V-Loc 180 VLOCL0004Recalls & reported problems
Covidien LP · Absorbable Wound Closure Device · Class 2 · Implant
All models of V-Loc 180 (61)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 10884521130357), received between Sep 20, 2017 and Jan 12, 2024 — classified as 11 malfunctions.
The most frequently reported specific device problem is “Break” (9 reports).
1 other recall cover devices cleared under the same FDA submission (K091087); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break9
- Detachment of Device or Device Component1
- Mechanical Problem1
Patient problems
- No Known Impact Or Consequence To Patient6
- No Clinical Signs, Symptoms or Conditions3
- No Consequences Or Impact To Patient1
- No Patient Involvement1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 12, 2024Break; Break
ACCORDING TO THE REPORTER, IN A LAPAROSCOPIC THORACOSCOPY-AIDED PULMONARY RESECTION, DURING THE WOUND CLOSURE (DERMIS SUTURE) USING THE CONTINUOUS DERMAL SUTURING TECHNIQUE, WHEN PASSING THROUGH THE LOOP AT THE START OF SUTURE, THE THREAD BROKE. ANOTHER PACKAGE WAS OPENED AND USED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18504566 ↗MalfunctionApr 7, 2021Break; Break
ACCORDING TO THE REPORTER, DURING OPEN HIP REPLACEMENT, UPON CLOSING OF DEEP DERMAL LAYER, THE THREE (3) SUTURES FROM THE SAME BOXES BROKE. THE SURGEON FIRST TRIED TO USE ONE OF THE DEVICES, UPON SNAPPING THE SUTURE, IT WAS DEVIATED TO AN HIGHER USP WHERE SUTURE BREAKING WAS ALSO EXPERIENCED AND BOTH OTHER ITEMS WERE RELATED TO THIS. SURGEON TOOK NEW BOXES AND THE ISSUE WAS RESOLVED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11628637 ↗MalfunctionMar 1, 2021Break; Break
ACCORDING TO THE REPORTER, DURING OPEN HIP REPLACEMENT, UPON CLOSING OF DEEP DERMAL LAYER, THE THREE (3) SUTURES FROM THE SAME BOXES BROKE. THE SURGEON FIRST TRIED TO USE ONE OF THE DEVICES, UPON SNAPPING THE SUTURE, IT WAS DEVIATED TO AN HIGHER USP WHERE SUTURE BREAKING WAS ALSO EXPERIENCED AND BOTH OTHER ITEMS WERE RELATED TO THIS. SURGEON TOOK NEW BOXES AND THE ISSUE WAS RESOLVED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11394646 ↗MalfunctionFeb 20, 2020Break; Break
ACCORDING TO THE REPORTER, WHILE SUTURING THE VAGINAL VAULT DURING A TOTAL LAPAROSCOPIC HYSTERECTOMY, THE FIRST SUTURE OPENED WAS BLUNT AND THE SURGEON WAS STRUGGLING TO GET THE SUTURE THROUGH THE TISSUE. OPENED A SECOND SUTURE, BUT THE THREAD SNAPPED AT THE END OF SUTURING. AFTER THE FIRST SUTURE SNAPPED, THE SURGEON USED A SECOND SUTURE WITH THE SAME BATCH CODE FROM THE SAME BOX IN ORDER TO COMPLETE THE CASE. THIS SUTURE ALSO SNAPPED, BUT THE SUTURING HAD FINISHED AT THAT POINT, AND WAS JUST CHECKING THE HEMOSTASIS, SO A THIRD SUTURE DID NOT OPEN. THERE WAS NO INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9732159 ↗MalfunctionJan 21, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, PRIOR TO PROCEDURE, WHEN OPENING THE PACKAGE AND TRYING TO USE THE SUTURE AT THE TIME OF CLOSING WOUND, THE NEEDLE AND THREAD WERE FOUND TO BE DISENGAGED AND UNABLE TO BE USED. THEREFORE THE DEVICE WAS STOPPED USING AND A NEW SUTURE WAS TAKEN OUT AND USED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 8265052 ↗MalfunctionApr 24, 2018Mechanical Problem; Mechanical Problem
ACCORDING TO THE REPORTER, DURING AN OPEN ORTHOPEDIC PROCEDURE, WHEN THE SURGEON WAS PERFORMING A REGULAR SUTURE THE THREE DEVICES HAD AN UNKNOWN ISSUE. THE PATIENT'S OUTCOME WAS ASKED AND WAS UNKNOWN.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7454024 ↗MalfunctionSep 28, 2017Break; Break
ACCORDING TO THE REPORTER, THE LOOP BROKE AFTER THE FIRST SUTURING. NO PATIENT INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6901616 ↗MalfunctionSep 27, 2017Break; Break
ACCORDING TO THE REPORTER, DURING THE CESAREAN SECTION PROCEDURE, THE PHYSICIAN MADE ONE STITCH ON TISSUE AND HAD THE NEEDLE GET THROUGH THE LOOP, TYING IT UP, THEN THE LOOP BROKE. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. NO HARM WAS CAUSED TO THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6897768 ↗MalfunctionSep 25, 2017Break; Break
ACCORDING TO THE REPORTER, DURING SUTURING, THE LOOP BROKE WHEN THE DOCTOR MADE 2 STITCHES AND HAVE THE NEEDLE GET THROUGH THE LOOP, TYING IT UP. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THE STATUS OF THE PATIENT IS NO PROBLEM.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6891038 ↗MalfunctionSep 25, 2017Break; Break
ACCORDING TO THE REPORTER, DURING A C-SECTION, A DOCTOR MADE ONE STITCH ON TISSUE AND HAD THE NEEDLE GET THROUGH THE LOOP, TYING IT UP, AND THEN THE LOOP BROKE. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THE STATUS OF THE PATIENT IS NO PROBLEM.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6890787 ↗MalfunctionSep 20, 2017Break; Break
ACCORDING TO THE REPORTER, DURING USE IN DERMAL SUTURE, THE LOOP BROKE. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. PATIENT STATUS IS WITH NO PROBLEM. NO ADVERSE EVENTS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6880655 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K091087. One submission often covers many models, so check the product and lot details in each record.
- Z-2919-2026 · Jun 24, 2026 · Open, ClassifiedV-Loc 180 Absorbable Wound Closure Device BOX BEDSIDE STAPLING IS KING BOX06614V4 BOX BY PASS FINAL BOX07405V1 BOX BYPASS CARDON BOX01056V2 BOX BYPASS DR LAGAE BOX0
Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices
- Coated VICRYL VCP433HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death1,093 reports
- Polysorb CL-915Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID503 reports
- Coated VICRYL VCP493HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID435 reports
- V-Loc 180 VLOCL0316Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall376 reports
- Coated VICRYL VCP345HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death405 reports
- V-Loc 180 VLOCL0604Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall328 reports
- V-Loc 180 VLOCA008LCovidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall294 reports
- V-Loc 180 VLOCL0326Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID330 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page