Skip to content

Alaris 10012241-0500Recalls & reported problems

BD Switzerland Sàrl · Closed Male Luer · Class 2

All models of Alaris (85)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This BD Alaris closed male luer is a connector used to transfer hazardous drugs through IV lines. Its 142 reports often describe leaks and tubing popping off the connector, and a few mention drug exposure. The reports include 6 injuries and no deaths. 1 recall, still open, began in May 2025 over leaks where the connector meets needle-free connectors, which can cause treatment delays and drug exposure. Reports fell to 12 in the last 12 months from 91 before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
138
Deaths reported
0
Injuries reported
6
Recalls naming it
1

Summary

The FDA's MAUDE database contains 138 adverse event reports that list this exact device (DI 10885403221705), received between Aug 9, 2018 and May 12, 2026 — classified as 6 injuries, 132 malfunctions.

The most frequently reported specific device problem is “Leak/Splash” (88 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 2 reportsDec 2024: 4 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 2 reportsMay 2025: 18 reportsJun 2025: 4 reportsJul 2025: 1 reportsAug 2025: 30 reportsSep 2025: 8 reportsOct 2025: 19 reportsNov 2025: 6 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202630
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. May 12, 2025Class IIOpen, Classified

    Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.

    Z-1973-2025 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Leak/Splash88
  • Fluid/Blood Leak30
  • Connection Problem6
  • Complete Blockage4
  • Break3
  • Material Fragmentation3
  • Material Integrity Problem3
  • Material Separation3

Patient problems

  • No Clinical Signs, Symptoms or Conditions102
  • No Information17
  • No Consequences Or Impact To Patient15
  • Underdose4
  • Chemical Exposure2
  • Insufficient Information1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 138 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 12, 2026Leak/Splash

    IT WAS REPORTED THAT THE BD TEXIUM CLOSED MALE LUER WITH FEMALE CAP HAD LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WE RECENTLY HAD AN INPATIENT LEAK INVOLVING TEXIUM. A NURSE DISCONNECTED BETWEEN THE TEXIUM AND THE SMARTSITE, AND A CHEMOTHERAPY SPILL OCCURRED. WE ARE ALSO TRYING TO UNDERSTAND WHETHER THIS MAY BE RELATED TO THE MAX ZERO NEEDLELESS CONNECTORS, SINCE THEY ARE DESIGNED TO PUSH FLUID OUT AND WE TYPICALLY EXPECT TO SEE FLUID PRESENT AT THE ENDO F THE CONNECTOR WHEN IT'S NOT PART OF A CLOSED SYSTEM. ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER: KINDLY PROVIDE THE MATERIAL NUMBER AND BATCH/LOT NUMBER OF THE MAX ZERO NEEDLELESS CONNECTORS. I'LL INQUIRE. ANY ADVERSE EVENT OR SERIOUS INJURY REPORTED TO PATIENT OR HEALTHCARE PROFESSIONAL? IF YES, PLEASE PROVIDE THE DETAILS. NO. CAN YOU PLEASE PROVIDE AN EXACT DATE OF EVENT IN FORMAT DD-MMM-YYYY? 23-APRIL-2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25153514 ↗
  • MalfunctionFeb 2, 2026Leak/Splash

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24239972 ↗
  • MalfunctionNov 21, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD ALARIS CLOSED MALE LUER WAS LEAKING. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING: IT WAS REPORTED BY CUSTOMER THAT LEAKAGE FROM THE TIP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23624924 ↗
  • MalfunctionNov 18, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE BD ALARIS CLOSED MALE LUER HAD LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WE¿VE BEEN RECEIVING MULTIPLE REPORTS OF LEAKAGE FROM THE TEXIUM TIP AGAIN. ADDITIONAL INFORMATION RECEIVED FROM THE CUSTOMER: WHAT WAS THE ISSUE YOU EXPERIENCED? LEAKAGE FROM THE TIP IMPACT OR CLINICAL EVENT THAT OCCURRED TO THE PATIENT, HEALTHCARE PROVIDER OR USER: THE PATIENT WOKE UP DUE TO LEAKAGE FROM THE TIP. SAFETY IMPACT WAS NOT MEASURED. COULDN¿T ESTIMATE THE AMOUNT OF LEAKAGE. BOTH SAFETY AND EFFICACY CONCERNS WITH LEAKAGE AND LESS DOSE ADMINISTRATION THAN INTENDED. ANY ADVERSE EVENT OR SERIOUS INJURY REPORTED TO PATIENT OR HEALTHCARE PROFESSIONAL? IF YES, PLEASE PROVIDE THE DETAILS. NOT ABLE TO ASSESS THE ADVERSE EVENTS OR SERIOUS HARM CAUSED BY ADMINISTERING LESS DOSE THAN INTENDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23578282 ↗
  • MalfunctionNov 10, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD ALARIS CLOSED MALE LUER WAS LEAKING. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING: IT WAS REPORTED BY THE CUSTOMER THAT LEAKING OF MEDICATION NOTED AT THE CONNECTION SITE OF THE TEXIUM CONNECTOR TO THE WHITE HUB ON THE PRIMARY TEXIUM TUBING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23514005 ↗
  • MalfunctionNov 10, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD ALARIS CLOSED MALE LUER WAS LEAKING. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. IT WAS REPORTED BY THE CUSTOMER THAT GREEN TEXIUM TIP LEAKING AT THE CONNECTION SITE OF THE TUBING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23513868 ↗
  • MalfunctionNov 10, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD ALARIS CLOSED MALE LUER WAS LEAKING. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM IT WAS REPORTED BY THE CUSTOMER THAT THE TEXIUM TIP OF THE SECONDARY TUBING STARTING TO LEAK.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23513629 ↗
  • MalfunctionNov 4, 2025Break; Break

    IT WAS REPORTED THAT THE BD ALARIS TEXIUM CLOSED MALE LUER WAS DAMAGED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: TEXIUM BROKE OFF. TEXIUM CLOSED MALE LUER LOT (10)25025321 EXP 02-12-2028.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23468837 ↗
  • MalfunctionOct 21, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT BD TEXIUM CLOSED MALE LUER CAP WAS LEAKING. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM IT WAS REPORTED BY CUSTOMER THAT THIS WAS A STAND ALONE TEXIUM OR WAS THIS A BONDED SET.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23349138 ↗
  • MalfunctionOct 21, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD TEXIUM CLOSED MALE LUER SET WAS LEAKING. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. IT WAS REPORTED BY THE CUSTOMER THAT THE BLUE ADAPTOR CAP WAS CRACKED AND LEAKING CHEMOTHERAPY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23347794 ↗
  • MalfunctionOct 19, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD TEXIUM CLOSED MALE LUER WAS LEAKING . THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. IT WAS REPORTED BY THE CUSTOMER THAT LEAKING FROM TIP DURING PREPARATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23334002 ↗
  • MalfunctionOct 19, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD TEXIUM CLOSED MALE LUER CAP WAS LEAKING. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING: IT WAS REPORTED BY CUSTOMER THAT WHILE DOXIL WAS INFUSING PT NOTICED THERE WAS DOXIL LEAKING ON TOP OF THE PUMP AND SOME ON THE FLOOR. THE LEAK OCCURED AT THE Y SITE BY THE TEXIUM HUB.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23334001 ↗
  • MalfunctionOct 19, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD ALARIS TEXIUM CLOSED MALE LUER WAS LEAKING. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING: IT WAS REPORTED BY CUSTOMER THAT THEY NOTICED THERE WERE 5-6 DROPS ON THE FLOOR AND THERE WAS LEAKAGE AROUND THE TEXIUM THAT WAS CONNECTED TO THE PRIMARY TUBING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23333976 ↗
  • MalfunctionOct 19, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD TEXIUM CLOSED MALE LUER WAS LEAKING. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM: IT WAS REPORTED BY CUSTOMER THAT THE PATIENT'S 3RD DESENSITIZATION BAG OF OXALIPLATIN WAS LEAKING FROM THE TEXIUM CLOSED MALE LUER CAP. THE INFUSION WAS ALREADY FINISHING UP WHEN THIS WAS SEEN, INFUSION WAS STOPPED. THE NEXT BAG OF OXALIPLATIN WAS HUNG AND IT WAS NOTICED RIGHT AWAY THAT THE TEXIUM CAP WAS LEAKING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23333975 ↗
  • MalfunctionOct 19, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD TEXIUM CLOSED MALE LUER CAP WAS LEAKING. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM. IT WAS REPORTED BY CUSTOMER THAT THEY WENT TO DISCONNECT PATIENT FROM INFUSION AND NOTICED SOME LIQUID ON THE FLOOR. DETERMINED IT WAS THE BENDEKA CHEMOTHERAPY INFUSION THAT WAS LEAKING FROM THE SECONDARY TUBING AT THE TEXIUM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23333974 ↗
  • MalfunctionOct 19, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD TEXIUM MALE LUER CAP WAS LEAKING. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM: IT WAS REPORTED BY THE CUSTOMER THAT A COUPLE DROPS OF LIQUID ON THE FLOOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23333944 ↗
  • MalfunctionOct 19, 2025Leak/Splash; Leak/Splash

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23333715 ↗
  • MalfunctionOct 19, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT BD ALARIS CLOSED MALE LUER WAS LEAKING. THE FOLLOWING INFORMATION WAS RECEIVED BY THE INITIAL REPORTER WITH THE FOLLOWING VERBATIM: IT WAS REPORTED BY CUSTOMER THAT ITEM IS LEAKING.

    Patient problems: Insufficient Information

    MAUDE report 23333606 ↗
  • MalfunctionOct 19, 2025Leak/Splash; Leak/Splash

    NO ADDITIONAL INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23333602 ↗
  • MalfunctionOct 15, 2025Connection Problem; Connection Problem

    EVENT DETAILS: CUSTOMER IS STATING THAT TEXIUM CONNECTOR SKU# 10012241-0500 IS NOT LOCKING ONTO THEIR BD SYRINGES WITH A STRONG, POSITIVE FEEL. BASICALLY, THE CONNECTOR CAN TOO EASILY BE DISCONNECTED. THEY ARE CONCERNED THAT THIS COULD LEAD TO LEAKING AND EXPOSURE RISKS FOR STAFF AND PATIENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23304389 ↗
  • MalfunctionOct 15, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE BD TEXIUM CLOSED MALE LUER WITH FEMALE CAP HAD LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: UMC LUBBOCK HAS EXPERIENCED SEVERAL TEXIUM LEAKS - STANDALONE AND BONDED - WHILE INFUSING CHEMO / HD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23296429 ↗
  • MalfunctionOct 15, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE BD TEXIUM CLOSED MALE LUER WITH FEMALE CAP HAD LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: (B)(6) HAS EXPERIENCED SEVERAL TEXIUM LEAKS STANDALONE AND BONDED WHILE INFUSING CHEMO / HD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23295690 ↗
  • MalfunctionOct 15, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE BD TEXIUM CLOSED MALE LUER WITH FEMALE CAP HAD LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: UMC LUBBOCK HAS EXPERIENCED SEVERAL TEXIUM LEAKS - STANDALONE AND BONDED - WHILE INFUSING CHEMO / HD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23295248 ↗
  • MalfunctionOct 15, 2025Fluid/Blood Leak; Fluid/Blood Leak

    IT WAS REPORTED THAT THE BD TEXIUM CLOSED MALE LUER WITH FEMALE CAP HAD LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: UMC LUBBOCK HAS EXPERIENCED SEVERAL TEXIUM LEAKS - STANDALONE AND BONDED - WHILE INFUSING CHEMO / HD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23294908 ↗
  • MalfunctionOct 15, 2025Leak/Splash; Leak/Splash

    IT WAS REPORTED THAT THE BD CLOSED MALE LUER HAD LEAKAGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: UMC LUBBOCK HAS EXPERIENCED SEVERAL TEXIUM LEAKS - STANDALONE AND BONDED - WHILE INFUSING CHEMO / HD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23287407 ↗

Other Closed antineoplastic and hazardous drug reconstitution and transfer system devices

All Closed antineoplastic and hazardous drug reconstitution and transfer system devices →