Skip to content

Argyle 8888501015Recalls & reported problems

Cardinal Health, Inc. · Flexible Yankauer,Sump Tip, Regular Capacity, No Vent · Class 2

Adverse event reports
9
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10884527019830), received between Oct 30, 2019 and Oct 29, 2025 — classified as 1 injury, 8 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component4
  • Suction Problem4
  • Decrease in Suction1

Patient problems

  • Foreign Body In Patient4
  • Blood Loss2
  • No Consequences Or Impact To Patient2
  • Cough1
  • Dysphagia/ Odynophagia1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 29, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER REPORTED THAT ON THE EVENING OF (B)(6) 2025, ANESTHESIA USED A YANKAUER SUCTION AT THE END OF THE CASE. AFTER THE PATIENT WAS DISCHARGED HOME, SHE ADMITTED TO HAVING SOME IRRITATION AND DIFFICULTY SWALLOWING. AFTER GOING TO BED THAT NIGHT, SHE WOKE UP COUGHING AND THE BLUE TIP FROM THE YANKAUER WAS IN HER MOUTH. PER CUSTOMER KNOWLEDGE, THE PATIENT HAD NO FURTHER ISSUES AND AGREED TO BRING THE TIP BACK TO THE HOSPITAL.

    Patient problems: Cough; Dysphagia/ Odynophagia; Foreign Body In Patient

    MAUDE report 23417505 ↗
  • MalfunctionFeb 21, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER REPORTED DURING A TOTAL ABDOMINAL HYSTERECTOMY IN LAPAROTOMY, THE CANNULA BASKET DETACHED IN THE PATIENT'S ABDOMINAL CAVITY, AND WAS PROMPTLY RECOVERED.

    Patient problems: Foreign Body In Patient

    MAUDE report 18747629 ↗
  • MalfunctionMay 18, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER REPORTED THAT ONE OF THE NURSES WAS USING THE COVIDIEN YANKAUER AND THE TIP OF THE DEVICE CAME OFF AND THE PATIENT ALMOST ASPIRATED ON IT. ADDITIONAL INFORMATION WAS RECEIVED FROM THE CUSTOMER AND STATED THAT THEY WERE DOING ORAL CARE AND THE PATIENT WAS UNABLE TO FOLLOW DIRECTION, NEEDED SUCTIONING. THE NURSE INSERTED YANKAUER INTO LEFT CHEEK TO SUCTION AND WHEN REMOVED, TIP WAS NOT PRESENT. THE RESPIRATORY THERAPIST (RT) AND REGISTERED NURSE (RN) ASSISTED TO KEEP PATIENT CALM AND HELP OPEN THE MOUTH TO ALLOW THE OTHER RN TO LOOK FOR TIP AND RETRIEVE. THE NURSE WAS ABLE TO RETRIEVE THE TIP FROM PATIENT¿S POSTERIOR ASPECT OF THE ORAL CAVITY NEAR THE OROPHARYNX. PER CUSTOMER, TONGUE DEPRESSOR, FLASHLIGHT AND TWEEZERS WERE USED TO RETRIEVE THE TIP. THERE WAS NO MEDICAL INTERVENTION OR TREATMENT REQUIRED BECAUSE OF THE INCIDENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 16958095 ↗
  • MalfunctionMar 4, 2020Decrease in Suction; Decrease in Suction

    THE CUSTOMER REPORTED THAT THE PROBE BECAME BLOCKED DURING THE PROCEDURE, CAUSING AN INADEQUATE BLOOD ASPIRATION AND BLOOD CLOTS. THE DEVICE HAD TO BE CHANGED. THERE WERE NO CONSEQUENCES FOR THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9788509 ↗
  • InjuryDec 16, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER REPORTED THE TIP FELL DOWN DURING SURGERY. THE SURGEON MANAGED TO GET IT BACK. NO INJURY ON THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 9476765 ↗
  • MalfunctionOct 31, 2019Suction Problem; Suction Problem

    THE CUSTOMER REPORTED THE UNIT IS NOT ADEQUATE TO SUCTION THE BLOOD AND BLOOD CLOTS, DESPITE THE CANNULA BEING CHANGED. THE PATIENT HAD PROLONGED BLEEDING AND INCREASED SEVERE BLOOD LOSS. THE PATIENT DID NOT REQUIRE A BLOOD TRANSFUSION. NO OTHER INFORMATION IS AVAILABLE AT THIS TIME.

    Patient problems: Blood Loss

    MAUDE report 9263293 ↗
  • MalfunctionOct 31, 2019Suction Problem; Suction Problem

    THE CUSTOMER REPORTED THE UNIT IS NOT ADEQUATE TO SUCTION THE BLOOD AND BLOOD CLOTS, DESPITE THE CANNULA BEING CHANGED. THE PATIENT HAD PROLONGED BLEEDING AND INCREASED SEVERE BLOOD LOSS. THE PATIENT DID NOT REQUIRE A BLOOD TRANSFUSION. NO OTHER INFORMATION IS AVAILABLE AT THIS TIME.

    Patient problems: Blood Loss

    MAUDE report 9263276 ↗
  • MalfunctionOct 30, 2019Suction Problem; Suction Problem

    THE CUSTOMER REPORTED THAT THE PROBE IS OCCLUDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9257363 ↗
  • MalfunctionOct 30, 2019Suction Problem; Suction Problem

    THE CUSTOMER REPORTED THAT THE PROBE IS OCCLUDED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9257351 ↗

Other APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED devices

All APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.