V-Loc 90 VLOCM0844Recalls & reported problems

Covidien LP · Absorbable Wound Closure Device · Class 2 · Implant

All models of V-Loc 90 (56)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 10884521139596), received between Sep 18, 2018 and Mar 10, 2025 — classified as 11 malfunctions.

The most frequently reported specific device problem is “Break” (6 reports).

1 other recall cover devices cleared under the same FDA submission (K100257); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break6
  • Detachment of Device or Device Component6
  • Device Slipped1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Insufficient Information3
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 10, 2025Break; Break

    IT WAS REPORTED THAT DURING A SUTURING PROCEDURE INVOLVING THE DEVICE, THE SUTURE BROKE WHILE IN USE. A SECOND SUTURE WAS USED FROM THE SAME PACK ALSO SNAPPED DURING THE SAME EVENT WITH THE SAME PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21561688 ↗
  • MalfunctionMar 10, 2025Break; Break

    IT WAS REPORTED THAT DURING A SUTURING PROCEDURE INVOLVING THE DEVICE, THE SUTURE BROKE WHILE IN USE. A SECOND SUTURE WAS USED FROM THE SAME PACK ALSO SNAPPED DURING THE SAME EVENT WITH THE SAME PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21561681 ↗
  • MalfunctionOct 30, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE NEEDLE DETACHED. THIS ISSUE OCCURRED ON TWO DEVICES. TO RESOLVE THE ISSUE, A NEW SUTURE WAS USED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20568791 ↗
  • MalfunctionOct 30, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE NEEDLE DETACHED. THIS ISSUE OCCURRED ON TWO DEVICES. TO RESOLVE THE ISSUE, A NEW SUTURE WAS USED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20568331 ↗
  • MalfunctionApr 26, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, INTRA-OPERATIVELY, THE LIGATURE WIRE BROKE AND LOOSENED SEVERAL TIMES. THE THREAD CAME OFF THE NEEDLE WITHOUT BEING POWERED ON. THE ISSUE OCCURRED ON THREE (3) SUTURES FROM THE SAME BATCH. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THE ISSUE CAUSED LOSS OF TIME.

    Patient problems: Insufficient Information

    MAUDE report 16815399 ↗
  • MalfunctionApr 26, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, INTRA-OPERATIVELY, THE LIGATURE WIRE BROKE AND LOOSENED SEVERAL TIMES. THE THREAD CAME OFF THE NEEDLE WITHOUT BEING POWERED ON. THE ISSUE OCCURRED ON THREE (3) SUTURES FROM THE SAME BATCH. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THE ISSUE CAUSED LOSS OF TIME.

    Patient problems: Insufficient Information

    MAUDE report 16815358 ↗
  • MalfunctionMar 24, 2023Break; Device Slipped; Detachment of Device or Device Component; Break; Device Slipped; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, INTRA-OPERATIVELY, THE LIGATURE WIRE BROKE AND LOOSENED SEVERAL TIMES. THE THREAD CAME OFF THE NEEDLE WITHOUT BEING POWERED ON. THE ISSUE OCCURRED ON THREE (3) SUTURES FROM THE SAME BATCH. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THE ISSUE CAUSED LOSS OF TIME.

    Patient problems: Insufficient Information

    MAUDE report 16612580 ↗
  • MalfunctionNov 22, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING AN INGUINAL HERNIA OPERATION, THE THREAD FROM THE THREE SUTURES CAME OFF THE NEEDLE WITHOUT TENSION, NO STITCHED WERE MADE. THERE WAS LOSS OF TIME DUE TO THE MALFUNCTION. A NEW SUTURE FROM DIFFERENT BOX WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15845918 ↗
  • MalfunctionJul 29, 2022Break; Break

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE THREAD OF THE DEVICE BROKE. THE DEVICE WAS NOT USED ON THE PATIENT. TO RESOLVE THE ISSUE, A NEW DEVICE WAS USED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15132224 ↗
  • MalfunctionJul 26, 2022Break; Break

    ACCORDING TO THE REPORTER, DURING A LEFT INGUINAL HERNIA REPAIR WITH MESH PROCEDURE, THE THREAD OF THE DEVICE BROKE. A SECOND DEVICE WAS OPENED, AND THE SURGEON MADE FEW STITCHES AND THE DEVICE BROKE AGAIN. TO RESOLVE THE ISSUE, A NEW DEVICE WAS USED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15104086 ↗
  • MalfunctionSep 18, 2018Break; Break

    ACCORDING TO THE REPORTER, TWO AND HALF HOUR INTO A ROBOTIC LAPAROSCOPIC GASTRIC BYPASS PROCEDURE AND THEY WERE HALF WAY THROUGH USING THE SUTURE IN THE JEJUNUM OF THE PATIENT WHEN THE THREAD OF THE SUTURE BROKE. THEY NOTED THAT THE SUTURING TECHNIQUE THEY WERE USING WAS RUNNING STITCH. THEY TIED THE THREAD AND THEY CONTINUED TO USE THE SUTURE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7886028 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K100257. One submission often covers many models, so check the product and lot details in each record.

  • Z-2918-2026 · Jun 24, 2026 · Open, Classified
    V-Loc 90 Absorbable Wound Closure Device VLOC 90 0 VL 22 GS-21 VLOCM0336 VLOC 90 0 VL 22 GS-25 VLOCM0436 VLOCM0436 VLOCM0003 V-LOC* 90 4-0 CLR 15CM P12X12 VLOCM0003 VL

Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices

All SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page