Skip to content

BIS PMB4000ACBLRecalls & reported problems

Covidien LP · BIS (TM) Advance Adapter Cable;Also for use with HealthCast (TM) insight patient monitor PM5000;Reusable cable · Class 2

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10884521808317), received between Jun 27, 2025 and Jun 27, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Temperature Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Temperature Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 27, 2025Temperature Problem; Temperature Problem

    IT WAS REPORTED THAT THE DEVICE WAS REPORTED TO HAVE DISCHARGED DUE TO BEING UNPLUGGED. AN ATTEMPT TO INSERT THE POWER ADAPTER INTO THE DOCKING STATION FOR CHARGING REVEALED THAT THE ADAPTER WOULD NOT STAY SECURELY IN PLACE. UPON INSPECTING THE DOCKING STATION, IT WAS DISCOVERED THAT A MINOR INCIDENT, CABLE BURN, HAD OCCURRED INSIDE THE DOCKING STATION. TO ADDRESS THIS, THE POWER CABLE WAS SECURED WITH CABLE TIES TO MINIMIZE STRAIN AT THE POINT OF INSERTION INTO THE DOCKING STATION. ADDITIONALLY, A 12V POWER SUPPLY UNIT WAS IMPLEMENTED BETWEEN THE 230V POWER SOURCE AND THE DOCKING STATION TO ACT AS A SAFEGUARD. IN THE EVENT OF A POWER SURGE, THIS UNIT WOULD BE MORE LIKELY TO SUSTAIN DAMAGE RATHER THAN THE DOCKING STATION ITSELF. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22347981 ↗

Other Index-generating electroencephalograph software devices

All Index-generating electroencephalograph software devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.