Optetrak Logic 02-012-35-2509Recalls & reported problems
Exactech, Inc. · Class 2 · Implant
All models of Optetrak Logic (408)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Exactech knee implant is a cemented, semi-constrained replacement for the kneecap and knee joint surfaces. The 36 reports mostly describe gradual wear of the plastic insert and femoral component years after surgery, along with loosening, bone loss (osteolysis), pain, swelling, and joint inflammation, often leading to revision surgery;35 reports describe injuries and no deaths. One open recall, started in August 2021, concerns inserts packaged in vacuum bags without an extra oxygen barrier layer.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 36
- Deaths reported
- 0
- Injuries reported
- 35
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 36 adverse event reports that list this exact device (DI 10885862001764), received between Dec 4, 2020 and Aug 4, 2026 — classified as 35 injuries, 1 malfunction.
The most frequently reported specific device problem is “Naturally Worn” (30 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K093360); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Aug 30, 2021Class IIOpen, Classified
Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.
Z-0021-2022 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Naturally Worn30
- Unstable7
- Insufficient Device Problem Information6
- Loosening of Implant Not Related to Bone-Ingrowth5
- Adverse Event Without Identified Device or Use Problem4
- Fracture2
- Mechanical Problem2
- Break1
Patient problems
- Failure of Implant21
- Pain16
- Insufficient Information12
- Osteolysis12
- Synovitis9
- Swelling/ Edema8
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available6
- Implant Pain6
Recent adverse event reports
The 25 most recent of 36 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 4, 2026Insufficient Device Problem Information
Patient problems: Insufficient Information
MAUDE report 50129791 ↗InjuryMay 7, 2026Naturally Worn
AS REPORTED, APPROXIMATELY 7 YEARS AND 10 MONTHS POST THE INITIAL RIGHT TOTAL KNEE ARTHROPLASTY, THE PATIENT'S TIBIAL AND PATELLAR POLYS WERE REVISED DUE TO WEAR. THE PATELLA POLY SHOWED MORE SIGNIFICANT WEAR, WITH MILD OSTEOLYSIS OF THE BONE. NO BONE GRAFT REQUIRED. THERE WAS NO REPORTED BREAKAGE OF A DEVICE OR SURGICAL DELAY/PROLONGATION. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.
Patient problems: Osteolysis
MAUDE report 25110894 ↗InjuryNov 21, 2025Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL TOTAL KNEE REPLACEMENT ON THE LEFT SIDE. SUBSEQUENTLY, THE PATIENT EXPERIENCED FEMORAL KNEE DEBONDING/LOOSENING AND BROKEN POLYETHYLENE LINER. AS A RESULT, APPROXIMATELY 10 YEARS AND 7 MONTHS POST-SURGERY, THE PATIENT UNDERWENT A REVISION. NO FURTHER IMPACT TO THE PATIENT WAS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS IS REPORT 2 OF 2 FOR THIS EVENT/PATIENT.
Patient problems: Insufficient Information
MAUDE report 23622342 ↗InjuryAug 28, 2025Unstable; Unstable
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL LEFT TOTAL KNEE ARTHROPLASTY. SUBSEQUENTLY, IT WAS REPORTED THAT THE PATIENT EXPERIENCED PAIN AND INSTABILITY. AS A RESULT, THE PATIENT UNDERWENT A LEFT KNEE REVISION TO EXCHANGE THE POLYETHYLENE LINER APPROXIMATELY 12 YEARS AFTER INITIAL IMPLANTATION. NO FURTHER PATIENT IMPACT REPORTED. NO FURTHER INFORMATION AVAILABLE.
Patient problems: Joint Laxity; Pain
MAUDE report 22917224 ↗InjuryJan 22, 2025Unstable; Naturally Worn; Unstable; Naturally Worn
IT WAS REPORTED THAT APPROXIMATELY 96 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, OSTEOLYSIS, PAIN, SWELLING/EDEMA, AND SYNOVITIS. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant; Osteolysis; Pain; Swelling/ Edema; Synovitis
MAUDE report 21210011 ↗InjurySep 16, 2024Naturally Worn; Naturally Worn
IT WAS REPORTED THAT APPROXIMATELY 83 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, LEFT KNEE PAIN; OSTEOLYSIS, LOOSENING OF FEMUR AND TIBIA, POLYETHYLENE WEAR ON THE TIBIAL INSERT. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Osteolysis; Pain
MAUDE report 20223452 ↗InjuryAug 28, 2024Naturally Worn; Naturally Worn
IT WAS REPORTED THAT APPROXIMATELY 163 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, ASEPTIC LOOSENING, POLYETHYLENE WEAR AND OSTEOLYSIS.. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Aseptic loosening; Joint effusion; Osteolysis; Synovitis
MAUDE report 20084181 ↗InjuryAug 27, 2024Naturally Worn; Naturally Worn
IT WAS REPORTED THAT APPROXIMATELY 133 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, PAIN, SWELLING. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Cyst(s); Pain; Swelling/ Edema
MAUDE report 20075206 ↗InjuryAug 21, 2024Naturally Worn; Insufficient Device Problem Information; Naturally Worn; Insufficient Device Problem Information
IT WAS REPORTED THAT APPROXIMATELY 116 MONTHS AFTER A RIGHT KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Insufficient Information
MAUDE report 20017815 ↗InjuryAug 20, 2024Naturally Worn; Naturally Worn
IT WAS REPORTED VIA LEGAL DOCUMENTATION APPROXIMATELY 100 MONTHS AFTER A LEFT KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, PAIN AND SUFFERING. A REVISION OP REPORT WAS PROVIDED. POST OPERATIVE DIAGNOSIS NOTED PREMATURE WEAR OF THE POLYETHYLENE LINER AND ASEPTIC LOOSENING OF LEFT TOTAL KNEE ARTHROPLASTY FEMORAL COMPONENT. INTRAOPERATIVELY THE SURGEON OBSERVED SIGNIFICANT WEAR TO THE LINER AND A LOOSE FEMORAL COMPONENT. IT WAS NOTED THE PATELLAR AND TIBIAL COMPONENTS WERE WELL FIXED WITHOUT ANY SIGNS OF SIGNIFICANT OSTEOLYSIS OR LOOSENING SURROUNDING THE IMPLANTS. THE LINER AND FEMORAL COMPONENTS WERE REVISED. NO MEDICAL OR SURGICAL INTERVENTION WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Aseptic loosening; Discomfort; Failure of Implant; Osteolysis; Pain; Synovitis
MAUDE report 20008433 ↗InjuryAug 13, 2024Fracture; Unstable; Naturally Worn; Fracture; Unstable; Naturally Worn
IT WAS REPORTED THAT APPROXIMATELY 137 MONTHS AFTER A TOTAL LEFT KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, CHRONIC SYNOVITIS, INSTABILITY. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant; Joint Laxity; Joint effusion; Synovitis
MAUDE report 19977185 ↗InjuryAug 8, 2024Mechanical Problem; Naturally Worn; Mechanical Problem; Naturally Worn
LEGAL CASE ¿ USA (MDL NO. 3044) (NGG) (MMH). PATIENT ID: IRIS SHOOTS RK 1ST REV. REFER TO (B)(4) RK 2ND REV. 8. PLAINTIFF WAS IMPLANTED WITH THE FOLLOWING EXACTECH DEVICE: KNEE. 9. LEG IN WHICH THE EXACTECH DEVICE WAS IMPLANTED: RIGHT. 10. DATE THE EXACTECH DEVICE WAS IMPLANTED: (B)(6) 2016. 11. STATE IN WHICH THE EXACTECH DEVICE WAS IMPLANTED: (B)(6). 12. DATE THE EXACTECH DEVICE WAS SURGICALLY REMOVED/REVISED: 2018. 13. ON (B)(6) 2016, PLAINTIFF UNDERWENT A RIGHT TOTAL KNEE ARTHROPLASTY IN WHICH AN EXACTECH OPTETREK LOGIC DEVICE WAS IMPLANTED. IN 2018, PLAINTIFF UNDERWENT A RIGHT TOTAL KNEE ARTHROPLASTY REVISION. PLAINTIFF SUFFERED FROM PAIN, SWELLING, AND DECREASED RANGE OF MOTION. THE IMPLANTATION OF THE EXACTECH DEVICE AND SUBSEQUENT REVISIONS HAVE CAUSED PERSONAL INJURIES, SEVERE PAIN AND SUFFERING, MENTAL ANGUISH, EMOTIONAL DISTRESS, FEAR, LOSS OF ENJOYMENT OF LIFE, MEDICAL EXPENSES, AND FINANCIAL LOSSES, AND SHE WILL SUSTAIN FUTURE DAMAGES INCLUDING BUT NOT LIMITED TO COST OF MEDICAL CARE, REHABILITATION, MENTAL AND EMOTIONAL DISTRESS, AND PAIN AND SUFFERING. ACCORDINGLY, PLAINTIFF IS SEEKING COMPENSATORY AND SPECIAL DAMAGES, INCLUDING ATTORNEYS' FEES AND COSTS, AND ALL OTH
Patient problems: Distress; Emotional Changes; Loss of Range of Motion; Pain; Swelling/ Edema
MAUDE report 19938813 ↗InjuryJun 18, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 148 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS POLYETHYLENE WEAR AND LOOSENING. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. POST OPERATIVE DIAGNOSIS NOTED OSTEOLYSIS OF BOTH MEDIAL AND LATERAL POSTERIOR CONDYLES AS WELL AS THE LATERAL DISTAL CONDYLE. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Joint effusion; Osteolysis; Pain
MAUDE report 19561247 ↗InjuryJun 6, 2024Unstable; Naturally Worn; Unstable; Naturally Worn
IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 122 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, BONE LOSS, PATELLAR COMPONENT WEAR, PAIN, SWELLING, STIFFNESS, TISSUE DAMAGE, INSTABILITY, OSTEOLYSIS, FEMORAL LOOSENING, AND SYNOVITIS. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure of Implant; Implant Pain; Joint Laxity; Osteolysis; Swelling/ Edema; Synovitis
MAUDE report 19476444 ↗InjuryMay 21, 2024Naturally Worn; Naturally Worn
IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 100 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, JOINT PAIN, JOINT SWELLING AND STIFFNESS. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure of Implant; Implant Pain; Swelling/ Edema
MAUDE report 19360026 ↗InjuryMay 20, 2024Naturally Worn; Naturally Worn
IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT A PATIENT HAD A LEFT KNEE ARTHROPLASTY ON (B)(6) 2014, AND THEN EXPERIENCED A REVISION SURGICAL PROCEDURE ON (B)(6) 2022, APPROXIMATELY 7 YEARS, 11 MONTHS AFTER INITIAL IMPLANT. THERE WAS NO OTHER PATIENT/MEDICAL INFORMATION PROVIDED. NO X-RAYS OR IMAGES WERE PROVIDED. THE DEVICE WILL NOT BE RETURNED. THERE IS NO OTHER INFORMATION AVAILABLE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Discomfort; Failure of Implant; Pain
MAUDE report 19353229 ↗InjuryMay 16, 2024Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth; Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 148 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS POLYETHYLENE WEAR AND LOOSENING. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. POST OPERATIVE DIAGNOSIS NOTED OSTEOLYSIS OF BOTH MEDIAL AND LATERAL POSTERIOR CONDYLES AS WELL AS THE LATERAL DISTAL CONDYLE. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant; Joint effusion; Osteolysis; Pain
MAUDE report 19332529 ↗InjuryMay 16, 2024Naturally Worn; Naturally Worn
IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT A PATIENT HAD A RIGHT KNEE ARTHROPLASTY ON (B(6) 2016, WITH NO REVISION SURGERY REPORTED. THERE WAS NO OTHER PATIENT/MEDICAL INFORMATION PROVIDED. NO X-RAYS OR IMAGES WERE PROVIDED. THERE IS NO OTHER INFORMATION AVAILABLE.
Patient problems: Failure of Implant; Insufficient Information
MAUDE report 19331288 ↗InjuryMay 15, 2024Naturally Worn; Naturally Worn
IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT A PATIENT HAD A LEFT KNEE ARTHROPLASTY ON (B)(6) 2016, WITH NO REVISION SURGERY REPORTED. THERE WAS NO OTHER PATIENT/MEDICAL INFORMATION PROVIDED. NO X-RAYS OR IMAGES WERE PROVIDED. THERE IS NO OTHER INFORMATION AVAILABLE.
Patient problems: Failure of Implant; Insufficient Information
MAUDE report 19327194 ↗InjuryMay 15, 2024Corroded; Fracture; Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth; Corroded; Fracture; Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 126 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, PROGRESSIVE PAIN AND SWELLING, JOINT DAMAGE, OSTEOLYSIS UNDER FEMORAL COMPONENT, OSTEOLYSIS WITH LYTIC LESIONS ON TIBIA, POLY OXIDATION AND BREAKAGE. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure of Implant; Implant Pain; Osteolysis; Swelling/ Edema
MAUDE report 19325765 ↗InjuryMay 15, 2024Crack; Naturally Worn; Crack; Naturally Worn
IT WAS REPORTED THAT APPROXIMATELY 147 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS POLYETHYLENE WEAR AND SYNOVITIS. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant; Synovitis
MAUDE report 19324685 ↗InjuryMay 10, 2024Naturally Worn; Naturally Worn
IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 78 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS POLYETHYLENE WEAR, PAIN, JOINT LAXITY, PITTING OF THE TIBIAL INSERT, POLYETHYLENE WEAR OF THE PATELLAR COMPONENT, OSTEOLYSIS OF THE TIBIA AND FEMUR, LOOSE FEMORAL COMPONENT, AND LIMITED RANGE OF MOTION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant; Implant Pain; Joint Laxity; Loss of Range of Motion; Osteolysis
MAUDE report 19285820 ↗InjuryMay 7, 2024Naturally Worn; Naturally Worn
IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 87 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT MAY REQUIRE A REVISION SURGERY FOR POLYETHYLENE WEAR AND OTHER UNSPECIFIED ISSUES. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant; Insufficient Information
MAUDE report 19260020 ↗InjuryApr 12, 2024Naturally Worn; Naturally Worn
IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT A PATIENT HAD A LEFT TOTAL KNEE ARTHROPLASTY ON (B)(6) 2014 AND THEN EXPERIENCED A REVISION SURGICAL PROCEDURE ON (B)(6) 2023 APPROXIMATELY 8 YEARS AND 8 MONTHS AFTER INITIAL IMPLANT. NO OTHER PATIENT INFORMATION / MEDICAL HISTORY REPORTED. NO IMAGES OF THE DEVICES ARE PROVIDED. THE DEVICE WILL NOT BE RETURNED. THERE IS NO OTHER INFORMATION AVAILABLE.
Patient problems: Failure of Implant; Insufficient Information
MAUDE report 19096791 ↗InjuryDec 6, 2023Naturally Worn; Naturally Worn
AS REPORTED, THE PATIENT HAD AN INITIAL RIGHT TKA ON (B)(6) 2019. THE PATIENT WAS REVISED ON (B)(6) 2023 DUE TO POLY WEAR AND THE LINER WAS EXCHANGED. THERE WAS NO REPORTED BREAKAGE OF A DEVICE OR SURGICAL DELAY/PROLONGATION. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.
Patient problems: Failure of Implant; Insufficient Information
MAUDE report 18280074 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K093360. One submission often covers many models, so check the product and lot details in each record.
- Z-2310-2012 · Aug 19, 2011 · TerminatedOPTETRAK LOGIC TIBIAL INSERT POSTERIOR STABILIZED***Match Component Colors 6 YELLOW, 11mm, Use with LOGIC Femoral and Tibial Components***REF 02-012-35-6011. There are three produ
Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page