Skip to content

PeritX 50-9210Recalls & reported problems

Carefusion 2200, Inc · PeritX Drainage Bag, 2000 mL · Class 2

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10885403222726), received between Apr 3, 2026 and Apr 11, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fluid/Blood Leak2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 11, 2026Fluid/Blood Leak

    IT WAS REPORTED BY THE CUSTOMER ON MARCH 11, 2026, THAT THE PERITX DRAINAGE BAG WAS LEAKING DURING USE, DESPITE CONFIRMATION THAT BOTH THE DRAINAGE END AND THE CLAMP WERE SECURELY CLOSED. DURING FOLLOW UP INFORMATION RECEIVED ON MARCH 29, 2026, IT WAS FURTHER REPORTED THAT THERE WAS NO DIRECT PATIENT OR HEALTHCARE PROVIDER INJURY. THE SOURCE OF THE LEAK COULD NOT BE CLEARLY IDENTIFIED. THE REPORTER STATED THAT AFTER ENSURING THE DRAINAGE CAP WAS TWISTED CLOSED, APPROXIMATELY 1000 ML OF FLUID WAS COLLECTED, THE DEVICE WAS DISCONNECTED FROM THE PATIENT, THE ROLLER CLAMP WAS CLOSED, AND THE PLEURX BAG WAS DISCARDED INTO A GARBAGE BAG PLACED ON THE FLOOR. SHORTLY THEREAFTER, AN ESTIMATED 500 ML OF FLUID LEAKED ONTO THE FLOOR. DUE TO THE VOLUME OF FLUID AROUND THE BAG, THE EXACT SOURCE OF THE LEAK COULD NOT BE VISUALLY DETERMINED. AS THE DRAINAGE CAP AND ROLLER CLAMP HAD BEEN CONFIRMED CLOSED, THE LEAK WAS SUSPECTED TO HAVE ORIGINATED FROM THE BAG ITSELF OR POSSIBLY FROM THE CONNECTION BETWEEN THE BAG AND THE TUBING. THE CUSTOMER STATED THAT THEY HAD HEARD OF SEVERAL OTHER INSTANCES OF BAGS LEAKING IN THE PAST MONTH. DURING FOLLOW UP, THE CUSTOMER INDICATED THEY WERE NOT AWARE OF ANY PR

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24848768 ↗
  • MalfunctionApr 3, 2026Fluid/Blood Leak

    IT WAS REPORTED BY THE CUSTOMER ON (B)(6) 2026, THAT THE PERITX DRAINAGE BAG WAS LEAKING DURING USE, DESPITE CONFIRMATION THAT BOTH THE DRAINAGE END AND THE CLAMP WERE SECURELY CLOSED. DURING FOLLOW UP INFORMATION RECEIVED ON MARCH 29, 2026, IT WAS FURTHER REPORTED THAT THERE WAS NO DIRECT PATIENT OR HEALTHCARE PROVIDER INJURY. THE SOURCE OF THE LEAK COULD NOT BE CLEARLY IDENTIFIED. THE REPORTER STATED THAT AFTER ENSURING THE DRAINAGE CAP WAS TWISTED CLOSED, APPROXIMATELY 1000 ML OF FLUID WAS COLLECTED, THE DEVICE WAS DISCONNECTED FROM THE PATIENT, THE ROLLER CLAMP WAS CLOSED, AND THE PLEURX BAG WAS DISCARDED INTO A GARBAGE BAG PLACED ON THE FLOOR. SHORTLY THEREAFTER, AN ESTIMATED 500 ML OF FLUID LEAKED ONTO THE FLOOR. DUE TO THE VOLUME OF FLUID AROUND THE BAG, THE EXACT SOURCE OF THE LEAK COULD NOT BE VISUALLY DETERMINED. AS THE DRAINAGE CAP AND ROLLER CLAMP HAD BEEN CONFIRMED CLOSED, THE LEAK WAS SUSPECTED TO HAVE ORIGINATED FROM THE BAG ITSELF OR POSSIBLY FROM THE CONNECTION BETWEEN THE BAG AND THE TUBING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24779862 ↗

Other Peritoneal, drainage catheter for refractory ascites, long-term indwelling devices

All Peritoneal, drainage catheter for refractory ascites, long-term indwelling devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.