MEDI-VAC 65652-611Recalls & reported problems
Cardinal Health 200, LLC · CRD Flex 100cc Canister W/Valve · Class 2
All models of MEDI-VAC (6)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 1
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10885380042638), received between Jun 21, 2018 and Nov 24, 2021 — classified as 1 death, 3 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break2
- Suction Failure2
- Suction Problem2
- Crack1
- Device Inoperable1
- Device Markings/Labelling Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Aspiration/Inhalation1
- Death1
- Insufficient Information1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 24, 2021Break; Suction Failure; Break; Suction Failure
CUSTOMER REPORTED THE CANISTERS ARE BREAKING WITHIN 4 MONTHS THAT THEIR SHELF LIFE IS BECOMING SHORTER. CANISTER LOST SUCTION ON HIGH, LOW OR BOTH AND/OR CRACKED AT THE BASE AND SOMETIMES HAD DECREASED SUCTION DURING UNKNOWN PROCEDURE. CUSTOMER HAS STATED THEY USE (B)(6) WIPES TO CLEAN THEM DOWN AND LEAVE THE CANISTERS ON HIGH SUCTIONS 24/7 AND CLAMP OFF THE TUBING, THEREFORE THE CANISTERS ARE UNDER CONSTANT PRESSURE. THERE WAS NO REPORTED ADVERSE EVENT TO PATIENT. NO FURTHER INFORMATION WAS PROVIDED BY CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12872694 ↗MalfunctionNov 24, 2021Break; Crack; Suction Problem; Device Markings/Labelling Problem; Suction Failure; Break; Suction Failure
CUSTOMER REPORTED THE CANISTERS ARE BREAKING WITHIN 4 MONTHS OF USE AND THAT THEIR SHELF LIFE IS BECOMING SHORTER. CANISTER LOST SUCTION ON HIGH, LOW OR BOTH AND/OR CRACKED AT THE BASE AND SOMETIMES HAD DECREASED SUCTION DURING UNKNOWN PROCEDURE. CUSTOMER HAS STATED THEY USE CLINELL WIPES TO CLEAN THEM DOWN AND LEAVE THE CANISTERS ON HIGH SUCTIONS 24/7 AND CLAMP OFF THE TUBING. THERE WAS NO REPORTED ADVERSE EVENT TO PATIENT. NO FURTHER INFORMATION WAS PROVIDED BY CUSTOMER.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 12872601 ↗MalfunctionJul 6, 2018Suction Problem
A (B)(6) PATIENT WAS TREATED FOR AN IDIOPATHIC THORACOLUMBAR SCOLIOSIS OF ADOLESCENCE. IN POST-SURGERY, DURING A GASTRIC ASPIRATION A MEDI-VAC SUCTION SYSTEM, CONNECTED TO CH12 NASOGASTRIC PROBE (FITTING BUCKET) WAS USED. SEVERAL SUCTION SETS AND SEVERAL REGULATORS (DEPRESSION RATE: FROM 0 TO -200 MBAR) WERE USED BY THE NURSES BEFORE OBTAINING THE ASPIRATION. CLINICAL CONSEQUENCES: PROCEDURE DELAYED. PRACTITIONER PERFORMED A MANUAL ASPIRATION USING A SYRINGE. HEALTHCARE PROFESSIONALS WERE DOUBTFUL ON THEIR ABILITY TO ACHIEVE A RAPID AND EFFECTIVE ASPIRATION IN AN EMERGENCY SITUATION. MEASURES: THE DEFECTIVE CANISTERS WERE DISCARDED. DESPITE THE PREVIOUS MEDI-VAC SYSTEM INFORMATION OF USE TO PERSONAL, THE INTENDED USE IS STILL UNKNOWN (NO CANISTER FIXATION, NO REINFORCED TUBING FROM VACUUM SOURCE TO THE CANISTER).
Patient problems: Aspiration/Inhalation
MAUDE report 7666814 ↗DeathJun 21, 2018Device Inoperable
(B)(6) PATIENT, HOSPITALIZED FOR THE CARE OF A PATIENT FEBRILE NEUTROPENIA FOLLOWING CHEMOTHERAPY FOR EVOLUTIONARY T-CELL LYMPHOMA. ON THE NIGHT OF (B)(6) TO (B)(6), THE PATIENT HAD A SUDDEN DYSPNEA WITH CARDIOPULMONARY ARREST, TAKEN CARE OF BY A DOCTOR (FROM INTENSIVE CARE UNIT). THE PATIENT VOMITS ABUNDANTLY, CAUSING A DEFECT OF VENTILATION. AN INTUBATION WAS NECESSARY. THE DOCTOR INTENDED TO PERFORM A BRONCHO-ASPIRATION USING THE MEDIC-VAC SYSTEM. THE FACILITY REPORTED THAT THIS WAS NOT POSSIBLE BECAUSE THE SYSTEM DID NOT WORK (NO DEPRESSION IN THE PATIENT'S MANIFOLD). THE USE OF A SECOND SYSTEM PERMITTED TO OBTAIN A SUCTION. CONSEQUENCES : PROCEDURE DELAYED. THE FACILITY REPORTED THAT VENTILATION WAS NOT POSSIBLE DUE TO ASPIRATION DEFECT. THE CARDIO-RESPIRATORY REANIMATION WAS STOPPED DUE TO THE SERIOUS NEUROLOGICAL PROGNOSIS. THE PATIENT DIED.
Patient problems: Death
MAUDE report 7627023 ↗
Other APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED devices
- StrykeFlow 0250070520Stryker Corporation · APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED2 recalls250 reports
- StrykeFlow 0250070505Stryker Corporation · APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED1 recall3 reports
- Sentinel Seal 8888571562Cardinal Health 200, LLC · APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED1 recall0 reports
- Aqua-Seal 8888571406Cardinal Health 200, LLC · APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED1 recall0 reports
- Sentinel Seal 8888571513Cardinal Health 200, LLC · APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED1 recall0 reports
- Sentinel Seal 8888571489Cardinal Health 200, LLC · APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED1 recall0 reports
- spineX BONE DUST COLLECTOR S-4500Oto-Med, Inc. · Apparatus, Suction, Operating-Room, Wall Vacuum Powered1 recall0 reports
- CRD 65651-510Cardinal Health 200, LLC · APPARATUS, SUCTION, OPERATING-ROOM, WALL VACUUM POWERED1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page