CARDINAL HEALTH CMT-40-1Recalls & reported problems
Cardinal Health 200, LLC · Orthopaedic cement, non-medicated · Class 2 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10885380066511), received between Feb 13, 2017 and Sep 14, 2020 — classified as 3 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Device Handling Problem” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Handling Problem4
- Inadequate or Insufficient Training4
- Adverse Event Without Identified Device or Use Problem3
- Loosening of Implant Not Related to Bone-Ingrowth3
- Patient Device Interaction Problem3
Patient problems
- Failure of Implant7
- No Known Impact Or Consequence To Patient4
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjurySep 14, 2020Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem; Loosening of Implant Not Related to Bone-Ingrowth; Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem; Loosening of Implant Not Related to Bone-Ingrowth
ON (B)(6) 2018, (B)(6) UNDERWENT LEFT-SIDED TOTAL KNEE REPLACEMENT SURGERY PERFORMED BY DR. (B)(6) AT (B)(6) HOSPITAL IN (B)(6). HE WAS IMPLANTED WITH BIOMET VANGUARD COMPONENTS. ON (B)(6) 2018, IN ORDER TO BOND THESE COMPONENTS, DR. (B)(6) UTILIZED THE DEFECTIVE CARDINAL HV BONE CEMENT. AFTER THE CARDINAL HV BONE CEMENT WAS IMPLANTED, MR. (B)(6) BEGAN EXPERIENCING SEVERE AND PERSISTENT PAIN, DISCOMFORT, INSTABILITY AND DIFFICULTY AMBULATING DUE TO ASEPTIC LOOSENING. ON (B)(6) 2019, PATIENT UNDERWENT REVISION SURGERY DUE TO GROSS LOOSENING OF HIS KNEE COMPONENTS IN HIS LEFT KNEE. THIS SURGERY WAS PERFORMED BY DR. (B)(6) AT (B)(6) HOSPITAL IN (B)(6). IN THIS PROCEDURE, DR. (B)(6) ASSESSED THE IMPLANT COMPONENTS AND FOUND "GROSS LOOSENING OF THE TIBIAL COMPONENT." NO X-RAYS ARE AVAILABLE. NO FURTHER INFORMATION HAVE BEEN PROVIDED.
Patient problems: Failure of Implant
MAUDE report 10527111 ↗InjuryMay 6, 2020Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem; Loosening of Implant Not Related to Bone-Ingrowth; Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem; Loosening of Implant Not Related to Bone-Ingrowth
ON (B)(6)2016, MS. (B)(6) UNDERWENT TOTAL LEFT KNEE REPLACEMENT BY IMPLANTING SURGEON, DR. (B)(6) WITH CARDINAL HEALTH ARTHROPLASTY BONE CEMENT. ON (B)(6) 2019, MS. (B)(6) TO UNDERGO UNDERGO A REVISION SURGERY DUE TO ASEPTIC LOOSENING OF HIS ARTIFICIAL KNEE COMPONENTS. NO X-RAYS ARE AVAILABLE. NO FURTHER INFORMATION HAVE BEEN PROVIDED.
Patient problems: Failure of Implant
MAUDE report 10028062 ↗InjuryMay 6, 2020Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem; Loosening of Implant Not Related to Bone-Ingrowth; Adverse Event Without Identified Device or Use Problem; Patient Device Interaction Problem; Loosening of Implant Not Related to Bone-Ingrowth
ON (B)(6) 2016, MR. (B)(6) UNDERWENT BILATERAL TOTAL KNEE REPLACEMENTS BY IMPLANTING SURGEON, DR. (B)(6) WITH CARDINAL HEALTH ARTHROPLASTY BONE CEMENT. ON (B)(6) 2017, MR. (B)(6) TO UNDERGO A RIGHT-SIDED REVISION SURGERY DUE TO ASEPTIC LOOSENING OF HIS ARTIFICIAL KNEE COMPONENTS. SIMILARLY, ON (B)(6) 2018, MR. (B)(6) UNDERGO A LEFT-SIDED REVISION SURGERY DUE TO ASEPTIC LOOSENING OF HIS ARTIFICIAL KNEE COMPONENTS. NO X-RAYS ARE AVAILABLE. NO FURTHER INFORMATION HAVE BEEN PROVIDED.
Patient problems: Failure of Implant
MAUDE report 10028060 ↗MalfunctionFeb 13, 2017Inadequate or Insufficient Training; Device Handling Problem; Inadequate or Insufficient Training; Device Handling Problem
MIXINNG TIME: 90 SECOUNDS; BONE CEMENT IMPLANTATION AFTER 2:20 MINUTES; TIBIA PLATE IMPLANTATION AFTER 3:20 MINUTES; CURING TIME 15 MINUTES (LONGER THAN EXPECTED). DELAY IN SURGERY (SOME MINUTES).
Patient problems: Failure of Implant; No Known Impact Or Consequence To Patient
MAUDE report 6325164 ↗MalfunctionFeb 13, 2017Inadequate or Insufficient Training; Device Handling Problem; Inadequate or Insufficient Training; Device Handling Problem
MIXINNG TIME: 90 SECONDS; BONE CEMENT IMPLANTATION AFTER 2:38 MINUTES; TIBIA PLATE IMPLANTATION AFTER 3:30 MINUTES; CURING TIME 15 MINUTES (LONGER THAN EXPECTED); DELAY IN SURGERY (SOME MINUTES).
Patient problems: Failure of Implant; No Known Impact Or Consequence To Patient
MAUDE report 6325161 ↗MalfunctionFeb 13, 2017Inadequate or Insufficient Training; Device Handling Problem; Inadequate or Insufficient Training; Device Handling Problem
MIXING TIME: 50 SECONDS; IMPLANTATION OF BONE CEMENT INTO FEMUR AFTER 2:04 MINUTES; INSERT OF FEMORAL COMPONENT AFTER 4:48 MINUTES; 15 MINUTES FLEXED KNEE (CEMENT CURED); DELAY IN SURGERY (SOME MINUTES).
Patient problems: Failure of Implant; No Known Impact Or Consequence To Patient
MAUDE report 6325129 ↗MalfunctionFeb 13, 2017Inadequate or Insufficient Training; Device Handling Problem; Inadequate or Insufficient Training; Device Handling Problem
MIXING TIME: 25 SECONDS; WAITING TIME: 32 SECONDS; IMPLANTATION OF TIBIA BASE PLATE AFTER 2:06 MINUTES; CEMENT WAS TOO THIN FOR IMPLANTATION AND HAS NOT REACHED THE DOUGH STATE USAGE, CEMENT RAN DOWN THE LEG. DELAY IN SURGERY (SOME MINUTES).
Patient problems: Failure of Implant; No Known Impact Or Consequence To Patient
MAUDE report 6325128 ↗
Other BONE CEMENT devices
- SMARTSET 3122-040Depuy (Ireland) · BONE CEMENT824 reports
- NA 3322-020Depuy (Ireland) · BONE CEMENT306 reports
- NA 3312-040Depuy (Ireland) · BONE CEMENT4 deaths258 reports
- NA 3332-040Depuy (Ireland) · BONE CEMENT4 deaths197 reports
- NA 5450-31-500Depuy (Ireland) · BONE CEMENT2 deaths168 reports
- NA 5450-32-500Depuy (Ireland) · BONE CEMENT1 death150 reports
- PALACOS® MV pro 80Heraeus Medical GmbH · Bone Cement2 recalls1 report
- PALACOS® R pro 80Heraeus Medical GmbH · Bone Cement2 recalls1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page