Skip to content

Palindrome 8888145060PRecalls & reported problems

Mozarc Medical US LLC · Precision Chronic Catheter Kit,Symmetrical Tip, Pre-Curved Shaft and Tal VenaTrac Stylet,14.5 Fr/Ch (4.8 mm) x 28 cm · Class 2 · Implant

All models of Palindrome (133)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10884521158306), received between Sep 15, 2021 and Sep 15, 2021 — classified as 1 malfunction.

The most frequently reported specific device problem is “Obstruction of Flow” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

48 other recalls cover devices cleared under the same FDA submission (K123196); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jun 8, 2022Class IOpen, Classified

    There is a potential leaking condition within the hub of specific chronic dialysis catheters, due to an inter-lumen void in the hub component, which may result in unanticipated fluid return during use.

    Z-1312-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Obstruction of Flow1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 15, 2021Obstruction of Flow; Obstruction of Flow

    ACCORDING TO THE REPORTER, DURING PROCEDURE, THE STYLET WOULD NOT PASS THROUGH THE LUMEN. STYLETS WERE THREADED WITH THE CORRECT ORIENTATION, BUT SOMETHING IN ONE OF THE LUMENS PREVENTED THE STYLET FROM THREADING COMPLETELY THROUGH THE LUMEN FROM ONE END TO THE OTHER - OCCLUSION WAS NOTED. NOTHING UNUSUAL OBSERVED DURING VISUAL INSPECTION. NO DAMAGE NOTED ON THE DEVICE PACKAGING. THE BOX THAT THE DEVICE CAME IN WAS INTACT. THE CATHETER WAS NOT REPAIRED AND HAD NO LEAK. TEGO WAS NOT UTILIZED. THERE WAS NO LUER ADAPTER ISSUE. THE STYLET WAS PART OF THE KIT BEING USED. NO OTHER PRODUCTS BEING UTILIZED WITH THE DEVICE. THERE WAS NO BLOOD LOSS AND NO BLOOD TRANSFUSION WAS REQUIRED. NO MEDICAL INTERVENTION/TREATMENT PROVIDED DUE TO THE EVENT. IT WAS SAID THAT THE PROCEDURE WAS PERFORMED AND COMPLETED WITH ANOTHER CATHETER (THEY PULLED OUT ANOTHER ONE). THERE WAS NO REPORTED PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12475361 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K123196. One submission often covers many models, so check the product and lot details in each record.

  • Z-1888-2025 · Apr 23, 2025 · Open, Classified
    Palindrome Precision H Chronic Hemodialysis Catheter Sport Pack. Radiopaque urethane catheter.
  • Z-1889-2025 · Apr 23, 2025 · Open, Classified
    MAHURKAR Chronic Hemodialysis Catheter Sport Pack. Radiopaque urethane catheter.
  • Z-1887-2025 · Apr 23, 2025 · Open, Classified
    Palindrome Precision SI Chronic Hemodialysis Catheter Sport Pack. Radiopaque urethane catheter.
  • Z-1886-2025 · Apr 23, 2025 · Open, Classified
    Palindrome Precision HSI Chronic Hemodialysis Catheter Sport Pack. Radiopaque urethane catheter.
  • Z-1885-2025 · Apr 23, 2025 · Open, Classified
    Palindrome Precision Chronic Hemodialysis Catheter Sport Pack. Radiopaque urethane catheter.
  • Z-1634-2024 · Mar 15, 2024 · Open, Classified
    Palindrome Precision HSI Chronic Catheter Sport Pack 14.5 Fr/Ch (4.8mm) x 19 cm (Symmetrical Tip, Heparin Coating, Silver Ion Sleeve and Tal VenaTrac Stylet- Intended for acute a
  • Z-1635-2024 · Mar 15, 2024 · Open, Classified
    Palindrome Precision H Chronic Catheter Kit 14.5 Fr/Ch (4.8mm) x 23 cm (Symmetrical Tip, Heparin Coating and Tal VenaTrac Stylet- Intended for acute and chronic hemodialysis, aph
  • Z-1313-2022 · Jun 8, 2022 · Open, Classified
    Palindrome Precision H Chronic Catheter, Sterile, Single Use, Rx Only. For vascular access for acute and chronic hemodialysis, apheresis, and infusion.
  • Z-1315-2022 · Jun 8, 2022 · Open, Classified
    Palindrome Precision HSI Chronic Catheter, Sterile, Single Use, Rx Only. For acute and chronic hemodialysis, apheresis, and infusion.
  • Z-1314-2022 · Jun 8, 2022 · Open, Classified
    Palindrome Precision SI Chronic Catheter, Sterile, Single Use, Rx Only. For vascular access for acute and chronic hemodialysis, apheresis, and infusion.
  • Z-2373-2021 · May 28, 2021 · Open, Classified
    Palindrome Precision RT Chronic Catheter Kit Symmetrical Tip, Reverse-Tunneled 15 Fr/Ch (5.0 mm) x 33 cm, Item Code 8888541033P
  • Z-2372-2021 · May 28, 2021 · Open, Classified
    Palindrome Precision RT Chronic Catheter Kit Symmetrical Tip, Reverse-Tunneled 15 Fr/Ch (5.0 mm) x 28 cm, Item Code 8888541028P
  • Z-2374-2021 · May 28, 2021 · Open, Classified
    Palindrome Precision RT Chronic Catheter Kit Symmetrical Tip, Reverse-Tunneled 15 Fr/Ch (5.0 mm) x 44 cm, Item Code 8888541044P
  • Z-2371-2021 · May 28, 2021 · Open, Classified
    Palindrome Precision RT Chronic Catheter Kit Symmetrical Tip, Reverse-Tunneled 15 Fr/Ch (5.0 mm) x 23 cm, Item Code 8888541023P
  • Z-2375-2021 · May 28, 2021 · Open, Classified
    Palindrome Precision RT Chronic Catheter Kit Symmetrical Tip, Reverse-Tunneled 15 Fr/Ch (5.0 mm) x 55 cm, Item Code 8888541055P

…and 33 more.

Other Catheter, hemodialysis, implanted devices

All Catheter, hemodialysis, implanted devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page