Skip to content

Hugo RAS MRASI0006Recalls & reported problems

Covidien LP · Large Needle Driver · Class 2

All models of Hugo RAS (10)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10884521850972), received between Mar 25, 2026 and Mar 25, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Display or Visual Feedback Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Display or Visual Feedback Problem1
  • Intermittent Communication Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 25, 2026Display or Visual Feedback Problem; Intermittent Communication Failure

    IT WAS REPORTED THAT DURING A ROBOTIC LAPAROSCOPIC PROSTATECTOMY, THE MONOPOLAR CURVED SHEARS (MCS) WERE NOT PERFORMING LIKE IT SHOULD DURING BLADDER TAKEDOWN, SPECIFICALLY WITH COLD CUTTING. THE PHYSICIAN WAS NOT SATISFIED WITH THE PERFORMANCE OF THE MCS. THE MCS WERE SWAPPED WITH ANOTHER MCS WHICH WAS SATISFACTORY TO THE PHYSICIAN AND THE SURGERY CONTINUED WITH VERY LITTLE DELAY. ADDITIONALLY, THE LARGE NEEDLE DRIVER HAD AN OUT OF THE BOX FAILURE. THE LND WAS NOT RECOGNIZED WITHIN THE SIXTY SECOND WINDOW WHICH CAUSED A YELLOW INSTRUMENT ERROR. INSTRUCTED BEDSIDE ASSIST TO DETACH THE INSTRUMENT, RECONNECTED INSTRUMENT BUT SAME ERROR. AFTER SIXTY SECONDS OF USE, IT ALARMED ONCE MORE. THE LND WAS SWITCHED FOR A NEW LND AND IT WAS RECOGNIZED AND THE SURGERY CONTINUED SUCCESSFULLY. THERE WAS NO ISSUE WITH THE STERILE INTERFACE MODULE (SIM). THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24688966 ↗

Other Modular electromechanical surgical system devices

All Modular electromechanical surgical system devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page