Skip to content

Palindrome 8888145048CRecalls & reported problems

Mozarc Medical US LLC · HSI Chronic Catheter Kit Symmetrical Tip, Heparin Coating, Silver Ion Sleeve and Tal VenaTrac Stylet · Class 2 · Implant

All models of Palindrome (133)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10884521083042), received between Aug 12, 2021 and Oct 25, 2022 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Insufficient Flow or Under Infusion” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

64 other recalls cover devices cleared under the same FDA submission (K112477); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jun 8, 2022Class IOpen, Classified

    There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.

    Z-1319-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Insufficient Flow or Under Infusion1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 25, 2022Insufficient Flow or Under Infusion; Insufficient Flow or Under Infusion

    ACCORDING TO THE REPORTER, DURING USE, THE CATHETER HAD POOR INTERMITTENT FLOWS, RARELY OVER BFR (BLOOD FLOW RATE) 300 ML/MIN AND NE EDED REGULAR ANTI-THROMBOTIC. DOSES OF UROKINASE PUSH PAULSE WERE 5 UNITS, UROKINASE LOCK WERE 4 UNITS, ACTILYSE WERE 2 UNITS. THE CATHETER WAS NOT REPAIRED AND HAD NO LEAK. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15670006 ↗
  • MalfunctionAug 12, 2021Material Deformation; Material Deformation

    ACCORDING TO THE REPORTER, DURING DIALYSIS, BUBBLES WERE SEEN ON THE EXTENSION TUBE (OUTSIDE) OF THE CATHETER. THE CATHETER WAS NOT REPAIRED AND HAD NO LEAK. CHLORHEXIDINE DIGLUCONATE 0.5% M/V IN ALCOHOL 70% V/V WERE THE CLEANING AGENTS USED ON THE DEVICE. TEGADERM WAS THE DRESSING BEING USED. TEGO WAS NOT UTILIZED. THERE WAS NO LUER ADAPTER ISSUE. A DIALYSIS MACHINE WAS BEING UTILIZED AND ATTACHED TO THE CATHETER THRICE A WEEK. THERE WAS NO BLOOD LOSS AND NO BLOOD TRANSFUSION WAS REQUIRED. NO MEDICAL INTERVENTION/TREATMENT PROVIDED DUE TO THE EVENT. TREATMENT WAS COMPLETED. NOTHING WAS DONE TO RESOLVE THE ISSUE. THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12315585 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K112477. One submission often covers many models, so check the product and lot details in each record.

  • Z-1318-2022 · Jun 8, 2022 · Open, Classified
    Palindrome SI Chronic Catheter, Sterile, Single Use, Rx Only. For acute and chronic hemodialysis, apheresis, and infusion.
  • Z-1317-2022 · Jun 8, 2022 · Open, Classified
    Palindrome H Chronic Catheter, Sterile, Single Use, Rx Only. For acute and chronic hemodialysis, apheresis, and infusion.
  • Z-1861-2015 · Apr 23, 2015 · Terminated
    Palindrome SI Chronic Catheter Kit w/VenaTrac Stylet 14.5 Fr x 23cm Item Code: 8888145063C
  • Z-1894-2015 · Apr 23, 2015 · Terminated
    Palindrome HSI Chronic Catheter Kit w/VenaTrac Stylet 14.5 Fr x 33cm Item Code: 8888145050
  • Z-1887-2015 · Apr 23, 2015 · Terminated
    Palindrome Precision SI Chronic Catheter Sport Pack w/VenaTrac Stylet 14.5 Fr x 33cm Item Code: 8888133508P
  • Z-1870-2015 · Apr 23, 2015 · Terminated
    Palindrome Precision SI Chronic Catheter Kit w/VenaTrac Stylet 14.5 Fr x 23cm Item Code: 8888145063P
  • Z-1898-2015 · Apr 23, 2015 · Terminated
    Palindrome Precision HSI Chronic Catheter Kit w/VenaTrac Stylet 14.5 Fr x 23cm Item Code: 8888145048P
  • Z-1853-2015 · Apr 23, 2015 · Terminated
    Palindrome SI Chronic Catheter Kit 14.5 Fr x 28cm Item Code: 8888245064
  • Z-1850-2015 · Apr 23, 2015 · Terminated
    Palindrome SI Chronic Catheter Kit 14.5 Fr x 19cm Item Code: 8888245062 The Palindrome SI Chronic Catheter is indicated for acute and chronic hemodialysis, apheresis, and infusion.
  • Z-1892-2015 · Apr 23, 2015 · Terminated
    Palindrome HSI Chronic Catheter Kit w/VenaTrac Stylet 14.5 Fr x 28cm Item Code: 8888145049
  • Z-1895-2015 · Apr 23, 2015 · Terminated
    Palindrome HSI Chronic Catheter Kit w/VenaTrac Stylet 14.5 Fr x 33cm Item Code: 8888145050C
  • Z-1886-2015 · Apr 23, 2015 · Terminated
    Palindrome Precision SI Chronic Catheter Sport Pack w/VenaTrac Stylet 14.5 Fr x 28cm Item Code: 8888128458P
  • Z-1871-2015 · Apr 23, 2015 · Terminated
    Palindrome Precision SI Chronic Catheter Kit w/VenaTrac Stylet 14.5 Fr x 23cm Item Code: 8888145063CP
  • Z-1854-2015 · Apr 23, 2015 · Terminated
    Palindrome SI Chronic Catheter Kit 14.5 Fr x 28cm Item Code: 8888245064
  • Z-1901-2015 · Apr 23, 2015 · Terminated
    Palindrome Precision HSI Chronic Catheter Kit w/VenaTrac Stylet 14.5 Fr x 28cm Item Code: 8888145049CP

…and 49 more.