BD Pyxis 324 (UDI 10885403533235)Recalls & reported problems
Carefusion 303, Inc. · BD Pyxis MedStation ES (Auxiliary) · Class 1
Quick summary
This BD Pyxis MedStation auxiliary unit is a medical cabinet. Reports mostly describe mechanical and circuit failures, loose connections, broken parts, and drawers or locks that stop working. There are 7,888 reports, with no deaths and 1 injury. No recalls name this device. Reports rose to 7,027 in the last 12 months from 861 in the 12 months before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 8,624
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8,624 adverse event reports that list this exact device (DI 10885403533235), received between Jan 6, 2025 and Sep 30, 2026 — classified as 1 injury, 8,616 malfunctions.
The most frequently reported specific device problem is “Mechanical Problem” (2,115 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Mechanical Problem2,115
- Loose or Intermittent Connection2,070
- Circuit Failure2,054
- Break1,396
- Component Missing624
- Contamination610
- Electrical /Electronic Property Problem482
- Activation Failure476
Patient problems
- No Clinical Signs, Symptoms or Conditions8,621
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available5
- Insufficient Information1
Recent adverse event reports
The 25 most recent of 8,624 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Loose or Intermittent Connection; Installation-Related Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50953948 ↗MalfunctionSep 30, 2026Component Missing
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50953946 ↗MalfunctionSep 30, 2026Break; Mechanical Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50953941 ↗MalfunctionSep 30, 2026Break; Mechanical Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50953924 ↗MalfunctionSep 30, 2026Mechanical Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50953916 ↗MalfunctionSep 30, 2026Electrical /Electronic Property Problem; Loose or Intermittent Connection
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50953914 ↗MalfunctionSep 30, 2026Loose or Intermittent Connection
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50952590 ↗MalfunctionSep 30, 2026Thermal Decomposition of Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50946186 ↗MalfunctionSep 30, 2026Break
It was reported that when using the BD Pyxis¿ MedStation¿ ES Auxiliary MCU-A drawer 5 pockets B1 and D3 experienced failures and could not be recovered. The user attempted troubleshooting, including rebooting the device; however, both pockets remained in a failed state and recovery was unsuccessful.The customer stated that this malfunction occurred while dispensing the medication which caused a delay to the patient care. There were no adverse events or injuries reported based on this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50944975 ↗MalfunctionSep 30, 2026Mechanical Problem
It was reported that when using the BD Pyxis¿ MedStation¿ ES Auxiliary, system auxiliary drawer 7 was failed. The customer stated that they tried to reboot and recover but the issue was not resolved.The customer stated that there was a delay in patient care.There were no adverse events or injuries reported based on this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50944972 ↗MalfunctionSep 30, 2026Electrical /Electronic Property Problem
It was reported that when using the BD Pyxis¿ MedStation¿ ES Auxiliary system cubie was failed and was unable to release. The customer stated that this malfunction caused a delay in dispensing medication to patients. However, there were no adverse events or injuries reported based on this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50944953 ↗MalfunctionSep 30, 2026Component Missing
It was reported that when using the BD Pyxis¿ MedStation¿ ES Auxiliary, a drawer failure occurred involving a full-height drawer that failed to stay closed.The customer stated that there was a delay in patient care.There were no adverse events or injuries reported based on this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50944938 ↗MalfunctionSep 30, 2026Loose or Intermittent Connection
It was reported that when using the BD Pyxis¿ MedStation¿ ES Auxiliary drawer 7 was failed and unable to recover.The customer stated that there was a delay in patient care.There were no adverse events or injuries reported based on this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50944928 ↗MalfunctionSep 30, 2026Circuit Failure; Loose or Intermittent Connection
It was reported that when using the BD Pyxis¿ MedStation¿ ES Auxiliary drawer was failed. Customer tried to reboot and recover, but issue doesn't resolve. The customer stated that this malfunction caused a delay in dispensing medication to patients. However, there were no adverse events or injuries reported based on this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50944925 ↗MalfunctionSep 30, 2026Component Missing
It was reported that when using the BD Pyxis¿ MedStation¿ ES Auxiliary drawer was failed and unable to recover.The customer stated that this malfunction occurred while dispensing the medication and they were unable to access/issue the medication, which caused a delay to the patient care.There were no adverse events or injuries reported based on this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50944921 ↗MalfunctionSep 30, 2026Contamination
It was reported that when using the BD Pyxis¿ MedStation¿ ES Auxiliary drawer 5 was not detected on bus.As per part investigation, it was determined that the complaint was attributed to visual anomalies indicative of fluid ingress detected on the board. However, there were no delays or adverse events or injuries reported based on this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50944382 ↗MalfunctionSep 30, 2026Circuit Failure; Contamination
It was reported that when using the BD Pyxis¿ MedStation¿ ES Auxiliary, the drawer failed.As per part investigation, it was determined that the fluid ingress condition of the part module controller.There were no adverse events or injuries reported based on this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50942961 ↗MalfunctionSep 30, 2026Break
It was reported that when using the BD Pyxis¿ MedStation¿ ES Auxiliary, the drawer 2 was failed and displayed a device not detected on bus error. Reboot and recovery attempts were performed, but the issue persisted. The customer stated that this malfunction caused a delay in dispensing medication to patients. However, there were no adverse events or injuries reported based on this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50942731 ↗MalfunctionSep 30, 2026Labelling, Instructions for Use or Training Problem
It was reported that when using the BD Pyxis¿ MedStation¿ ES Auxiliary, a cached credential issue occurred. Following a major technology outage, users at EHSC were unable to log in to the Pyxis cabinets after the Ethernet connection was disconnected. The issue prevented access to the cabinets and ultimately required staff to open the back of the units to manually access cubies. The customer reported that other sites within the region were able to log in without issue, indicating the problem was isolated to the affected location.There were no delay or adverse events or injuries reported based on this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50942539 ↗MalfunctionSep 30, 2026Component Missing
It was reported that when using the BD Pyxis¿ MedStation¿ ES Auxiliary VGH4SP1 drawer 6.2 would not open automatically and required users to manually pull the drawer open. Additionally, staff reported difficulties closing the drawer, indicating intermittent drawer operation issues.The customer stated that this malfunction occurred while dispensing the medication and they were unable to access the medication, which caused a delay to the patient care. There were no adverse events or injuries reported based on this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50942482 ↗MalfunctionSep 29, 2026Circuit Failure; Mechanical Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50933764 ↗MalfunctionSep 29, 2026Circuit Failure; Mechanical Problem
It was reported that when using the BD Pyxis¿ MedStation¿ ES Auxiliary drawer was failed and required maintenance. The customer unplugged and reconnected the power cable, attempted a reboot, and performed a recovery; however, the issue persisted.The customer stated that there was a delay in patient care.There were no adverse events or injuries reported based on this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50933759 ↗MalfunctionSep 29, 2026Contamination; Mechanical Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50932760 ↗MalfunctionSep 29, 2026Contamination
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50932713 ↗MalfunctionSep 29, 2026Break
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50932667 ↗
Other Cabinet, table and tray, anesthesia devices
- BD Pyxis 323Carefusion 303, Inc. · Cabinet, table and tray, anesthesia2 deaths28,053 reports
- BD Pyxis 323Carefusion 303, Inc. · CABINET, TABLE AND TRAY, ANESTHESIA14 recalls2 deaths24,375 reports
- BD Pyxis 324Carefusion 303, Inc. · CABINET, TABLE AND TRAY, ANESTHESIA5 recalls1 death5,301 reports
- BD Pyxis 343Carefusion 303, Inc. · CABINET, TABLE AND TRAY, ANESTHESIA2 recalls1 death5,196 reports
- Pyxis 327Carefusion 303, Inc. · Cabinet, table and tray, anesthesia1 death4,304 reports
- BD Pyxis MEDBANK-TOWERCarefusion 303, Inc. · CABINET, TABLE AND TRAY, ANESTHESIA30 recalls2,581 reports
- Pyxis 327Carefusion 303, Inc. · CABINET, TABLE AND TRAY, ANESTHESIA10 recalls3,137 reports
- BD Pyxis 1145-00Carefusion 303, Inc. · CABINET, TABLE AND TRAY, ANESTHESIA11 recalls1,705 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.