Maxon 8886621731-2Recalls & reported problems
Covidien LP · Monofilament Absorbable Suture · Class 2 · Implant
All models of Maxon (114)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10884521757820), received between Dec 9, 2022 and May 27, 2026 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (6 reports).
1 other recall cover devices cleared under the same FDA submission (K990951); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component6
- Break1
- Component Missing1
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Insufficient Information1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 27, 2026Component Missing
IT WAS REPORTED THAT DURING A PROCEDURE, THE NEEDLE WAS NOT ATTACHED. IT WAS NOT INCLUDED IN THE PACKET. THE DEVICE WAS NOT USED. TO RESOLVE THE ISSUE, A NEW DEVICE WAS USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25299538 ↗MalfunctionJan 3, 2026Detachment of Device or Device Component
ACCORDING TO THE REPORTER, PREOPERATIVELY OF A VETERINARY CASE, THE TWO NEEDLES DETACHED WHEN PULLING OUT FROM THE RETAINER AND PREPARING FOR SUTURING. A NEW SUTURE USED WITHOUT ANY PROBLEMS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23952855 ↗MalfunctionJan 3, 2026Detachment of Device or Device Component
ACCORDING TO THE REPORTER, PREOPERATIVELY OF A VETERINARY CASE, THE TWO NEEDLES DETACHED WHEN PULLING OUT FROM THE RETAINER AND PREPARING FOR SUTURING. A NEW SUTURE USED WITHOUT ANY PROBLEMS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23952839 ↗MalfunctionApr 24, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER DURING A PROCEDURE, WHILE PERFORMING A CONTINUOUS SUTURING TECHNIQUE ON THE SUBCUTANEOUS DERMAL, THE NEEDLE AND THREAD DETACHED. THE SURGEON HANDLED IT GENTLY. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21901604 ↗MalfunctionJan 24, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER DURING A ORTHOPEDIC PROCEDURE, THE NEEDLE WAS DETACHED, WHEN IT WAS BEING PULLED OUT FROM THE RETAINER AND PREPARING FOR SUTURING (BEFORE PATIENT USE). THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21232112 ↗MalfunctionJan 24, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER DURING A ORTHOPEDIC PROCEDURE, THE NEEDLE WAS DETACHED, WHEN IT WAS BEING PULLED OUT FROM THE RETAINER AND PREPARING FOR SUTURING (BEFORE PATIENT USE). THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21231628 ↗MalfunctionMar 16, 2023Break; Break
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC APPENDECTOMY PROCEDURE, THE THREAD WAS CUT BY THE FORCE AFTER INTRACORPOREAL LIGATION, WHICH CANNOT BE CUT DURING NORMAL USE. A DIFFERENT PRODUCT WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16553316 ↗MalfunctionDec 9, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, INTRA-OPERATIVELY, BEFORE SUTURING, THE NEEDLE AND SUTURE WERE DISCONNECTED AFTER A FEW MINUTES WHEN THE PRODUCT WAS OPENED.
Patient problems: Insufficient Information
MAUDE report 15950266 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K990951. One submission often covers many models, so check the product and lot details in each record.
- Z-2542-2017 · May 3, 2017 · TerminatedSutures. MAXON: MAXON 4/0 5X18" GREEN CV-24 D-TACH; Item Number: 8886621933; MAXON 5/0 5X18 GR CV-23 D-TACH; Item Number: XX2109; MAXON* 4/0 5X30 GREEN V-20 D*TACH; Item Number:
Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices
- Coated VICRYL VCP433HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death1,093 reports
- Polysorb CL-915Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID503 reports
- Coated VICRYL VCP493HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID435 reports
- V-Loc 180 VLOCL0316Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall376 reports
- Coated VICRYL VCP345HEthicon, LLC · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 death405 reports
- V-Loc 180 VLOCL0604Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall328 reports
- V-Loc 180 VLOCA008LCovidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID1 recall294 reports
- V-Loc 180 VLOCL0326Covidien LP · SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID330 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026