Skip to content

Kangaroo 388103Recalls & reported problems

Cardinal Health, Inc. · Interface Cable For use with Kangaroo Feeding Tubes with IRIS Technology · Class 2

All models of Kangaroo (145)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10884521165014), received between Dec 18, 2019 and Oct 25, 2023 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “No Display/Image” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • No Display/Image2
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Consequences Or Impact To Patient2
  • Pneumothorax1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 25, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    CUSTOMER REPORTS: A SWEDISH ICU EXPERIENCED AN INTERFACE CABLE FAILURE DURING A PLACEMENT WHICH UNFORTUNATELY RESULTED IN THE NURSE CAUSING A PNEUMOTHORAX. THE SAFETY TEAM IS PERFORMING AN INVESTIGATION TO CONFIRM WHO'S AT FAULT AND IF THE INTERFACE CABLE HAS BEEN IDENTIFIED AS THE CAUSE. PER ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER ON 10/24/23, THE CLINICAL ATTEMPTED TO USE THE IRIS FEEDING TUBE, HOWEVER PRIOR TO PLACEMENT, THERE WAS NO SIGNAL ON THE CAMERA AND THE DEVICE COULD NOT BE USED. THE CUSTOMER STATED THEY BELIEVE IT WAS AN ISSUE WITH THE INTERFACE CABLE. THE CLINICIAN USED ANOTHER FEEDING TUBE AND PLACED THE DEVICE BLINDLY WHICH CAUSED THE PNEUMOTHORAX. THE CLINICIAN DID NOT KNOW IF THE FEEDING TUBE USED WAS A CARDINAL HEALTH DEVICE.

    Patient problems: Pneumothorax

    MAUDE report 18006979 ↗
  • MalfunctionDec 18, 2019No Display/Image; No Display/Image

    THE CUSTOMER REPORTED THAT THE SCREEN TURNED BLACK DURING PLACEMENT. THE HEALTHCARE PROVIDER REPLACED THE CABLE AND THERE WAS NO HARM TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9490474 ↗
  • MalfunctionDec 18, 2019No Display/Image; No Display/Image

    THE CUSTOMER REPORTED THAT THE SCREEN TURNED BLACK DURING PLACEMENT OF THE TUBE. THE HEALTHCARE PROVIDER REPLACED THE CABLE. THERE WAS NO PATIENT HARM.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9490417 ↗

Other TUBES, GASTROINTESTINAL (AND ACCESSORIES) devices

All TUBES, GASTROINTESTINAL (AND ACCESSORIES) devices →

Sources and updates