Skip to content

BIS 186-0200Recalls & reported problems

Covidien LP · Pediatric Sensor For use with BIS Monitoring Systems · Class 2

All models of BIS (19)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10884521134331), received between Mar 1, 2021 and Feb 4, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Patient Device Interaction Problem4
  • Defective Component1
  • Erratic or Intermittent Display1
  • Low Readings1

Patient problems

  • Skin Inflammation/ Irritation2
  • Burn(s)1
  • Insufficient Information1
  • No Clinical Signs, Symptoms or Conditions1
  • Pressure Sore/Ulcer1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 4, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT A PEDIATRIC PATIENT IN THE INTENSIVE CARE UNIT HAD PRESSURE ULCERS FROM THE SENSOR. THE PATIENT, WHO WAS PARALYZED, HAD THE SENSOR APPLIED TO THE FOREHEAD AFTER THE AREA WAS CLEANED WITH AN ALCOHOL WIPE. THE SENSOR WAS USED FROM 8 PM ON (B)(6), 2025, UNTIL 10 AM ON (B)(6) 2025. UPON REMOVAL, RED SORES AND SCABS WERE OBSERVED ON THE PATIENT'S FOREHEAD, INDICATING THAT THE SENSOR HAD BROKEN THE SKIN AND CAUSED A PRESSURE ULCER. THE PATIENT DID NOT HAVE ANY PREEXISTING SKIN CONDITIONS OR KNOWN ALLERGIES, AND THE SENSOR WAS USED ACCORDING TO THE INSTRUCTIONS FOR USE AND PROTOCOL. THE SKIN INTEGRITY ISSUE PERSISTED FOR THREE WEEKS.

    Patient problems: Skin Inflammation/ Irritation

    MAUDE report 21309187 ↗
  • MalfunctionJan 30, 2025Patient Device Interaction Problem; Patient Device Interaction Problem

    IT WAS REPORTED THAT DURING USE, A CHILD EXPERIENCED WOUNDS AND THREE BURNS ON THE FOREHEAD AFTER USING SENSOR. THE PATCH HAS BEEN CHANGED SEVERAL TIMES.

    Patient problems: Burn(s)

    MAUDE report 21272709 ↗
  • MalfunctionDec 12, 2024Low Readings; Low Readings

    ACCORDING TO THE REPORTER DURING USE, THE BIS VALUE WAS DISPLAYED AS 0 AT THE STAGE WHEN IT WAS WAS NOT UNDER ANESTHESIA. THERE WAS NO REPORTED PATIENT OUTCOME.

    Patient problems: Insufficient Information

    MAUDE report 20912975 ↗
  • MalfunctionMay 5, 2023Erratic or Intermittent Display; Defective Component; Patient Device Interaction Problem; Erratic or Intermittent Display; Defective Component; Patient Device Interaction Problem

    ACCORDING TO THE REPORTER, DURING USE, THE DEVICE HAD ERRATIC READING. IT WAS NOT WORKING EFFECTIVELY FOR THE FULL LENGTH OF THE CASE REQUIRING SIGNIFICANT MANIPULATION TO WORK. THE USER PUT ADDITIONAL WATER BASED GEL ON EACH SENSOR, APPLIED IT TO FOREHEAD, HOLD IT IN PLACE WITH ADDITIONAL WIDE AREA ADHESIVE TAPE, APPLIED GENTLE PRESSURE TO EXPEL ANY AIR AT THE INTERFACE AND WRAPPING THE STEM OF THE SENSOR IN GAUZE, TO ALLOW THE MONITOR TO WORK EFFECTIVELY AND STICK EASILY. THE PATIENT RECOVERED DESPITE THE PEDIATRIC SENSORS LEAVING MARKS ON THE SKIN.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Skin Inflammation/ Irritation

    MAUDE report 16876206 ↗
  • MalfunctionMar 1, 2021Patient Device Interaction Problem; Patient Device Interaction Problem

    ACCORDING TO THE REPORTER, POST OPERATIVELY, THE DEVICE WAS UTILIZED DURING A LIVER TRANSPLANT IN A PATIENT WITH HYPERBILIRUBINEMIA. THE SURGERY TOOK 10 HOURS AND 30 MINUTES OF DURATION AND AFTER THE END OF SURGERY AND DEVICE DISCARDED, THE MEDICAL TEAM SAW THE SKIN INJURIES WHERE DEVICE WAS PLACED. THE SKIN INJURY INCLUDED BLISTERS ON THE SKIN IN A CIRCLE FORMAT THAT WAS SIMILAR TO THE INSIDE OF THE DEVICE'S ELECTRODES. AFTER A WEEK THE PATIENT DOES NOT HAVE ANY PROBLEMS IN THE LOCATION, THE BLISTERS ON THE SKIN DISAPPEARED.

    Patient problems: Pressure Sore/Ulcer

    MAUDE report 11395290 ↗

FDA clearance and classification

Cleared through 510(k) K143506 on Apr 16, 2015 (decision: Substantially Equivalent). Submitted by Covidien as “Bis Standard Sensor,1-Channel (3 Electrodes); Bis Pedi Sensor,1-Channel,3 Electrodes; Bis Quatro Sensor; Bis Pedi-Xp Sensor,2-Channel (4 Electrodes); Bis Extend Sensor; Bis Bilateral Sensor,4-Channel (6 Elecrodes); Bis Insight Sensor,Cvi-Enabled,2-Channel (4 Electrodes)”. The FDA review took 127 days. FDA record ↗

Cleared through Special 510(k) K041670 on Jul 6, 2004 (decision: Substantially Equivalent). Submitted by Aspect Medical Systems, Inc. as “Bis Pediatric Sensor”. The FDA review took 15 days. FDA record ↗

The FDA regulates product code GXY (electrode, cutaneous) under 21 CFR 882.1320 as a Class II device. 21 CFR 882.1320 ↗

Common questions about the BIS 186-0200

Has the BIS 186-0200 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10884521134331), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the BIS 186-0200?

The FDA's MAUDE database lists 5 adverse event reports for this device, received between Mar 1, 2021 and Feb 4, 2025. In those reports, the most common specific device problems are “Patient Device Interaction Problem” (4), “Defective Component” (1) and “Erratic or Intermittent Display” (1); the most common patient problems are “Skin Inflammation/ Irritation” (2), “Burn(s)” (1) and “Pressure Sore/Ulcer” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the BIS 186-0200?

None of the reports are classified as deaths, 0 as injuries and 5 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the BIS 186-0200 FDA approved?

It was cleared through Special 510(k) K041670 on Jul 6, 2004; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the BIS 186-0200 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the BIS 186-0200, and is it still sold?

The FDA's device database lists Covidien LP as the company responsible for it (brand BIS, model 186-0200). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Electrode, Cutaneous devices

All Electrode, Cutaneous devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page