REPLACED BY ANOTHER MEDTRONIC PRODUCT. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21168302 ↗
  • MalfunctionDec 5, 2020Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING OPTHALMIC SURGERY, THE NEEDLE DETACHED FROM THE THREAD OF 2 SUTURES. ANOTHER SUTURE WAS USED TO RESOLVE THE ISSUE IN ORDER TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10950811 ↗
  • MalfunctionDec 5, 2020Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING OPTHALMIC SURGERY, THE NEEDLE DETACHED FROM THE THREAD OF 2 SUTURES. ANOTHER SUTURE WAS USED TO RESOLVE THE ISSUE IN ORDER TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10950756 ↗
  • MalfunctionMay 21, 2020Break; Break

    ACCORDING TO THE REPORTER, DURING AN OPEN EYE PROCEDURE, THE DEVICE THREAD BROKE EASILY. THE SURGEON THEN USED ANOTHER DEVICE TO RESOLVE THE ISSUE IN ORDER TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10079995 ↗
  • MalfunctionMar 2, 2020Break; Break

    ACCORDING TO THE REPORTER DURING A PROCEDURE THE DEVICE BROKE.

    Patient problems: No Information

    MAUDE report 9775240 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K900531. One submission often covers many models, so check the product and lot details in each record.

  • Z-1173-2023 · Dec 2, 2022 · Open, Classified
    Monofilament Nylon Suture (McKesson Branded) and Coated Braided Silk Suture (McKesson Branded) (1) S661GX 5-0 NYLON 18 DS18 (2)S683GX 4-0 SILK 18 DS18 (3) S683GX
  • Z-1174-2023 · Dec 2, 2022 · Open, Classified
    Monosof Monofilament Nylon sutures: Product Description CN-724 MONOSOF* 1 BLK 100CM GS25 X36 CN-791 MONOSOF* 2-0 BLK 100CM GCC90X36 CN-791-12 MONOSOF* 2-0 BLK 10

Other SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE devices

All SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page