MAHURKAR Elite 8888231324Recalls & reported problems
Mozarc Medical US LLC · Acute Triple Lumen Catheter Tray Straight Extensions with IC Components · Class 2
All models of MAHURKAR Elite (64)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10884521127500), received between Jun 13, 2018 and Mar 22, 2023 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K120674); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break1
- Difficult to Remove1
- Material Fragmentation1
Patient problems
- Foreign Body In Patient1
- No Information1
- Unintended Radiation Exposure1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMar 22, 2023Material Fragmentation; Material Fragmentation
ACCORDING TO THE REPORTER, DURING INSERTION WHILE THE MEDICAL DOCTOR WAS REWIRING THE LEFT FEMORAL CENTRAL LINE FOR A TRIALYSIS LINE, THE DISTAL END OF THE GUIDEWIRE BROKE INTO 2 PIECES AND REMAINED IN THE CENTRAL LUMEN OF THE TRIALYSIS LINE. THE PORT OR THE MIDDLE ACCESS SITE (BROWN) WAS CLAMPED, LABELED, AND CAPPED TO PREVENT DISLODGEMENT OF THE WIRE. THE BROKEN PIECES WERE LEFT INSIDE THE PATIENT'S BODY. CONTINUOUS RENAL REPLACEMENT THERAPY WAS INITIATED PER ORDER AND VASCULAR SURGERY WAS CONSULTED FOR LINE REMOVAL. THE CATHETER WAS NOT REPAIRED. THERE WAS NO LEAK AND NO LUER ADAPTER ISSUE. THERE WAS NO CLEANING AGENT USED ON THE DEVICE. THE INSERTION SITE WAS TREATED WITH CHLORHEXIDINE PRIOR TO PRODUCT PLACEMENT. THERE WAS NOTHING UNUSUAL OBSERVED ON THE DEVICE PRIOR TO USE AND THERE WERE NO ABNORMAL CONDITIONS NOTED PRIOR TO THE ISSUE. NO OTHER PRODUCTS WERE UTILIZED WITH THE DEVICE. FLUSHING WAS DONE BY USING A SALINE SOLUTION AND HAD A NORMAL RESULT. THE GUIDE WIRE USED WAS THE ONE INCLUDED IN THE KIT. THERE WAS NO OCCLUSION. THERE WAS NO EXCESSIVE FORCE USED ON INSERTION AND WITHDRAWAL OF THE GUIDE WIRE. IT WAS NOT NECESSARY TO REMOVE THE GUIDE WIRE AND NEEDLE IN ONE ACTION
Patient problems: Foreign Body In Patient; Unintended Radiation Exposure
MAUDE report 16591885 ↗MalfunctionJun 13, 2018Break; Difficult to Remove; Break; Difficult to Remove
ACCORDING TO THE REPORTER, DURING THE PROCEDURE, THERE WAS RESISTANCE ON THE WIRE WHEN THE DOCTOR WAS RETRIEVING IT AFTER PLACEMENT AS WELL AS DURING INSERTION. UPON RETRIEVAL, THE WIRE FELL APART. IT WAS ALSO REPORTED THAT THERE WAS A SIGNIFICANT FORCE REQUIRED TO REMOVE THE GUIDE WIRE.
Patient problems: No Information
MAUDE report 7597379 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K120674. One submission often covers many models, so check the product and lot details in each record.
- Z-2319-2026 · Apr 29, 2026 · Open, ClassifiedMahurkar Elite PASS Trays; Description (Product Number/CFN): 12 FR x 13cm Acute Dual Lumen Catheter with Straight Extensions, PASS Tray (8888221413), 12 FR x 16cm Acute Dual Lumen
- Z-1007-2018 · Oct 14, 2017 · TerminatedMAHURKAR"* Elite 12 FR x 16cm Acute Dual Lumen Dialysis Catheter with Curved Extensions, IC Trays. Product Number: 8888222316.
Other CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED devices
- Power-Trialysis catheter 5605150Bard Access Systems, Inc. · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls13 reports
- Power-Trialysis catheter 5605240Bard Access Systems, Inc. · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls8 reports
- MAHURKAR 8888101002HPMozarc Medical US LLC · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls1 death8 reports
- Power-Trialysis Slim-Cath catheter 5855240Bard Access Systems, Inc. · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls4 reports
- Power-Trialysis catheter 5615200Bard Access Systems, Inc. · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls4 reports
- MAHURKAR 8888101003HPMozarc Medical US LLC · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls4 reports
- MAHURKAR 8888102003HPMozarc Medical US LLC · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls4 reports
- MAHURKAR 8888102004HPMozarc Medical US LLC · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page