Covidien 8884715122ERecalls & reported problems

Cardinal Health, Inc. · True content to be populated as part of rebranding · Class 2 · Implant

All models of Covidien (7)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 10884521661622), received between Jun 11, 2021 and Oct 15, 2025 — classified as 15 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak11
  • Break3
  • Material Rupture2
  • Unintended Deflation2
  • Deflation Problem1
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions13
  • Insufficient Information1
  • Unintended Extubation1

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 15, 2025Unintended Deflation; Unintended Deflation

    THE CUSTOMER REPORTED COMPLETE DISPLACEMENT OF THE GASTROSTOMY TUBE WAS EVIDENT. UPON DISCHARGE, THE INTERNAL RETENTION BALLOON WAS OBSERVED TO BE DEFLATED, WHICH COULD HAVE FACILITATED ITS ACCIDENTAL DISLODGEMENT. THE PATIENT WAS TAKEN TO SURGERY WITH APPROPRIATE MANAGEMENT AND REMAINS IN THE ICU.

    Patient problems: Insufficient Information

    MAUDE report 23294551 ↗
  • MalfunctionFeb 14, 2024Unintended Deflation; Unintended Deflation

    THE CUSTOMER REPORTED WHEN MANIPULATING THE PATIENT, THE GASTROSTOMY TUBE WAS DISPLACED. THE BALLOON WAS TESTED WITH WATER, AND IT WAS FOUND TO BE DYSFUNCTIONAL. A FOLEY CATHETER WAS PASSED WITHOUT RESISTANCE, AND THE BALLOON WAS INSUFFLATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18702869 ↗
  • MalfunctionJul 18, 2023Fluid/Blood Leak; Fluid/Blood Leak

    THE CUSTOMER REPORTED EVIDENCE AT THE TIME OF FEEDING THAT ONE PARTITION GASTROTOMY PORT WHERE THE FEEDING EXITS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17347428 ↗
  • MalfunctionMar 10, 2023Deflation Problem; Detachment of Device or Device Component; Deflation Problem; Detachment of Device or Device Component

    THE CUSTOMER REPORTED THAT THE 12 FR KANGAROO G-TUBE HAD A BROKEN MED PORT AND BALLOON. THE TUBE WAS INSERTED ON (B)(6) 2023 AND IT WAS BROKEN ON (B)(6) 2023. THERE WAS NO HARM REPORTED. ADDITIONAL INFORMATION WAS RECEIVED AND STATED THAT THE Y-PORT WAS DETACHED FROM THE TUBE. PER CUSTOMER, THERE WAS NO WATER IN THE BALLOON AT THE TIME OF THE CHANGE AND WHEN THE TUBE WAS REMOVED. THE CUSTOMER IS NOT SURE IF THE BALLOON WAS DEFLATED WITH A PIN HOLE OR LEAK. UNFORTUNATELY, THEY DID NOT LOOK AT IT THAT CLOSELY TO INVESTIGATE IF IT WAS RUPTURED OR LEAKING. THERE WERE NO FRAGMENTED PIECES LEFT INSIDE THE PATIENT INVOLVED. THERE WAS NO HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16526384 ↗
  • MalfunctionMar 6, 2023Break; Fluid/Blood Leak; Break; Fluid/Blood Leak

    THE CHARGE TECHNOLOGIST FOR INTERVENTIONAL RADIOLOGY REPORTED THAT ONE OF THE TECHNICIANS SAID THE ENFIT END BROKE BELOW THE ACTUAL CONNECTION ON THE KANGAROO FEEDING TUBE FOR ONE OF THE INPATIENTS. ADDITIONAL INFORMATION RECEIVED STATED THE DEVICE HAD A BREAK AT THE FEED AND MEDICATION PORT CAUSING A LEAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16485626 ↗
  • MalfunctionMar 6, 2023Break; Fluid/Blood Leak; Break; Fluid/Blood Leak

    THE CHARGE TECHNOLOGIST FOR INTERVENTIONAL RADIOLOGY REPORTED THE ENFIT END IS BREAKING BELOW THE ACTUAL CONNECTION ON THE KANGAROO FEEDING TUBES. ADDITIONAL INFORMATION RECEIVED STATED THE DEVICE HAD A BREAK AT THE FEED AND MEDICATION PORT CAUSING A LEAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16485472 ↗
  • MalfunctionMar 6, 2023Break; Fluid/Blood Leak; Break; Fluid/Blood Leak

    THE CHARGE TECHNOLOGIST FOR INTERVENTIONAL RADIOLOGY REPORTED THE ENFIT END IS BREAKING BELOW THE ACTUAL CONNECTION ON THE KANGAROO FEEDING TUBES. ADDITIONAL INFORMATION RECEIVED STATED THE DEVICE HAD A BREAK AT THE FEED AND MEDICATION PORT CAUSING A LEAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16485285 ↗
  • MalfunctionJan 31, 2023Material Rupture; Material Rupture

    THE CUSTOMER REPORTED THAT THE BALLOON RUPTURED AND FELL OUT.

    Patient problems: Unintended Extubation

    MAUDE report 16271979 ↗
  • MalfunctionNov 14, 2022Fluid/Blood Leak; Material Rupture; Fluid/Blood Leak; Material Rupture

    THE CUSTOMER REPORTED THAT THE TUBE HAD A LEAK AND THEY VERIFIED IT VIA THE BREAKAGE OF THE UNION OF THE KEY. THE PATIENT WAS A MALE UNDER THE AGE OF ONE. HE WAS IN THE HOSPITAL WITH A MEDICAL DIAGNOSIS OF MIXED CHRONIC MALNUTRITION, NEUROGENIC DYSPHAGIA AND DILATATED HEART DISEASE. PER ADDITIONAL INFORMATION RECEIVED, THERE WAS NO MEDICAL INTERVENTION REQUIRED. THERE WAS NO PATIENT INJURY REPORTED. "BREAKAGE OF THE UNION OF THE KEY" REFERS TO THE TUBE RUPTURING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15792617 ↗
  • MalfunctionNov 24, 2021Fluid/Blood Leak; Fluid/Blood Leak

    THE CUSTOMER REPORTED A CRACK BETWEEN THE Y OF THE CONNECTOR CAUSING FORMULA TO LEAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12869231 ↗
  • MalfunctionJun 11, 2021Fluid/Blood Leak; Fluid/Blood Leak

    THE CUSTOMER REPORTED THAT IN THE ENDOSCOPY DEPARTMENT, THE HOSPITAL RECENTLY SWITCHED TO A NEW SYSTEM FOR PEG PROBES. THEY NOW USE THE KANGAROO GASTROSTOMY FEEDING TUBES WITH ENFIT Y-PORTS. THE ENDOSCOPIC DEPARTMENT STATED THAT THE PATIENT RETURNED WITH A DEFECTIVE CONNECTION. THE NURSE HAD TO TAKE A NEW PRODUCT AND TAKE THE CONNECTOR FROM THE NEW PRODUCT TO SWITCH IT WITH THE DAMAGED CONNECTOR. A REPRESENTATIVE PHOTO WAS PROVIDED AND SHOWED THAT THE Y-PORT CRACKED IN THE CENTER. THERE WAS NO PATIENT INJURY. ADDITIONAL INFORMATION PROVIDED ON JUNE 04, 2021 STATED THAT THE DEVICE DID LEAK BECAUSE OF THE Y-PORT BEING TORN. AFTER ATTACHING THE FEEDING TUBE ON THE ENFIT PORT, THEY WERE NOT ABLE TO UNSCREW THE FEEDING TUBE FROM THE ENFIT PORT, THEREFOR THE Y-PORT TORE AND LEAKED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11982247 ↗
  • MalfunctionJun 11, 2021Fluid/Blood Leak; Fluid/Blood Leak

    THE CUSTOMER REPORTED THAT IN THE ENDOSCOPY DEPARTMENT, THE HOSPITAL RECENTLY SWITCHED TO A NEW SYSTEM FOR PEG PROBES. THEY NOW USE THE KANGAROO GASTROSTOMY FEEDING TUBES WITH ENFIT Y-PORTS. THE ENDOSCOPIC DEPARTMENT STATED THAT THE PATIENT RETURNED WITH A DEFECTIVE CONNECTION. THE NURSE HAD TO TAKE A NEW PRODUCT AND TAKE THE CONNECTOR FROM THE NEW PRODUCT TO SWITCH IT WITH THE DAMAGED CONNECTOR. A REPRESENTATIVE PHOTO WAS PROVIDED AND SHOWED THAT THE Y-PORT CRACKED IN THE CENTER. THERE WAS NO PATIENT INJURY. ADDITIONAL INFORMATION PROVIDED ON (B)(6) 2021 STATED THAT THE DEVICE DID LEAK BECAUSE OF THE Y-PORT BEING TORN. AFTER ATTACHING THE FEEDING TUBE ON THE ENFIT PORT, THEY WERE NOT ABLE TO UNSCREW THE FEEDING TUBE FROM THE ENFIT PORT, THEREFOR THE Y-PORT TORE AND LEAKED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11982240 ↗
  • MalfunctionJun 11, 2021Fluid/Blood Leak; Fluid/Blood Leak

    THE CUSTOMER REPORTED THAT IN THE ENDOSCOPY DEPARTMENT, THE HOSPITAL RECENTLY SWITCHED TO A NEW SYSTEM FOR PEG PROBES. THEY NOW USE THE KANGAROO GASTROSTOMY FEEDING TUBES WITH ENFIT Y-PORTS. THE ENDOSCOPIC DEPARTMENT STATED THAT THE PATIENT RETURNED WITH A DEFECTIVE CONNECTION. THE NURSE HAD TO TAKE A NEW PRODUCT AND TAKE THE CONNECTOR FROM THE NEW PRODUCT TO SWITCH IT WITH THE DAMAGED CONNECTOR. A REPRESENTATIVE PHOTO WAS PROVIDED AND SHOWED THAT THE Y-PORT CRACKED IN THE CENTER. THERE WAS NO PATIENT INJURY. ADDITIONAL INFORMATION PROVIDED ON JUNE 04, 2021 STATED THAT THE DEVICE DID LEAK BECAUSE OF THE Y-PORT BEING TORN. AFTER ATTACHING THE FEEDING TUBE ON THE ENFIT PORT, THEY WERE NOT ABLE TO UNSCREW THE FEEDING TUBE FROM THE ENFIT PORT, THEREFOR THE Y-PORT TORE AND LEAKED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11982230 ↗
  • MalfunctionJun 11, 2021Fluid/Blood Leak; Fluid/Blood Leak

    THE CUSTOMER REPORTED THAT IN THE ENDOSCOPY DEPARTMENT, THE HOSPITAL RECENTLY SWITCHED TO A NEW SYSTEM FOR PEG PROBES. THEY NOW USE THE KANGAROO GASTROSTOMY FEEDING TUBES WITH ENFIT Y-PORTS. THE ENDOSCOPIC DEPARTMENT STATED THAT THE PATIENT RETURNED WITH A DEFECTIVE CONNECTION. THE NURSE HAD TO TAKE A NEW PRODUCT AND TAKE THE CONNECTOR FROM THE NEW PRODUCT TO SWITCH IT WITH THE DAMAGED CONNECTOR. A REPRESENTATIVE PHOTO WAS PROVIDED AND SHOWED THAT THE Y-PORT CRACKED IN THE CENTER. THERE WAS NO PATIENT INJURY. ADDITIONAL INFORMATION PROVIDED ON JUNE 04, 2021 STATED THAT THE DEVICE DID LEAK BECAUSE OF THE Y-PORT BEING TORN. AFTER ATTACHING THE FEEDING TUBE ON THE ENFIT PORT, THEY WERE NOT ABLE TO UNSCREW THE FEEDING TUBE FROM THE ENFIT PORT, THEREFOR THE Y-PORT TORE AND LEAKED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11982228 ↗
  • MalfunctionJun 11, 2021Fluid/Blood Leak; Fluid/Blood Leak

    THE CUSTOMER REPORTED THAT IN THE ENDOSCOPY DEPARTMENT, THE HOSPITAL RECENTLY SWITCHED TO A NEW SYSTEM FOR PEG PROBES. THEY NOW USE THE KANGAROO GASTROSTOMY FEEDING TUBES WITH ENFIT Y-PORTS. THE ENDOSCOPIC DEPARTMENT STATED THAT THE PATIENT RETURNED WITH A DEFECTIVE CONNECTION. THE NURSE HAD TO TAKE A NEW PRODUCT AND TAKE THE CONNECTOR FROM THE NEW PRODUCT TO SWITCH IT WITH THE DAMAGED CONNECTOR. A REPRESENTATIVE PHOTO WAS PROVIDED AND SHOWED THAT THE Y-PORT CRACKED IN THE CENTER. THERE WAS NO PATIENT INJURY. ADDITIONAL INFORMATION PROVIDED ON (B)(6) 2021 STATED THAT THE DEVICE DID LEAK BECAUSE OF THE Y-PORT BEING TORN. AFTER ATTACHING THE FEEDING TUBE ON THE ENFIT PORT, THEY WERE NOT ABLE TO UNSCREW THE FEEDING TUBE FROM THE ENFIT PORT, THEREFOR THE Y-PORT TORE AND LEAKED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11982216 ↗

Other Tubes, gastrointestinal (and accessories) devices

All Tubes, gastrointestinal (and accessories) devices →