Skip to content

Maxon GMMT-563-MGRecalls & reported problems

Covidien LP · Monofilament Absorbable Suture · Class 2 · Implant

All models of Maxon (114)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10884523004458), received between May 15, 2024 and May 15, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

1 other recall cover devices cleared under the same FDA submission (K990951); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 3 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 15, 2024Break; Break

    ACCORDING TO THE REPORTER, DURING A ROBOTIC-ASSISTED AND LAPAROSCOPIC SIGMOID COLECTOMY, AT WOUND CLOSURE, AN INTERRUPTED SUTURE WAS USED AS USUAL, BUT WHEN THE SURGEON TRIED TO TIGHTEN THE DEVICE, BOTH SUTURES BROKE. EVEN WHEN THE SURGICAL TEAM TRIED TO HOLD AND STRETCH THE THIRD SUTURE WITH BOTH HANDS, THERE WAS NO SUTURE BREAKAGE, AND WHEN TRIED LIGATING IT AGAIN, THE SUTURE BROKE. THE USAGE WAS DISCONTINUED, AND ANOTHER DEVICE FROM A DIFFERENT MANUFACTURER WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19323229 ↗
  • MalfunctionMay 15, 2024Break; Break

    ACCORDING TO THE REPORTER, DURING A ROBOTIC-ASSISTED AND LAPAROSCOPIC SIGMOID COLECTOMY, AT WOUND CLOSURE, AN INTERRUPTED SUTURE WAS USED AS USUAL, BUT WHEN THE SURGEON TRIED TO TIGHTEN THE DEVICE, BOTH SUTURES BROKE. EVEN WHEN THE SURGICAL TEAM TRIED TO HOLD AND STRETCH THE THIRD SUTURE WITH BOTH HANDS, THERE WAS NO SUTURE BREAKAGE, AND WHEN TRIED LIGATING IT AGAIN, THE SUTURE BROKE. THE USAGE WAS DISCONTINUED, AND ANOTHER DEVICE FROM A DIFFERENT MANUFACTURER WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19323225 ↗
  • MalfunctionMay 15, 2024Break; Break

    ACCORDING TO THE REPORTER, DURING A ROBOTIC-ASSISTED AND LAPAROSCOPIC SIGMOID COLECTOMY, AT WOUND CLOSURE, AN INTERRUPTED SUTURE WAS USED AS USUAL, BUT WHEN THE SURGEON TRIED TO TIGHTEN THE DEVICE, BOTH SUTURES BROKE. EVEN WHEN THE SURGICAL TEAM TRIED TO HOLD AND STRETCH THE THIRD SUTURE WITH BOTH HANDS, THERE WAS NO SUTURE BREAKAGE, AND WHEN TRIED LIGATING IT AGAIN, THE SUTURE BROKE. THE USAGE WAS DISCONTINUED, AND ANOTHER DEVICE FROM A DIFFERENT MANUFACTURER WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19323089 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K990951. One submission often covers many models, so check the product and lot details in each record.

  • Z-2542-2017 · May 3, 2017 · Terminated
    Sutures. MAXON: MAXON 4/0 5X18" GREEN CV-24 D-TACH; Item Number: 8886621933; MAXON 5/0 5X18 GR CV-23 D-TACH; Item Number: XX2109; MAXON* 4/0 5X30 GREEN V-20 D*TACH; Item Number:

Other SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices

All SUTURE, ABSORBABLE, SYNTHETIC, POLYGLYCOLIC ACID devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page