Skip to content

Vital Sync VITALSYNCSW03 (UDI 10884521844933)Recalls & reported problems

Covidien LP · VITAL SYNC V3 SOFTWARE · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10884521844933), received between Mar 7, 2024 and Mar 7, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Computer Software Problem” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 30, 2026Open, Classified

    Due to complaints and investigations stating that alarms from the primary patient bedside monitor was not alarming/being transmitted and received on the remote patient monitoring system.

    Z-1557-2026 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Computer Software Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 7, 2024Computer Software Problem; Computer Software Problem

    ACCORDING TO THE REPORTER, PRIOR TO USE, THE REMOTE SURVEILLANCE SYSTEM HAD NO ANY AUDIBLE ALARMS ON THE CNS. TROUBLESHOOTING WAS REMOTED INTO, LOOKED AT THE TILES FOR ROOMS. IT WAS COLOR RED AND ORANGE BUT THERE WAS NO AUDIBLE ALERT. CLICKED ON THE TILE AND SEE THAT THE ALARM WAS SILENCED. THE RESOLUTION WAS TO UPDATE THE OPERATING SYSTEM TO CORRECT THE ISSUE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18862687 ↗

Other TRANSMITTERS AND RECEIVERS, PHYSIOLOGICAL SIGNAL, RADIOFREQUENCY devices

All TRANSMITTERS AND RECEIVERS, PHYSIOLOGICAL SIGNAL, RADIOFREQUENCY devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.