MAHURKAR Elite 8888233316Recalls & reported problems
Mozarc Medical US LLC · Acute Triple Lumen Catheter Tray Curved Extensions with IC Components · Class 2
All models of MAHURKAR Elite (64)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10884521127722), received between Feb 14, 2023 and Feb 14, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Obstruction of Flow” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K120674); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Obstruction of Flow1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 14, 2023Obstruction of Flow; Obstruction of Flow
ACCORDING TO THE REPORTER, MD (MEDICAL DOCTOR) WAS UNABLE TO PLACE THE CATHETER OVER THE GUIDEWIRE. IT WOULD STOP RIGHT AT THE HUB AREA OF THE CATHETER. PRODUCT WAS UNABLE TO BE PLACED PROPERLY. THE CATHETER WAS OCCLUDED ONLY WITH PASSAGE OF WIRE. OTHERWISE, CATHETER WAS FLUSHED NORMALLY PRIOR TO USE. NO OTHER PRODUCTS BEING UTILIZED WITH THE DEVICE. THE GUIDE WIRE WAS NOT STUCK, THE GUIDEWIRE WAS ABLE TO BE REMOVED FROM CATHETER WHEN THE WIRE DID NOT PASS THROUGH CATHETER. IT WAS NOT NECESSARY TO REMOVE THE GUIDE WIRE TOGETHER (AT THE SAME TIME) WITH CATHETER. NO TOOLS USED TO REMOVE THE GUIDE WIRE. NO EXCESSIVE FORCE APPLIED ON THE PRODUCT. THE GUIDE WIRE WAS STILL INTACT. THE GUIDE WIRE WAS NOT FRAYED, COILED, UNRAVELED. THE GUIDE WIRE USED WAS THE ONE INCLUDED IN THE KIT. THE DIMENSION OF THE DEVICE AND THE GUIDE WIRE MATCHED WHAT WAS INDICATED ON THE LABEL. THERE WAS NO DIMENSION ISSUE NOTED. NO OTHER DEFECTS/DAMAGES FOUND ON THE PRODUCT. NOTHING UNUSUAL OBSERVED ON THE DEVICE PRIOR TO USE. THE CATHETER WAS NOT REPAIRED. THERE WAS NO LEAK. THERE WAS NO LUER ADAPTER ISSUE. NO CLEANING AGENT USED ON THE DEVICE. TEGO WAS NOT UTILIZED. THE INSERTION SITE WAS NOT TREATED PRIOR TO P
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16371324 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K120674. One submission often covers many models, so check the product and lot details in each record.
- Z-2319-2026 · Apr 29, 2026 · Open, ClassifiedMahurkar Elite PASS Trays; Description (Product Number/CFN): 12 FR x 13cm Acute Dual Lumen Catheter with Straight Extensions, PASS Tray (8888221413), 12 FR x 16cm Acute Dual Lumen
- Z-1007-2018 · Oct 14, 2017 · TerminatedMAHURKAR"* Elite 12 FR x 16cm Acute Dual Lumen Dialysis Catheter with Curved Extensions, IC Trays. Product Number: 8888222316.
Other CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED devices
- Power-Trialysis catheter 5605150Bard Access Systems, Inc. · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls13 reports
- Power-Trialysis catheter 5605240Bard Access Systems, Inc. · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls8 reports
- MAHURKAR 8888101002HPMozarc Medical US LLC · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls1 death8 reports
- Power-Trialysis Slim-Cath catheter 5855240Bard Access Systems, Inc. · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls4 reports
- Power-Trialysis catheter 5615200Bard Access Systems, Inc. · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls4 reports
- MAHURKAR 8888101003HPMozarc Medical US LLC · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls4 reports
- MAHURKAR 8888102003HPMozarc Medical US LLC · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls4 reports
- MAHURKAR 8888102004HPMozarc Medical US LLC · CATHETER, HEMODIALYSIS, TRIPLE LUMEN, NON-IMPLANTED2 recalls3 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page