Skip to content

Maxon 8886 6213-31Recalls & reported problems

Covidien LP · Monofilament Absorbable Suture · Class 2 · Implant

All models of Maxon (114)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10884523002317), received between Nov 13, 2017 and Mar 20, 2025 — classified as 2 injuries, 6 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (6 reports).

1 other recall cover devices cleared under the same FDA submission (K990951); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component6
  • Adverse Event Without Identified Device or Use Problem1
  • Unraveled Material1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Code Available2
  • Bowel Perforation1
  • Failure to Anastomose1
  • Fever1
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 20, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, PREOPERATIVELY, WHILE PREPARING BEFORE PATIENT USE, WHEN THE SUTURE WAS BEING PULLED OUT FROM THE RETAINER, THE NEEDLE DETACHED. TO RESOLVE THE ISSUE, A NEW SUTURE WAS USED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21655036 ↗
  • MalfunctionOct 21, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING A LAPAROTOMY, RESECTION OF MUCUS CYST, AT THE TIME OF SEWING THE LIPS TOGETHER WHILE DOING A CONTINUOUS SUTURING TECHNIQUE, AFTER OPENING THE PACKAGE, THE NEEDLE AND THREAD SEPARATED FROM THE THREADING AREA. A NEW SUTURE FROM THE SAME BATCH WAS USED WITHOUT ANY PROBLEM. NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15654758 ↗
  • MalfunctionSep 16, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, PRE-OPERATIVELY, THE DEVICE'S NEEDLE DETACHED WHEN THE SUTURE WAS REMOVED FROM THE MOUNT. A NEW DEVICE WAS USED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12485000 ↗
  • InjuryMar 27, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, POST OPERATIVE OPERATION FOR DUODENAL ATRESIA OPEN PROCEDURE, THE PATIENT UNDERWENT SECOND OPERATION AND ANASTOMOTIC LEAK WAS FOUND. ADDITIONAL OPERATION WAS PERFORMED TO CORRECT THE ISSUE. THE PATIENT HAD DISTENDED ABDOMEN AND FEVER POST OPERATION. THE X-RAY SHOWED FREE AIR.

    Patient problems: Bowel Perforation; Fever; No Code Available; Radiation Exposure, Unintended

    MAUDE report 9889046 ↗
  • InjuryMar 16, 2020Unraveled Material; Unraveled Material

    ACCORDING TO THE REPORTER, POST-OPERATIVE FROM A TRANSANAL PULL THROUGH, PATIENT REVISITED BECAUSE HIS FECES HAD SUTURES IN IT. A SECOND OPERATION WAS PERFORMED AND FOUND THAT ALMOST ALL SUTURES DISLODGED FROM THE ANASTOMOTIC POINT. LAPAROTOMY WAS DONE AND ANASTOMOTIC LEAK WAS FOUNDED. RESUTURING WAS PERFORMED.

    Patient problems: Failure to Anastomose; No Code Available

    MAUDE report 9834934 ↗
  • MalfunctionOct 17, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING AN OPEN WHIPPLE PROCEDURE, THE NEEDLE BROKE OFF FROM THE THREAD ON A COUPLE OF SUTURES WITH THE SAME LOT NUMBER. ANOTHER SUTURE WAS USED FROM THE SAME LOT NUMBER WAS USED TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9203532 ↗
  • MalfunctionJun 12, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, PRIOR TO THE PROCEDURE, WHEN THE PRODUCT WAS OPENED, THE THREAD WAS DISENGAGED FROM THE NEEDLE. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Patient Involvement

    MAUDE report 8693327 ↗
  • MalfunctionNov 13, 2017Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, ON THE FIRST STITCH, THE NEEDLE CAME OFF THE THREAD. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. THE STATUS OF THE PATIENT IS NO PROBLEM.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7024854 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K990951. One submission often covers many models, so check the product and lot details in each record.

  • Z-2542-2017 · May 3, 2017 · Terminated
    Sutures. MAXON: MAXON 4/0 5X18" GREEN CV-24 D-TACH; Item Number: 8886621933; MAXON 5/0 5X18 GR CV-23 D-TACH; Item Number: XX2109; MAXON* 4/0 5X30 GREEN V-20 D*TACH; Item Number:

Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices

All SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page