ProGrip LPG1614Recalls & reported problems
Sofradim Production · Laparoscopic Self-Fixating Mesh Polyester with Polylactic Acid Grips and Collagen Film · Class 2 · Implant
All models of ProGrip (24)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10884521513242), received between Nov 4, 2019 and Feb 8, 2024 — classified as 2 injuries, 10 malfunctions.
The most frequently reported specific device problem is “Difficult to Insert” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Difficult to Insert7
- Loss of or Failure to Bond4
- Difficult to Remove3
- Adverse Event Without Identified Device or Use Problem1
- Material Split, Cut or Torn1
- Product Quality Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
- No Code Available1
- Pain1
- Tissue Breakdown1
- Unintended Radiation Exposure1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 8, 2024Difficult to Insert; Product Quality Problem; Difficult to Insert; Product Quality Problem
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC TOTALLY EXTRA-PERITONEAL (TEP) UNILATERAL INGUINAL HERNIA PROCEDURE, WHEN THE SURGEON WAS UNFOLDING THE MESH TO PLACE AND FIXATE IT INTO THE TISSUE, THE SURGEON REALIZED IT WAS INCORRECTLY FOLDED. THE SURGEON EXTRACTED THE MESH OUT OF THE PATIENT THROUGH THE PORT, AND THE MESH GOT STUCK IN THE SEAL, AND THE SEAL DISLODGED. MOREOVER, THERE WAS A RAPID LOSS OF ABDOMINAL PRESSURE. IT WAS NOTED THAT THE MESH WAS ABLE TO BE PULLED OUT WITH THE SEAL ATTACHED TO IT. THE SURGEON REUSED THE SAME PORT WITHOUT THE SEAL, AND THE SURGICAL ASSISTANT USED THE CAMERA AND THEIR HAND WITH A SURGICAL PACK TO COVER THE SEAL AND TO STOP GAS FROM LEAKING. TO RESOLVE THE ISSUE, A NEW PORT WAS OPENED, BUT IT WAS NOT USED, AND USED A NEW PIECE OF MESH IN THE PATIENT IN ORDER TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18667418 ↗MalfunctionJan 10, 2024Difficult to Insert; Difficult to Insert
ACCORDING TO THE REPORTER, DURING A BILATERAL TOTAL EXTRAPERITONEAL (TEP) LAPAROSCOPIC INGUINAL HERNIA REPAIR, THE FIRST MESH WAS CUT, FOLDED, AND PLACED DOWN THE STRUCTURAL BALLOON PORT WITH NO ISSUE. THE NEXT MESH WAS CUT AND FOLDED THE SAME WAY, PLACED DOWN THE STRUCTURAL PORT, AND BECAME STUCK IN THE MIDDLE OF THE PORT. THE SURGEON FORCED IT DOWN, AND IT GOT CAUGHT IN THE SEAL AND TEARED IT OFF. THE SEAL OF THE STRUCTURAL BALLOON FELL INTO THE PATIENT'S CAVITY AND WAS ABLE TO BE RETRIEVED. A NEW STRUCTURAL BALLOON AND MESH WERE USED WITH THE SAME LOT FOR BOTH DEVICES TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18488810 ↗MalfunctionAug 18, 2023Difficult to Remove; Difficult to Remove
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC RIGHT INGUINAL HERNIA REPAIR USING TOTALLY EXTRAPERITONEAL (TEP) TECHNIQUE, AFTER THE DISSECTION WAS COMPLETED AND THE MESH WAS READY TO BE PLACED, WHEN THE SURGEON TRIED TO PLACE THE MESH DOWN THE PORT AND THE VALVE SEAL DISENGAGED AND WAS PUSHED INTO THE PORT, IT BECOME STUCK INSIDE THE PORT WITH THE MESH AND WAS UNABLE TO BE PUSHED DOWN OR TAKEN OUT. THE SURGEON USED A NEW DEVICE AND A NEW PIECE OF MESH AND THE ISSUE DID NOT RE-OCCUR. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17576977 ↗MalfunctionAug 15, 2023Loss of or Failure to Bond; Difficult to Insert; Loss of or Failure to Bond; Difficult to Insert
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC TOTAL EXTRAPERITONEAL HERNIA REPAIR (TEP), THE GREY SEAL, FELL INTO THE CAVITY OF THE PATIENT. THE GREY SEAL GOT DETACHED FROM THE PORT AND BECAME STUCK HALFWAY ALONG THE MESH. ADDITIONALLY, THE MESH HAD BECOME ADHERED TO MUSCLE AND OTHER TISSUES NEAR THE PORT TIP. TO RESOLVE THE ISSUE, THE FAULTY PORT WAS REMOVED, AND THE MESH HAD TO BE FORCEFULLY PULLED OUT THROUGH THE PORT INCISION. THE GREY SEAL WAS ALSO REMOVED FROM THE MESH. AS A RESULT OF THIS PROBLEM, THERE WAS TISSUE DAMAGE TO THE RECTUS ABDOMINIS MUSCLE AND OTHER TISSUES SURROUNDING THE PORT SITE. THE DAMAGE OCCURRED WHILE FORCIBLY REMOVING THE MESH FROM THE PORT SITE. TO RESOLVE THE ISSUE AND COMPLETE THE CASE, ANOTHER TROCAR WAS OPENED. MOREOVER, THE DEVICE ISSUE CAUSED A RAPID LOSS OF ABDOMINAL PRESSURE.
Patient problems: Tissue Breakdown
MAUDE report 17553610 ↗MalfunctionAug 15, 2023Loss of or Failure to Bond; Difficult to Insert; Loss of or Failure to Bond; Difficult to Insert
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC TOTALLY EXTRAPERITONEAL (TEP) BILATERAL INGUINAL HERNIA REPAIR, AS THE SECOND PIECE OF MESH WAS BEING PASSED THROUGH THE PORT, THE GREY SEAL CAME COMPLETELY OFF. THE SURGEON ATTEMPTED TO REPAIR THE PORT, BUT THE SEAL WOULD NOT REMAIN IN PLACE AND INSUFFLATION GAS CONTINUED TO LEAK. DUE TO RAPID LOSS OF ABDOMINAL PRESSURE, THE MESH USED BECAME ATTACHED TO UNINTENDED TISSUE. PROCEDURE TIME WAS EXTENDED (BY LESS THAN 30 MINUTES), WHILE THE SURGEON REMOVED THE MESH FROM THESE TISSUES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17551025 ↗MalfunctionAug 11, 2023Loss of or Failure to Bond; Difficult to Insert; Loss of or Failure to Bond; Difficult to Insert
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC TOTALLY EXTRAPERITONEAL (TEP) BILATERAL INGUINAL HERNIA REPAIR, AS THE SECOND PIECE OF MESH WAS BEING PASSED THROUGH THE PORT, THE GREY SEAL CAME COMPLETELY OFF. THE SURGEON ATTEMPTED TO REPAIR THE PORT, BUT THE SEAL WOULD NOT REMAIN IN PLACE AND INSUFFLATION GAS CONTINUED TO LEAK. DUE TO RAPID LOSS OF ABDOMINAL PRESSURE, THE MESH USED BECAME ATTACHED TO UNINTENDED TISSUE. PROCEDURE TIME WAS EXTENDED (BY LESS THAN 30 MINUTES), WHILE THE SURGEON REMOVED THE MESH FROM THESE TISSUES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17514075 ↗MalfunctionAug 7, 2023Loss of or Failure to Bond; Difficult to Insert; Loss of or Failure to Bond; Difficult to Insert
ACCORDING TO THE REPORTER, IN A LAPAROSCOPIC INGUINAL HERNIA REPAIR, AS THE MESH WAS BEING INTRODUCED THROUGH THE TROCAR, THE VALVE SEAL COMPLETELY CAME OFF. THE TROCAR WAS ATTEMPTED TO BE FIXED. THE SURGEON TRIED RE-FITTING THE OUTER GREY SEAL. DUE TO RAPID AND CONTINUED LOSS OF ABDOMINAL PRESSURE, THE BALLOON WAS DEFLATED, REMOVED, AND THE DEVICE WAS REPLACED. THE MESH WHICH WAS BEING USED UNINTENDEDLY GOT STUCK TO THE TISSUE. THE LENGTH OF THE PROCEDURE INCREASED WHILE THE SURGEON DETACHED THE MESH FROM THIS UNINTENDED TISSUE. THE MESH WAS STILL ABLE TO BE USED, AND WAS SUCCESSFULLY IMPLANTED. NO DEVICE COMPONENT FELL INTO THE PATIENT¿S CAVITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17477126 ↗MalfunctionAug 1, 2023Difficult to Remove; Difficult to Remove
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC INGUINAL HERNIA REPAIR, AN ISSUE OCCURRED AS THE MESH WAS BEING INSERTED THROUGH THE TROCAR, CAUSING THE OUTER GREY SEAL TO COMPLETELY COME OFF. THE SURGEON MANAGED TO REATTACH THE GREY SEAL USING FORCEPS. HOWEVER, THE ABDOMINAL PRESSURE COULD NOT BE MAINTAINED. THE MESH USED BECAME ATTACHED TO UNINTENDED TISSUE DUE TO THE LACK OF ABDOMINAL PRESSURE. THERE WAS A RAPID LOSS OF ABDOMINAL PRESSURE RESULTED FROM THE DEVICE ISSUE. TO FULLY RESOLVE THE ISSUE, THE SURGEON NEEDED TO SPEND ADDITIONAL TIME REMOVING THE MESH FROM THIS UNINTENDED TISSUE, LEADING TO A MODERATE INCREASE IN THE PROCEDURE TIME, NOT EXCEEDING 30 MINUTES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17437972 ↗MalfunctionJul 31, 2023Difficult to Remove; Difficult to Remove
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC INGUINAL HERNIA REPAIR, AN ISSUE OCCURRED AS THE MESH WAS BEING INSERTED THROUGH THE TROCAR, CAUSING THE OUTER GREY SEAL TO COMPLETELY COME OFF. THE SURGEON MANAGED TO REATTACH THE GREY SEAL USING FORCEPS. HOWEVER, THE ABDOMINAL PRESSURE COULD NOT BE MAINTAINED. THE MESH USED BECAME ATTACHED TO UNINTENDED TISSUE DUE TO THE LACK OF ABDOMINAL PRESSURE. THERE WAS A RAPID LOSS OF ABDOMINAL PRESSURE RESULTED FROM THE DEVICE ISSUE. TO FULLY RESOLVE THE ISSUE, THE SURGEON NEEDED TO SPEND ADDITIONAL TIME REMOVING THE MESH FROM THIS UNINTENDED TISSUE, LEADING TO A MODERATE INCREASE IN THE PROCEDURE TIME, NOT EXCEEDING 30 MINUTES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17430445 ↗MalfunctionMay 10, 2023Difficult to Insert; Difficult to Insert
ACCORDING TO THE REPORTER, DURING LAPAROSCOPIC INGUINAL HERNIA, THE INSERTION OF MESH CAUSED THE VALVE TO COME UNSTUCK AND BE PUSHED DOWN THE TROCAR. NO COMPONENT FELL INTO THE CAVITY. THE SURGEON TRIED TO RETRIEVE VALVE TO CONTINUE, BUT NEEDED TO REPLACE ENTIRE STRUCTURAL BALLOON PORT AS USER WAS UNABLE TO CONTINUE DUE TO THE DISLODGED VALVE BLOCKING THE TROCAR. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16904498 ↗InjuryAug 12, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, AFTER FIXING THE HERNIA DEFECT IN A LAPAROSCOPIC INGUINAL HERNIA REPAIR, PATIENT WAS COMPLAINING ABOUT THE CHRONIC PAIN WHERE MESH WAS IMPLANTED. THE PATIENT IS BEEN SENT TO DO SCANNING TO LOCATE THE RIGHT REASON FOR PAIN IN INGUINAL AREA.
Patient problems: Pain; Unintended Radiation Exposure
MAUDE report 12316115 ↗InjuryNov 4, 2019Material Split, Cut or Torn; Material Split, Cut or Torn
ACCORDING TO THE REPORTER, POST LAPAROSCOPIC BILATERAL INGUINAL REPAIR, THE PATIENT RETURNED FOR A ROBOTIC INGUINAL HERNIA REPAIR FOR REOCCURRENCE. THE MESH IMPLANTS WERE REMAIN INSTITUTE AND SHOW GOOD INTEGRATION EXCEPT FOR THE HOLE IN THE RIGHT SIDED MESH. HERNIA DEFECT REPAIR WAS DONE TO RESOLVE THE ISSUE.
Patient problems: No Code Available
MAUDE report 9274639 ↗
FDA clearance and classification
The FDA regulates product code FTL (mesh, surgical, polymeric) under 21 CFR 878.3300 as a Class II device. 21 CFR 878.3300 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the ProGrip LPG1614
Has the ProGrip LPG1614 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10884521513242), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the ProGrip LPG1614?
The FDA's MAUDE database lists 12 adverse event reports for this device, received between Nov 4, 2019 and Feb 8, 2024. In those reports, the most common specific device problems are “Difficult to Insert” (7), “Loss of or Failure to Bond” (4) and “Difficult to Remove” (3); the most common patient problems are “Pain” (1), “Tissue Breakdown” (1) and “Unintended Radiation Exposure” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ProGrip LPG1614?
None of the reports are classified as deaths, 2 as injuries and 10 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the ProGrip LPG1614 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Safe: it poses no known hazards in any MRI environment.
Who makes the ProGrip LPG1614, and is it still sold?
The FDA's device database lists Sofradim Production as the company responsible for it (brand ProGrip, model LPG1614). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Mesh, surgical, polymeric devices
- PROCEED PVPMJohnson & Johnson International Inc. · Mesh, surgical, polymeric1,233 reports
- PROCEED PCDN1Johnson & Johnson International Inc. · Mesh, surgical, polymeric744 reports
- PROCEED PCDG1Johnson & Johnson International Inc. · Mesh, surgical, polymeric1 recall1 death637 reports
- PROCEED PVPSJohnson & Johnson International Inc. · Mesh, surgical, polymeric622 reports
- ProGrip LPG1510Sofradim Production · Mesh, surgical, polymeric2 deaths616 reports
- VENTRALEX 0010302Davol Inc. · Mesh, surgical, polymeric4 deaths530 reports
- BARD 0112680Davol Inc. · Mesh, surgical, polymeric1 death493 reports
- ProGrip TEM1509GSofradim Production · Mesh, surgical, polymeric2 deaths470 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page