Salem Sump 266197 (REF 266197-)Recalls & reported problems
Cardinal Health 200, LLC · Anti-Reflux Valve · Class 2
All models of Salem Sump (31)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10884527022762), received between Jun 19, 2025 and Feb 5, 2026 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Break” (5 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Sep 18, 2025Class IIOpen, Classified
Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.
Z-0209-2026 · Root cause: Use error · Recall details · FDA record ↗
Most reported problems
Device problems
- Break5
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Anxiety1
- Discomfort1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 5, 2026Break
THE CUSTOMER REPORTED THIS PRODUCT HAS BEEN BREAKING OFF AT THE WHITE HALF WITH THE PERFORATED HOLES. THIS LEAVES THE BLUE HALF LODGED IN THE NG. THIS HALF IS ALMOST IMPOSSIBLE TO REMOVE UNLESS MANIPULATED WITH THE USE OF LARGE PLIERS AND SCISSORS, WHICH THEY WERE ABLE TO DO FOR THIS PATIENT. THIS DID CAUSE STRESS AND UNNECESSARY DISCOMFORT TO THE PATIENT. THIS IS NOT THE FIRST TIME THIS PRODUCT HAS BROKE WHILE IN USE WITH A PATIENT AND IT ALWAYS BREAKS AT THAT SAME SPOT.
Patient problems: Anxiety; Discomfort
MAUDE report 24274966 ↗MalfunctionFeb 5, 2026Break
THE CUSTOMER REPORTED AN ONGOING ISSUE WITH THE SPIGOT SNAPPING ON THE WHITE PART WHEN TRYING TO REMOVE FROM NG TUBE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24274952 ↗MalfunctionFeb 5, 2026Break
THE CUSTOMER REPORTED THAT ONE OF THE ANTI REFLUX VALVES FROM THE LAST DELIVERY WAS BROKEN IN THE PACKAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24272234 ↗MalfunctionJul 7, 2025Break; Break
THE CUSTOMER REPORTED THAT THE ANTI REFLUX CAP HAS BROKEN IN HALF. PER CUSTOMER, THE NURSE WAS WITH THE PATIENT IN BED DURING THE INCIDENT, THE PATIENT DID NOT KNOW WHY THE CAP HAD BROKEN IN HALF. THE ISSUE WAS DETECTED DURING THE COMMISSIONING OF THE PRODUCT; THIS HAS HAD NO CONSEQUENCES FOR THE PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22430825 ↗MalfunctionJun 19, 2025Break; Break
THE CUSTOMER REPORTED THAT THE ANTI REFLUX CAP HAS ALREADY BROKEN IN HALF 2 TIMES. PER CUSTOMER, FIRST TIME, THE NURSE COULDN'T GET THE CAP OFF AND THEN IT BROKE IN HALF. SECOND TIME WAS IN BED WITH THE PATIENT, THE PATIENT DID NOT KNOW WHY IT HAD BROKEN IN HALF. THE ISSUE WAS DETECTED DURING THE COMMISSIONING OF THE PRODUCT; THIS HAS HAD NO CONSEQUENCES FOR THE PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22292865 ↗
Other TUBES, GASTROINTESTINAL (AND ACCESSORIES) devices
- Bard® DigniShield® Stool Management System SMS002C. R. Bard, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)2 deaths525 reports
- CORTRAK* 2 ENTERAL ACCESS SYSTEM 20-0950Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall22 deaths134 reports
- Bard Tri-Funnel Replacement Gastrostomy Tube 000720Bard Access Systems, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 death170 reports
- EndoVive Safety PEG Kit M00566470Boston Scientific Corporation · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall1 death111 reports
- Kangaroo 8884711253Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)152 reports
- NutraGlide Nasal Feeding Tube ENFT-06090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- NutraGlide Nasal Feeding Tube ENFT-05090-IApplied Medical Technology, Inc. · Tubes, Gastrointestinal (And Accessories)2 recalls0 reports
- Kangaroo 8884711253ECardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)100 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page