Skip to content

BIS 186-0299-RFBRecalls & reported problems

Covidien LP · BISX4 Module,Refurbished · Class 2

All models of BIS (19)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10884521130180), received between Jan 24, 2024 and Jun 10, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Incorrect Measurement” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K072286); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect Measurement2
  • Communication or Transmission Problem1
  • Device Alarm System1
  • Image Display Error/Artifact1
  • Material Split, Cut or Torn1
  • Temperature Problem1
  • Use of Device Problem1

Patient problems

  • Insufficient Information3
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 10, 2026Communication or Transmission Problem; Temperature Problem; Material Split, Cut or Torn

    IT WAS REPORTED THAT THE DEVICE WAS NOT RECOGNIZED BY THE MONITOR, AND THE DEVICE WIRE WAS PEELED. THERE WAS NO REPORTED PATIENT INVOLVEMENT OR ADVERSE OUTCOME. MEDTRONIC'S INITIAL EVALUATION OF THE INCIDENT DEVICE FOUND THAT THE PRINTED CIRCUIT BOARD SMELLED LIKE BURNING.

    Patient problems: Insufficient Information

    MAUDE report 25457410 ↗
  • MalfunctionJan 26, 2026Device Alarm System; Image Display Error/Artifact; Use of Device Problem

    IT WAS REPORTED THAT THE UNIT FAILED TO RECOGNIZE PIC (PATIENT INTERFACE CABLE), AS THE DEVICE CONTINUED TO DISPLAY 'PLEASE CONNECT PIC' DESPITE THE CONNECTION OF KNOWN GOOD CABLES AND THERE WAS NO ALARM. TRIED DIFFERENT KNOWN GOOD CABLE AND MODULE BUT STILL THE SAME. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24177341 ↗
  • MalfunctionMay 1, 2024Incorrect Measurement; Incorrect Measurement

    ACCORDING TO THE REPORTER, THE MONITOR HAD NO RELIABLE READINGS. THERE WAS NO REPORTED PATIENT INVOLVEMENT OR OUTCOME.

    Patient problems: Insufficient Information

    MAUDE report 19221185 ↗
  • MalfunctionJan 24, 2024Incorrect Measurement; Incorrect Measurement

    ACCORDING TO THE REPORTER, THE UNIT DID NOT HAVE REAL DATA. THERE WAS NO REPORTED PATIENT INVOLVEMENT AND OUTCOME.

    Patient problems: Insufficient Information

    MAUDE report 18575199 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K072286. One submission often covers many models, so check the product and lot details in each record.

  • Z-2958-2018 · Jul 26, 2018 · Open, Classified
    Li-ion battery packs used in BIS Vista and BIS View Monitoring Systems. Product Number/CFN: 186-0208 UDI-Device Identifier (GTIN/UPN): 10884521130319

Other Index-generating electroencephalograph software devices

All Index-generating electroencephalograph software devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page