Skip to content

MAHURKAR Elite 8888221316Recalls & reported problems

Mozarc Medical US LLC · Acute Dual Lumen Catheter Tray,Straight Extensions with IC* Components,12 Fr/Ch (4.0 mm) x 16 cm · Class 2

All models of MAHURKAR Elite (64)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10884521126374), received between Aug 23, 2024 and Sep 17, 2024 — classified as 2 injuries.

The most frequently reported specific device problem is “Insufficient Flow or Under Infusion” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K120674); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Flow or Under Infusion1
  • Separation Problem1

Patient problems

  • Foreign Body In Patient1
  • Insufficient Information1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjurySep 17, 2024Insufficient Flow or Under Infusion; Separation Problem; Insufficient Flow or Under Infusion; Separation Problem

    ACCORDING TO THE REPORTER, THE PATIENT WAS ADMITTED WITH ALTERED MENTAL STATUS AND ACUTE KIDNEY INJURY, LATER DIAGNOSED WITH MICROANGIOPATHIC HEMOLYTIC ANEMIA. ON (B)(6) 2024, AN EXPERIENCED INTENSIVE CARE UNIT (ICU) PROVIDER INSERTED A LEFT INTERNAL JUGULAR DOUBLE LUMEN CENTRAL LINE CATHETER FOR PLASMAPHERESIS TREATMENT. THE PROVIDER FOLLOWED THE STANDARD ICU PRACTICE WITHOUT COMPLICATIONS. THAT NIGHT, DESPITE THE LINES FLUSHING WELL, THERE WAS POOR BLOOD RETURN AND SLUGGISH FLOW NOTED DURING PLASMAPHERESIS TREATMENT. THE APPROXIMATE AGE OF THE DEVICE WAS ONE DAY. THERE WAS NOTHING ODD NOTED ABOUT THE DEVICE WHEN REMOVED FROM PACKAGING. THERE WERE NO OTHER DEFECTS/DAMAGES FOUND ON THE PRODUCT. THERE WERE NO OTHER PRODUCTS BEING UTILIZED WITH THE DEVICE. THERE WAS NO LEAK. THERE WAS NO LUER ADAPTER ISSUE. TEGO WAS NOT UTILIZED. THE LINE WAS NOT HARD TO FLUSH WITH SYRINGE. NO ANTI COAGULATION WAS USED. NO CLEANING AGENT OF THE NEW PRODUCT/CATHETER WAS USED PRIOR TO INSERTION. CHLOROPREP SOLUTION ON PATIENT SKIN TO PREP PRIOR TO INSERTION. CONSEQUENTLY, THE ICU TEAM (DIFFERENT PROVIDER) DECIDED TO INSERT A SECOND DOUBLE LUMEN CENTRAL LINE CATHETER WITH THE SAME LOT NUMBER AND REMOVE

    Patient problems: Insufficient Information

    MAUDE report 20236291 ↗
  • InjuryAug 23, 2024

    THE EVENT INVOLVED AN ELDERLY FEMALE WITH A PAST MEDICAL HISTORY OF THROMBOCYTOPENIA, HYPERTENSION, AND HYPERLIPIDEMIA. THE PATIENT WAS ADMITTED WITH ALTERED MENTAL STATUS AND ACUTE KIDNEY INJURY, LATER DIAGNOSED WITH MICROANGIOPATHIC HEMOLYTIC ANEMIA. ON [REDACTED DATE], AN EXPERIENCED INTENSIVE CARE UNIT (ICU) PROVIDER INSERTED A LEFT INTERNAL JUGULAR DOUBLE LUMEN CENTRAL LINE CATHETER FOR PLASMAPHERESIS TREATMENT. THE PROVIDER FOLLOWED THE STANDARD ICU PRACTICE WITHOUT COMPLICATIONS. THAT NIGHT, DESPITE THE LINES FLUSHING WELL, THERE WAS POOR BLOOD RETURN AND SLUGGISH FLOW NOTED DURING PLASMAPHERESIS TREATMENT. CONSEQUENTLY, THE ICU TEAM DECIDED TO INSERT A SECOND DOUBLE LUMEN CENTRAL LINE CATHETER AND REMOVE THE FIRST CENTRAL LINE CATHETER. THE REMOVAL OF THE LEFT INTERNAL JUGULAR CATHETER WAS PERFORMED AT THE ICU BEDSIDE PER THE STANDARD OF PRACTICE. THE ICU PROVIDER THAT REMOVED THE CATHETER INSPECTED AND NOTED THAT THE CATHETER TIP WAS INTACT. A CHEST COMPUTERIZED TOMOGRAPHY (CT) ON [REDACTED DATE] REVEALED A FOREIGN BODY IN THE RIGHT LOWER PULMONARY ARTERY. ON [REDACTED DATE] HAD THE FOREIGN BODY REMOVED WITHOUT COMPLICATIONS. THE FOREIGN BODY WAS DETERMINED TO BE A RETAINE

    Patient problems: Foreign Body In Patient

    MAUDE report 20047742 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K120674. One submission often covers many models, so check the product and lot details in each record.

  • Z-2319-2026 · Apr 29, 2026 · Open, Classified
    Mahurkar Elite PASS Trays; Description (Product Number/CFN): 12 FR x 13cm Acute Dual Lumen Catheter with Straight Extensions, PASS Tray (8888221413), 12 FR x 16cm Acute Dual Lumen
  • Z-1007-2018 · Oct 14, 2017 · Terminated
    MAHURKAR"* Elite 12 FR x 16cm Acute Dual Lumen Dialysis Catheter with Curved Extensions, IC Trays. Product Number: 8888222316.

Other CATHETER, HEMODIALYSIS, NON-IMPLANTED devices

All CATHETER, HEMODIALYSIS, NON-IMPLANTED devices →

Sources and updates