PleurX 50-7210Recalls & reported problems
Carefusion 2200, Inc · PleurX Vacuum Bottle, With Drainage Line, 1000 mL · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10885403090660), received between Dec 11, 2020 and Apr 6, 2022 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Material Separation” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Material Separation3
- Detachment of Device or Device Component1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 6, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
WHEN THE NURSE WANTED TO DRAIN THE PATIENT WITH A BOTTLE, THE TUBING (DRAINAGE LINE) FELL OFF THE BOTTLE NECK. SHE CLOSED THE CLAMP ON THE DRAINAGE LINE, TOOK A NEW BOTTLE AND THE SAME THING HAPPENED AGAIN. WAS THE DRAINAGE LINE DISCONNECTED FROM THE BOTTLE AT ANY POINT DURING THE DRAIN OR WAS IT NOTICED PRIOR TO DRAINAGE? YES, DURING THE DRAIN AND IT WAS NOTICED. IF IT'S BEFORE USE AND DID THEY ATTEMPT TO RECONNECT IT? DURING USE, AND NO ATTEMPT TO RECONNECT. NEED TO VERIFY THE LINE WAS DISCONNECTED PRIOR TO USE AND THE BOTTLE WAS NOT USED. BOTTLE WAS USED. WAS THE CLAMP PROPERLY CLOSED UNTIL AFTER THE PLUNGER WAS PRESSED, ALL OF OUR PATIENTS ARE TRAINED. SO THIS SHOULD BE DONE CORRECTLY. WHERE IS THE CATHETER LOCATED? ABDOMEN OR CHEST? UNKNOWN. IS THERE A PHOTO OR SAMPLE AVAILABLE SHOWING THE REPORTED ISSUE? THERE IS NO PHOT AND NO SAMPLE AVAILABLE. WHAT WAS THE IMPACT TO THE PATIENT? NO IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14021505 ↗MalfunctionAug 6, 2021Material Separation; Material Separation
REF (B)(4),LOT 0001347543: WHEN PUTTING IN SUCTION, DIFFICULTY IN PIERCING THE MEMBRANE. DURING INSTALLATION IN THE PATIENT'S BED, THE TUBING DISCONNECTS FROM THE JAR. NO CONSEQUENCES FOR THE PATIENT BUT BLOOD EXPOSURE ACCIDENT FOR THE MEDICAL STAFF (THE DOCTOR WAS CONTAMINATED BY BLOOD FROM THE TUBING WHEN IT CAME OFF THE JAR). WAS THE DRAIN LINE DISCONNECTED FROM THE CYLINDER AT ANY TIME DURING DRAINAGE OR WAS IT NOTICED BEFORE DRAINAGE? DURING DRAINAGE (AFTER A FEW MINUTES) FOR INCIDENT 21-172. IF THE FAULT WAS NOTICED BEFORE USE AND WAS AN ATTEMPT MADE TO RECONNECT IT? WAS THE LINE DISCONNECTED BEFORE USE AND THE BOTTLE DISCARDED IMMEDIATELY? NO DURING USE FOR INCIDENT 21-172 WAS THE BOTTLE USED? YES BUT DISCARDED AFTER DISCONNECTION WITH THE TUBING FOR INCIDENT 21-172. WAS THE CLAMP PROPERLY CLOSED UNTIL THE PISTON WAS PRESSED? YES FOR ALL THREE INCIDENTS. WHERE WAS THE CATHETER LOCATED? CHEST. IS THERE A PHOTO SHOWING THE REPORTED DEFECT IF POSSIBLE? NO. WAS THERE A VACUUM PROBLEM? DIFFICULTY IN STRIKING THE MEMBRANE, POOR SUCTION FOR ALL 3 INCIDENTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12289222 ↗MalfunctionFeb 26, 2021Material Separation; Material Separation
THE TUBE DISCONNECTED FROM THE BOTTLE. THEY CLOSE THE ROLLER CLAMP AND ACTIVATE THE VACUUM. WHEN THEY OPEN THE ROLLER CLAMP, THE TUBING FALLS OFF FROM THE BOTTLE. THEY IMMEDIATELY CLOSE THE ROLLER CLAMP. THEY USE A NEW BOTTLE AND ITS WORKING. FLUID IS DRAINED WITHOUT ANY PROBLEMS. FIRST THEY SUSPECT THAT AIR HAS ENTERED INTO THE PATIENT'S PLEURAL SPACE, BUT THE PATIENT IS NOT AFFECTED AND IS DOING FINE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11387878 ↗MalfunctionDec 11, 2020Material Separation; Material Separation
TODAY, I WOULD LIKE TO POINT OUT A PROBLEM THAT I HAVE NOTICED SEVERAL TIMES ON THE PLEURX DEVICES, THE BAD FIXING OF THE TUBING ON THE BOTTLE ITSELF. WHEN YOU REALIZE THIS BEFORE YOU PUT THE SUCTION ON, WE CAN PUSH IT BACK AND THE VACUUM WILL NOT GO OUT. LOT NUMBER IS 0001347542. 30NOV2020: EMAIL SENT REQUESTING IF ADDITIONAL RESPONSE HAS BEEN RECEIVED. PER CUSTOMER RESPONSE; AT LEAST 2 TIMES, TUBING DISCONNECTED FROM THE BASE ON THE BOTTLE, AND ONCE BY PUSHING IT BACK BEFORE VACUUMING THE DRAINAGE COULD TAKE PLACE WITH GOOD SUCTION, (COPY GIVEN BACK TO THE DR (B)(6) PHARMACY) NO INCIDENCE ON THE CARE, NO PATIENT DAMAGE THIS WAS ESTABLISHED BEFORE CONNECTING THE VACUUM DEVICE TO THE PATIENT. IT IS A THORACIC PLEURX.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10991342 ↗
Other Apparatus, Suction, Patient Care devices
- PleurX 50-7050Carefusion 2200, Inc · APPARATUS, SUCTION, PATIENT CARE137 reports
- PleurX 50-7500BCarefusion 2200, Inc · Apparatus, Suction, Patient Care66 reports
- Aspira® 00884450394915Merit Medical Systems, Inc. · APPARATUS, SUCTION, PATIENT CARE1 recall2 reports
- Aspira® 00884450394946Merit Medical Systems, Inc. · APPARATUS, SUCTION, PATIENT CARE1 recall2 reports
- Aspira® 00884450394861Merit Medical Systems, Inc. · APPARATUS, SUCTION, PATIENT CARE1 recall1 report
- Aspira® 00884450394984Merit Medical Systems, Inc. · APPARATUS, SUCTION, PATIENT CARE1 recall0 reports
- Aspira® 00884450394939Merit Medical Systems, Inc. · APPARATUS, SUCTION, PATIENT CARE1 recall0 reports
- Aspira® 00884450394960Merit Medical Systems, Inc. · APPARATUS, SUCTION, PATIENT CARE1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page