Skip to content

Kendall SCD 295251Recalls & reported problems

Cardinal Health, Inc. · 700 Sequential Compression System · Class 2

All models of Kendall SCD (12)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Kendall sequential compression system is used with inflatable limb sleeves. Its 24 reports mostly describe damaged or exposed power cords, electrical problems, overheating, and occasional fires; they include 1 injury and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
24
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 24 adverse event reports that list this exact device (DI 10884521129580), received between May 11, 2017 and Feb 15, 2023 — classified as 1 injury, 23 malfunctions.

The most frequently reported specific device problem is “Cut In Material” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Cut In Material6
  • Temperature Problem6
  • Power Problem3
  • Fire2
  • Split2
  • Appropriate Device Problem Term/Code Not Available1
  • Arcing1
  • Break1

Patient problems

  • No Consequences Or Impact To Patient9
  • No Information7
  • No Patient Involvement6
  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 24 most recent of 24 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 15, 2023Fire; Fire

    THE CUSTOMER REPORTED THAT THE DEVICE CAUGHT FIRE DURING OPERATION AT ONE OF THEIR THEATRES. THERE WERE VISIBLE BURN MARKS NEAR THE POINT WHERE THE MAIN CABLE CONNECTS TO THE UNIT. THE DEVICE IS STILL INTACT AND THE EVENT DID NOT REQUIRE EXTINGUISHMENT. THERE WAS NO PATIENT HARM OR INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16382238 ↗
  • MalfunctionFeb 1, 2021Arcing; Arcing

    THE CUSTOMER REPORTED THAT THE DEVICE WAS BLOWING SOCKETS WHEN PLUGGED IN. NO PATIENT WAS HARMED. ADDITIONAL INFORMATION RECEIVED STATED IT APPEARS THAT THE UNIT TRIPPED A DISTRIBUTION PANEL. THE MAIN LEAD HAD BEEN PULLED FROM THE DEVICE AND THIS MAY HAVE CAUSED THE PANEL TO TRIP AS A CURRENT/VOLTAGE SPIKE COULD HAVE RESULTED DURING THE ACCIDENTAL DISCONNECTION. UPON EVALUATION ON 25-JAN-2021, THE POWER CORD WAS OBSERVED TO HAVE BURN MARKS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11263228 ↗
  • MalfunctionNov 12, 2019Fire; Fire

    THE CUSTOMER REPORTED THAT THE BENDING PART OF THE POWER LINE WAS ON FIRE AFTER BEING PLUGGED IN. THERE WAS NO INJURY CAUSED BY THE EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9310773 ↗
  • MalfunctionAug 10, 2018Disconnection; Power Problem; Disconnection; Power Problem

    CUSTOMER STATES THAT THE POWER CORD PLUG IS BROKEN. UPON TRIAGE IT WAS NOTED THAT THE LIVE PIN ON THE PLUG IS BROKEN.

    Patient problems: No Patient Involvement

    MAUDE report 7771981 ↗
  • MalfunctionJul 16, 2018Break; Material Split, Cut or Torn; Material Split, Cut or Torn

    THE CUSTOMER REPORTS THAT THE UNIT HAS A DAMAGED POWER CORD WITH COPPER EXPOSED.

    Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement

    MAUDE report 7689614 ↗
  • MalfunctionJun 19, 2018Cut In Material; Cut In Material

    THE CUSTOMER STATES A POWER ISSUE AND THE POWER CORD HAS EXPOSED COPPER WIRES.

    Patient problems: No Information

    MAUDE report 7613928 ↗
  • MalfunctionMay 29, 2018Cut In Material; Cut In Material

    THE CUSTOMER STATED THAT THE UNIT HAD A DAMAGED POWER CORD WITH COPPER EXPOSED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7548606 ↗
  • MalfunctionMay 22, 2018Cut In Material; Cut In Material

    THE CUSTOMER STATED THAT THERE IS POWER CORD IS DAMAGED AND HAS EXPOSED COPPER.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7534375 ↗
  • InjuryApr 9, 2018Cut In Material; Cut In Material

    UPON TRIAGE THE SERVICE TECH FOUND THE PUMP TO HAVE A DAMAGED FAN AND A DAMAGED POWER CORD WITH EXPOSED COPPER.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7406806 ↗
  • MalfunctionApr 6, 2018Power Problem; Power Problem

    UPON TRIAGE, IT WAS DISCOVERED THAT THE POWER CORD HAD AN ISSUE WITH A BROKEN LIVE PIN.

    Patient problems: No Information

    MAUDE report 7404524 ↗
  • MalfunctionMar 22, 2018Cut In Material; Cut In Material

    THE CUSTOMER STATES AN ERROR E6 OCCURRED. UPON TRIAGE IT WAS NOTED THAT THERE WERE EXPOSED COPPER WIRES ON THE POWER CORD.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7362558 ↗
  • MalfunctionFeb 26, 2018Temperature Problem; Temperature Problem

    THE CUSTOMER STATES A POWER ISSUE. UPON TRIAGE IT WAS NOTED THAT THE CONTROL BOARD IS PHYSICALLY BURNT.

    Patient problems: No Patient Involvement

    MAUDE report 7296276 ↗
  • MalfunctionFeb 16, 2018Temperature Problem; Temperature Problem

    THE CUSTOMER REPORTS THE PUMP HAS A POWER ISSUE AND IS DAMAGED. UPON TRIAGE THE SERVICE TECH FOUND THE PUMP TO HAVE A DAMAGED BED HOOK, POWER CORD MISSING A FUSE DOOR AND THE CONTROL PCB IS BURNT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7277604 ↗
  • MalfunctionJan 25, 2018Detachment Of Device Component; Detachment Of Device Component

    THE CUSTOMER STATES THAT THE LIVE PIN ON THE POWER CORD WAS MISSING. THE SERVICE CENTER CONFIRMED THE POWER CORD WAS REPLACED DUE TO IT HAD 1 DAMAGED LIVE PIN.

    Patient problems: No Patient Involvement

    MAUDE report 7218272 ↗
  • MalfunctionDec 22, 2017Cut In Material; Cut In Material

    THE CUSTOMER STATES THAT THE POWER CORD IS BROKEN. UPON TRIAGE IT WAS NOTED THAT THE COPPER WIRES WERE EXPOSED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7141582 ↗
  • MalfunctionDec 14, 2017Temperature Problem; Temperature Problem

    THE POWER CORD SHOWS DAMAGE AS THE INSULATION IS BURNT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 7117151 ↗
  • MalfunctionNov 27, 2017Power Problem; Power Problem

    ACCORDING TO THE REPORTER, THE UNIT WAS FOR PREVENTIVE MAINTENANCE WITH COMPLAINT OF DAMAGED POWER CORD AND SOFTWARE NEEDED TO BE UPGRADED. UPON TRIAGE IT WAS NOTED THAT THE LIVE PIN ON THE PLUG WAS BROKEN OFF.

    Patient problems: No Information

    MAUDE report 7061031 ↗
  • MalfunctionNov 1, 2017Temperature Problem; Temperature Problem

    ACCORDING TO THE REPORTER, THE DEVICE HAD HAS ARCING DAMAGE ON THE MAINS LEAD AND THE INLET CONNECTOR.

    Patient problems: No Information

    MAUDE report 6994226 ↗
  • MalfunctionNov 1, 2017Temperature Problem; Temperature Problem

    THE CUSTOMER STATES THE POWER CORD HAD A THERMAL ISSUE.

    Patient problems: No Information

    MAUDE report 6993077 ↗
  • MalfunctionSep 5, 2017Split; Split

    ACCORDING TO THE REPORTER, THE DEVICE WIRES WERE EXPOSED ON THE CABLE (EXPOSED COPPER) . THERE WAS NO ALLEGATION OF PATIENT DEATH OR SERIOUS INJURY. MOST OF THE PROTECTOR IS MISSING DUE TO THE CORD GOT CAUGHT ON THE BED EXPOSING WIRES.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6841374 ↗
  • MalfunctionAug 23, 2017Split; Split

    ACCORDING TO THE REPORTER, THE DEVICE'S MAIN INSULATION HAD BEEN CUT THROUGH STATING THE COPPER WIRES WERE EXPOSED ON THE MAINS LEAD.

    Patient problems: No Information

    MAUDE report 6814972 ↗
  • MalfunctionAug 23, 2017Temperature Problem; Temperature Problem

    ACCORDING TO THE REPORTER, THE MAIN INSULATION HAD BEEN CUT THROUGH EXPOSING COPPER WIRES.

    Patient problems: No Information

    MAUDE report 6814682 ↗
  • MalfunctionJun 5, 2017Thermal Decomposition of Device; Thermal Decomposition of Device

    THE CUSTOMER REPORTS: SHORT CIRCUIT BECAUSE OF GETTING WET. DURING INCOMING INSPECTION AT THE SERVICE CENTER THE FOLLOWING WAS FOUND: POWER CORD BURNT AND CHECK VALVE LEAKING.

    Patient problems: No Patient Involvement

    MAUDE report 6610622 ↗
  • MalfunctionMay 11, 2017Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    THE CUSTOMER STATE THAT UNIT WAS RETURNED FOR PREVENTATIVE MAINTENANCE AND UPON INSPECTION SERVICE DETECTED THE FOLLOWING ISSUED: BED HOOK AND POWER CORD COVER WORN, POWER CORD DAMAGED, COPPER WIRES EXPOSED.

    Patient problems: No Known Impact Or Consequence To Patient; No Patient Involvement

    MAUDE report 6559991 ↗

Other Sleeve, Limb, Compressible devices

All Sleeve, Limb, Compressible devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page