V. Mueller RH551-111Recalls & reported problems
Steris Corporation · V. Mueller WEIL-BLAKESLEY FORCEPS FENESTRATED, STRAIGHT, 00 2.5 X 5 MM CUP WORKING LENGTH 4-1/2 inches 11.5CM · Class 1
All models of V. Mueller (117)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10885403110221), received between Jul 11, 2025 and Aug 26, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Material Fragmentation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 26, 2025Break; Material Fragmentation; Break; Material Fragmentation
ON 04AUG2025, AVALIGN GSI QUALITY RECEIVED A CUSTOMER COMPLAINT NOTIFICATION. COMPLAINT FROM CUSTOMER FOR ONE UNIT OF CUSTOMER PART# RH551-111. IT WAS REPORTED THAT A PIECE OF INSTRUMENT BROKE OFF IN THE PATIENT'S NASAL CAVITY, AND THE BROKEN PIECE WAS RETRIEVED AND INTACT. THERE WAS NO ADDITIONAL INTERVENTION, THE PATIENT WAS DISCHARGED WITH INJURY. THE PROCEDURE WAS NOT DELAYED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22896626 ↗MalfunctionJul 11, 2025Break; Break
THE USER FACILITY REPORTED VIA MEDWATCH REPORT (B)(4) THAT DURING A PATIENT PROCEDURE A PIECE OF THEIR WEIL-BLAKESLY FORCEP BROKE. THE PIECE WAS SUCCESSFULLY RETRIEVED FROM THE PATIENT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH ANOTHER DEVICE. NO REPORT OF INJURY OR PROCEDURE DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22482777 ↗
FDA clearance and classification
The FDA regulates product code KAE (forceps, ent) under 21 CFR 874.4420 as a Class I device. 21 CFR 874.4420 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the V. Mueller RH551-111
Has the V. Mueller RH551-111 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10885403110221), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the V. Mueller RH551-111?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Jul 11, 2025 and Aug 26, 2025. In those reports, the most common specific device problems are “Break” (2) and “Material Fragmentation” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the V. Mueller RH551-111?
None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the V. Mueller RH551-111 FDA approved?
Devices of this type (product code KAE) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Steris Corporation.
Who makes the V. Mueller RH551-111, and is it still sold?
The FDA's device database lists Steris Corporation as the company responsible for it (brand V. Mueller, model RH551-111). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other FORCEPS, ENT devices
- AESCULAP BH951RAesculap AG · FORCEPS, ENT1 recall0 reports
- AESCULAP BH963RAesculap AG · FORCEPS, ENT1 recall0 reports
- AESCULAP BH957RAesculap AG · FORCEPS, ENT1 recall0 reports
- AESCULAP BH961RAesculap AG · FORCEPS, ENT1 recall0 reports
- AESCULAP BH959RAesculap AG · FORCEPS, ENT1 recall0 reports
- AESCULAP BH965RAesculap AG · FORCEPS, ENT1 recall0 reports
- AESCULAP BH952RAesculap AG · FORCEPS, ENT1 recall0 reports
- Integra® 3711063Integra Lifesciences Production Corporation · FORCEPS, ENT2 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page