Skip to content

CARDINAL HEALTH 6708888Recalls & reported problems

Cardinal Health 200, LLC · Cardinal Health NPWT PRO NPWT Therapy Device · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10885380103520), received between Sep 1, 2020 and Apr 13, 2022 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Suction Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Suction Problem2
  • Decrease in Suction1
  • Defective Alarm1
  • Device Alarm System1
  • Device Displays Incorrect Message1
  • Leak/Splash1
  • Self-Activation or Keying1
  • Suction Failure1

Patient problems

  • Insufficient Information2
  • Fluid Discharge1
  • No Clinical Signs, Symptoms or Conditions1
  • Post Operative Wound Infection1
  • Sepsis1
  • Septic Shock1
  • Wound Dehiscence1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 13, 2022Device Alarm System; Suction Failure

    CUSTOMER REPORTED THE DEVICE WAS NOT SUCTIONING AND THERE WAS A FAILURE TO ALARM. THERE WAS NO PATIENT INJURY, AS THE DEVICE WAS IMMEDIATELY REMOVED FROM PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14107033 ↗
  • InjuryJan 7, 2022Leak/Splash; Suction Problem; Suction Problem

    CARDIAC SURGEON REPORT STATED THAT A PATIENT USING A NEGATIVE PRESSURE WOUND THERAPY DEVICE AND HAS REPORTED THE PATIENT TO BE IN SEPTIC SHOCK DUE TO UNIT REPORTEDLY LEAKING. PATIENT WAS ADMITTED TO ICU.

    Patient problems: Sepsis; Septic Shock

    MAUDE report 13193890 ↗
  • InjuryDec 10, 2021Decrease in Suction; Suction Problem; Suction Problem

    CUSTOMER REPORTED NURSE MANAGER HAD A MALE PATIENT WITH STERNAL INCISION THAT DEHISCED, SO THE NEGATIVE PRESSURE WOUND THERAPY PRO DEVICE WAS PLACED ON THIS COMPLEX CHEST WOUND. WHEN HE CAME HOME, HE HAD A LOT OF DRAINAGE, AND WAS ON IV ANTIBIOTICS, THE ANTIBIOTICS CEASED BUT THEY WERE STILL DOING 5X VISITS A WEEK BECAUSE THEY FELT THE SUCTION WASN'T WORKING PROPERLY AND THE PATIENT KEPT EXPERIENCING ISSUES WITH POOLING UNDER THE DRESSING. THE DRESSING HAD A GOOD SEAL, BUT THE PUMP WAS JUST NOT REMOVING ALL THE EXUDATE. THEY CHANGED THE PUMP FOR A NEW PRO DEVICE ON NOV 10 WORRIED THAT IT WAS A PUMP ISSUE, AND HE DID FEEL THERE WAS IMPROVED SUCTION WITH IT. UNFORTUNATELY, BECAUSE OF THE POOLING OF FLUID, HE ALSO HAD A LOT OF MACERATION BY THEN. HE ALSO DEVELOPED A FLUID FILLED POCKET AT THE BOTTOM OF THE WOUND. OVER THE WEEKEND (NOV 13-14) THE CLIENT WAS ADMITTED TO HOSPITAL WITH AN INFECTED WOUND SITE.

    Patient problems: Fluid Discharge; Post Operative Wound Infection; Wound Dehiscence

    MAUDE report 12978298 ↗
  • MalfunctionSep 1, 2020Device Displays Incorrect Message

    NO ALARM WENT OFF AND THE MACHINE CONTINUED TO ATTEMPT TO GET SUCTION/SEAL. TWO NEGATIVE PRESSURE WOUND THERAPY DEVICES WERE ATTEMPTED ON THE SAME PATIENT. THE LOT/SERIAL NUMBER FOR THIS ONE IS (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 10480270 ↗
  • MalfunctionSep 1, 2020Defective Alarm; Self-Activation or Keying

    NO ALARM WENT OFF AND THE MACHINE CONTINUED TO ATTEMPT TO GET SUCTION/SEAL. TWO NEGATIVE PRESSURE WOUND THERAPY DEVICES WERE ATTEMPTED ON THE SAME PATIENT. THE LOT/SERIAL NUMBER FOR THIS ONE IS (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 10480135 ↗

Other negative pressure wound therapy Powered suction pump devices

All negative pressure wound therapy Powered suction pump devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page