Mon-a-Therm 90044 (UDI 10884521828384)Recalls & reported problems
Covidien LP · General Purpose Temperature Probe 400TM · Class 2
All models of Mon-a-Therm (12)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10884521828384), received between Apr 26, 2024 and Nov 3, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Incorrect Measurement” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Apr 23, 2026Class IIOpen, Classified
Temperature probe devices lack FDA clearance.
Z-2272-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Incorrect Measurement2
- Low Readings1
Patient problems
- Hypothermia1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 3, 2025Incorrect Measurement; Incorrect Measurement
IT WAS REPORTED THAT THE SENSOR DISPLAYED AN ABNORMAL BODY TEMPERATURE READING OF 33 DEGREES AT THE BEGINNING OF THE OPERATION. THERE WAS AN INABILITY TO MONITOR THE PATIENT¿S TEMPERATURE DURING THE OPERATION, AND POOR MANAGEMENT OF HYPOTHERMIA OCCURRED DUE TO THE LACK OF TEMPERATURE MONITORING. THE ISSUE WAS DESCRIBED AS RECURRING, WITH PREVIOUS MEASURED TEMPERATURES OF 0 OR 17 DEGREES. IT WAS NOT POSSIBLE TO CHANGE THE TEMPERATURE PROBE DURING THE OPERATION IF A MALFUNCTION WAS DETECTED.
Patient problems: Hypothermia
MAUDE report 23460369 ↗MalfunctionApr 26, 2024Incorrect Measurement; Low Readings; Incorrect Measurement; Low Readings
ACCORDING TO THE REPORTER, DURING USE, THE READING FROM THE TEMPERATURE PROBE WAS LOW COMPARED TO OTHER DEVICES' READINGS. THE UNIT GAVE 35.2 READING WHILE THE ORAL TEMPERATURE WAS 36.3. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19193856 ↗
FDA clearance and classification
Cleared through 510(k) K240077 on Aug 20, 2024 (decision: Substantially Equivalent). Submitted by Covidien as “Mon-a-Therm General Purpose Temperature Probe 400TM (90050, 90044)”. The FDA review took 223 days. FDA record ↗
Cleared through 510(k) K874514 on Jan 29, 1988 (decision: Substantially Equivalent). Submitted by Mallinckrodt Critical Care as “Hi-Lo Temp(R) General Purpose Temperature Probe”. The FDA review took 88 days. FDA record ↗
The FDA regulates product code FLL (thermometer, electronic, clinical) under 21 CFR 880.2910 as a Class II device. 21 CFR 880.2910 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the Mon-a-Therm 90044
Has the Mon-a-Therm 90044 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2272-2026 (Class II), began on Apr 23, 2026: Temperature probe devices lack FDA clearance. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Mon-a-Therm 90044?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Apr 26, 2024 and Nov 3, 2025. In those reports, the most common specific device problems are “Incorrect Measurement” (2) and “Low Readings” (1); the most common patient problem is “Hypothermia” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Mon-a-Therm 90044?
None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Mon-a-Therm 90044 FDA approved?
It was cleared through 510(k) K874514 on Jan 29, 1988; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the Mon-a-Therm 90044 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the Mon-a-Therm 90044, and is it still sold?
The FDA's device database lists Covidien LP as the company responsible for it (brand Mon-a-Therm, model 90044). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Thermometer, electronic, clinical devices
- Welch Allyn, Inc. 901054Welch Allyn, Inc. · Thermometer, electronic, clinical1 recall221 reports
- Welch Allyn, Inc. 901054Welch Allyn, Inc. · Thermometer, electronic, clinical2 recalls13 reports
- NA B105 VSP3.0 Patient MonitorGE Medical Systems Information Technologies, Inc. · Thermometer, electronic, clinical2 recalls0 reports
- Mon-a-Therm 90050Covidien LP · Thermometer, electronic, clinical1 recall3 reports
- CARESCAPE B450 MBA313GE Healthcare Finland Oy · Thermometer, electronic, clinical1 recall1 report
- Temperature Probes Esophageal/Rectal Temperature ProbePhilips Medical Systems Hsg · Thermometer, electronic, clinical1 recall1 report
- NA 989803162621Philips Medical Systems Hsg · Thermometer, electronic, clinical1 recall0 reports
- Temperature Probes Skin Surface Temperature ProbePhilips Medical Systems Hsg · Thermometer, electronic, clinical1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.