Safe-T TPT1005SPRecalls & reported problems
Carefusion 2200, Inc · Safe-T PLUS Thora-Para 5 Fr Catheter Drainage Tray · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10885403220869), received between Jan 5, 2021 and Jan 5, 2021 — classified as 1 injury.
The most frequently reported specific device problem is “Disconnection” (1 report).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- May 28, 2020Terminated
The Universal Drainage Set (UDS) may contain valves that leak internally which may result in the introduction of up to 30cc of air into the patient.
Z-2659-2020 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Disconnection1
- Material Separation1
Patient problems
- Pneumothorax1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJan 5, 2021Disconnection; Material Separation; Disconnection; Material Separation
MATERIAL NO.: TPT1005SP; BATCH NO.: 0001381128. IT WAS REPORTED THAT WHEN REMOVING THE METAL STYLET, THE BALL VALVE AND MEMBRANE MECHANISM PULLED OUT OF THE 3 WAY HOUSING. PER ATTACHED FORM: PER THE PHYSICIAN: "AFTER PUNCTURE, WHEN REMOVING THE METAL STYLET, THE BALL VALVE AND MEMBRANE MECHANISM PULLED OUT OF THE 3WAY HOUSING. THIS ALLOWED AIR TO TRAVEL QUICKLY INTO THE PLEURAL SPACE. I WAS NOT ABLE TO COVER IT WITH MY FINGER DUE TO ITS IRREGULAR SHAPE, AND IT TOOK ME A MOMENT TO TURN THE 3WAY TO CLOSED POSITION." CUSTOMER RESPONSE: PLEASE CONFIRM WHETHER OR NOT THERE WAS ANY PATIENT IMPACT. PATIENT DEVELOPED A MODERATE PNEUMOTHORAX. DO YOU HAVE PHOTOS SHOWING THE REPORTED ISSUE? PICTURES ATTACHED. WHAT PROCEDURE WAS BEING DONE WHEN THE REPORTED ISSUE OCCURRED? THORACENTESIS. PLEASE USE THE ATTACHED PRE-PAID SHIPPING LABEL TO RETURN INSTRUMENT FOR INVESTIGATION. THE DEVICE WAS NOT SAVED.
Patient problems: Pneumothorax
MAUDE report 11119794 ↗
Other Thoracentesis tray devices
- Safe-T-Centesis PIG1260TCarefusion 2200, Inc · Thoracentesis tray2 recalls27 reports
- Safe-T-Centesis PIG1260KCarefusion 2200, Inc · Thoracentesis tray1 recall73 reports
- NA OTP5000Carefusion 2200, Inc · Thoracentesis tray2 recalls13 reports
- Safe-T-Centesis PIG1280KCarefusion 2200, Inc · Thoracentesis tray108 reports
- NA TPT1000Carefusion 2200, Inc · Thoracentesis tray1 recall1 death19 reports
- Safe-T-Centesis PIG1280TCarefusion 2200, Inc · Thoracentesis tray1 recall8 reports
- Safe-T TPT1000SPCarefusion 2200, Inc · Thoracentesis tray1 recall6 reports
- NA TPT1005Carefusion 2200, Inc · Thoracentesis tray1 recall1 report
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.