MAHURKAR Elite 8888221316Recalls & reported problems
Mozarc Medical US LLC · Acute Dual Lumen Catheter Tray,Straight Extensions with IC* Components,12 Fr/Ch (4.0 mm) x 16 cm · Class 2
All models of MAHURKAR Elite (64)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10884521126374), received between Aug 23, 2024 and Sep 17, 2024 — classified as 2 injuries.
The most frequently reported specific device problem is “Insufficient Flow or Under Infusion” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K120674); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Flow or Under Infusion1
- Separation Problem1
Patient problems
- Foreign Body In Patient1
- Insufficient Information1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjurySep 17, 2024Insufficient Flow or Under Infusion; Separation Problem; Insufficient Flow or Under Infusion; Separation Problem
ACCORDING TO THE REPORTER, THE PATIENT WAS ADMITTED WITH ALTERED MENTAL STATUS AND ACUTE KIDNEY INJURY, LATER DIAGNOSED WITH MICROANGIOPATHIC HEMOLYTIC ANEMIA. ON (B)(6) 2024, AN EXPERIENCED INTENSIVE CARE UNIT (ICU) PROVIDER INSERTED A LEFT INTERNAL JUGULAR DOUBLE LUMEN CENTRAL LINE CATHETER FOR PLASMAPHERESIS TREATMENT. THE PROVIDER FOLLOWED THE STANDARD ICU PRACTICE WITHOUT COMPLICATIONS. THAT NIGHT, DESPITE THE LINES FLUSHING WELL, THERE WAS POOR BLOOD RETURN AND SLUGGISH FLOW NOTED DURING PLASMAPHERESIS TREATMENT. THE APPROXIMATE AGE OF THE DEVICE WAS ONE DAY. THERE WAS NOTHING ODD NOTED ABOUT THE DEVICE WHEN REMOVED FROM PACKAGING. THERE WERE NO OTHER DEFECTS/DAMAGES FOUND ON THE PRODUCT. THERE WERE NO OTHER PRODUCTS BEING UTILIZED WITH THE DEVICE. THERE WAS NO LEAK. THERE WAS NO LUER ADAPTER ISSUE. TEGO WAS NOT UTILIZED. THE LINE WAS NOT HARD TO FLUSH WITH SYRINGE. NO ANTI COAGULATION WAS USED. NO CLEANING AGENT OF THE NEW PRODUCT/CATHETER WAS USED PRIOR TO INSERTION. CHLOROPREP SOLUTION ON PATIENT SKIN TO PREP PRIOR TO INSERTION. CONSEQUENTLY, THE ICU TEAM (DIFFERENT PROVIDER) DECIDED TO INSERT A SECOND DOUBLE LUMEN CENTRAL LINE CATHETER WITH THE SAME LOT NUMBER AND REMOVE
Patient problems: Insufficient Information
MAUDE report 20236291 ↗InjuryAug 23, 2024
THE EVENT INVOLVED AN ELDERLY FEMALE WITH A PAST MEDICAL HISTORY OF THROMBOCYTOPENIA, HYPERTENSION, AND HYPERLIPIDEMIA. THE PATIENT WAS ADMITTED WITH ALTERED MENTAL STATUS AND ACUTE KIDNEY INJURY, LATER DIAGNOSED WITH MICROANGIOPATHIC HEMOLYTIC ANEMIA. ON [REDACTED DATE], AN EXPERIENCED INTENSIVE CARE UNIT (ICU) PROVIDER INSERTED A LEFT INTERNAL JUGULAR DOUBLE LUMEN CENTRAL LINE CATHETER FOR PLASMAPHERESIS TREATMENT. THE PROVIDER FOLLOWED THE STANDARD ICU PRACTICE WITHOUT COMPLICATIONS. THAT NIGHT, DESPITE THE LINES FLUSHING WELL, THERE WAS POOR BLOOD RETURN AND SLUGGISH FLOW NOTED DURING PLASMAPHERESIS TREATMENT. CONSEQUENTLY, THE ICU TEAM DECIDED TO INSERT A SECOND DOUBLE LUMEN CENTRAL LINE CATHETER AND REMOVE THE FIRST CENTRAL LINE CATHETER. THE REMOVAL OF THE LEFT INTERNAL JUGULAR CATHETER WAS PERFORMED AT THE ICU BEDSIDE PER THE STANDARD OF PRACTICE. THE ICU PROVIDER THAT REMOVED THE CATHETER INSPECTED AND NOTED THAT THE CATHETER TIP WAS INTACT. A CHEST COMPUTERIZED TOMOGRAPHY (CT) ON [REDACTED DATE] REVEALED A FOREIGN BODY IN THE RIGHT LOWER PULMONARY ARTERY. ON [REDACTED DATE] HAD THE FOREIGN BODY REMOVED WITHOUT COMPLICATIONS. THE FOREIGN BODY WAS DETERMINED TO BE A RETAINE
Patient problems: Foreign Body In Patient
MAUDE report 20047742 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K120674. One submission often covers many models, so check the product and lot details in each record.
- Z-2319-2026 · Apr 29, 2026 · Open, ClassifiedMahurkar Elite PASS Trays; Description (Product Number/CFN): 12 FR x 13cm Acute Dual Lumen Catheter with Straight Extensions, PASS Tray (8888221413), 12 FR x 16cm Acute Dual Lumen
- Z-1007-2018 · Oct 14, 2017 · TerminatedMAHURKAR"* Elite 12 FR x 16cm Acute Dual Lumen Dialysis Catheter with Curved Extensions, IC Trays. Product Number: 8888222316.
Other CATHETER, HEMODIALYSIS, NON-IMPLANTED devices
- MAHURKAR 8888135191Mozarc Medical US LLC · CATHETER, HEMODIALYSIS, NON-IMPLANTED1 recall175 reports
- MAHURKAR 8888135241Mozarc Medical US LLC · CATHETER, HEMODIALYSIS, NON-IMPLANTED1 recall1 death100 reports
- ARROW IPN036723Teleflex Incorporated · CATHETER, HEMODIALYSIS, NON-IMPLANTED2 recalls2 reports
- ARROW IPN036412Teleflex Incorporated · CATHETER, HEMODIALYSIS, NON-IMPLANTED2 recalls1 report
- DuoGlide catheter 5625240Bard Access Systems, Inc. · CATHETER, HEMODIALYSIS, NON-IMPLANTED2 recalls1 report
- ARROW IPN036428Teleflex Incorporated · CATHETER, HEMODIALYSIS, NON-IMPLANTED2 recalls0 reports
- ARROW IPN036419Teleflex Incorporated · CATHETER, HEMODIALYSIS, NON-IMPLANTED2 recalls0 reports
- ARROW IPN927343Teleflex Incorporated · Catheter, hemodialysis, non-implanted2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page