Skip to content

Optetrak Logic 02-012-35-2509Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Optetrak Logic (408)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Exactech knee implant is a cemented, semi-constrained replacement for the kneecap and knee joint surfaces. The 36 reports mostly describe gradual wear of the plastic insert and femoral component years after surgery, along with loosening, bone loss (osteolysis), pain, swelling, and joint inflammation, often leading to revision surgery;35 reports describe injuries and no deaths. One open recall, started in August 2021, concerns inserts packaged in vacuum bags without an extra oxygen barrier layer.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
36
Deaths reported
0
Injuries reported
35
Recalls naming it
1

Summary

The FDA's MAUDE database contains 36 adverse event reports that list this exact device (DI 10885862001764), received between Dec 4, 2020 and Aug 4, 2026 — classified as 35 injuries, 1 malfunction.

The most frequently reported specific device problem is “Naturally Worn” (30 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K093360); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 9 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 6 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520269
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 30, 2021Class IIOpen, Classified

    Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

    Z-0021-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Naturally Worn30
  • Unstable7
  • Insufficient Device Problem Information6
  • Loosening of Implant Not Related to Bone-Ingrowth5
  • Adverse Event Without Identified Device or Use Problem4
  • Fracture2
  • Mechanical Problem2
  • Break1

Patient problems

  • Failure of Implant21
  • Pain16
  • Insufficient Information12
  • Osteolysis12
  • Synovitis9
  • Swelling/ Edema8
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available6
  • Implant Pain6

Recent adverse event reports

The 25 most recent of 36 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 4, 2026Insufficient Device Problem Information

    Patient problems: Insufficient Information

    MAUDE report 50129791 ↗
  • InjuryMay 7, 2026Naturally Worn

    AS REPORTED, APPROXIMATELY 7 YEARS AND 10 MONTHS POST THE INITIAL RIGHT TOTAL KNEE ARTHROPLASTY, THE PATIENT'S TIBIAL AND PATELLAR POLYS WERE REVISED DUE TO WEAR. THE PATELLA POLY SHOWED MORE SIGNIFICANT WEAR, WITH MILD OSTEOLYSIS OF THE BONE. NO BONE GRAFT REQUIRED. THERE WAS NO REPORTED BREAKAGE OF A DEVICE OR SURGICAL DELAY/PROLONGATION. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.

    Patient problems: Osteolysis

    MAUDE report 25110894 ↗
  • InjuryNov 21, 2025Break; Break

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL TOTAL KNEE REPLACEMENT ON THE LEFT SIDE. SUBSEQUENTLY, THE PATIENT EXPERIENCED FEMORAL KNEE DEBONDING/LOOSENING AND BROKEN POLYETHYLENE LINER. AS A RESULT, APPROXIMATELY 10 YEARS AND 7 MONTHS POST-SURGERY, THE PATIENT UNDERWENT A REVISION. NO FURTHER IMPACT TO THE PATIENT WAS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS IS REPORT 2 OF 2 FOR THIS EVENT/PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 23622342 ↗
  • InjuryAug 28, 2025Unstable; Unstable

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL LEFT TOTAL KNEE ARTHROPLASTY. SUBSEQUENTLY, IT WAS REPORTED THAT THE PATIENT EXPERIENCED PAIN AND INSTABILITY. AS A RESULT, THE PATIENT UNDERWENT A LEFT KNEE REVISION TO EXCHANGE THE POLYETHYLENE LINER APPROXIMATELY 12 YEARS AFTER INITIAL IMPLANTATION. NO FURTHER PATIENT IMPACT REPORTED. NO FURTHER INFORMATION AVAILABLE.

    Patient problems: Joint Laxity; Pain

    MAUDE report 22917224 ↗
  • InjuryJan 22, 2025Unstable; Naturally Worn; Unstable; Naturally Worn

    IT WAS REPORTED THAT APPROXIMATELY 96 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, OSTEOLYSIS, PAIN, SWELLING/EDEMA, AND SYNOVITIS. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant; Osteolysis; Pain; Swelling/ Edema; Synovitis

    MAUDE report 21210011 ↗
  • InjurySep 16, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT APPROXIMATELY 83 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, LEFT KNEE PAIN; OSTEOLYSIS, LOOSENING OF FEMUR AND TIBIA, POLYETHYLENE WEAR ON THE TIBIAL INSERT. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Osteolysis; Pain

    MAUDE report 20223452 ↗
  • InjuryAug 28, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT APPROXIMATELY 163 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, ASEPTIC LOOSENING, POLYETHYLENE WEAR AND OSTEOLYSIS.. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Aseptic loosening; Joint effusion; Osteolysis; Synovitis

    MAUDE report 20084181 ↗
  • InjuryAug 27, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT APPROXIMATELY 133 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, PAIN, SWELLING. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Cyst(s); Pain; Swelling/ Edema

    MAUDE report 20075206 ↗
  • InjuryAug 21, 2024Naturally Worn; Insufficient Device Problem Information; Naturally Worn; Insufficient Device Problem Information

    IT WAS REPORTED THAT APPROXIMATELY 116 MONTHS AFTER A RIGHT KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 20017815 ↗
  • InjuryAug 20, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION APPROXIMATELY 100 MONTHS AFTER A LEFT KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, PAIN AND SUFFERING. A REVISION OP REPORT WAS PROVIDED. POST OPERATIVE DIAGNOSIS NOTED PREMATURE WEAR OF THE POLYETHYLENE LINER AND ASEPTIC LOOSENING OF LEFT TOTAL KNEE ARTHROPLASTY FEMORAL COMPONENT. INTRAOPERATIVELY THE SURGEON OBSERVED SIGNIFICANT WEAR TO THE LINER AND A LOOSE FEMORAL COMPONENT. IT WAS NOTED THE PATELLAR AND TIBIAL COMPONENTS WERE WELL FIXED WITHOUT ANY SIGNS OF SIGNIFICANT OSTEOLYSIS OR LOOSENING SURROUNDING THE IMPLANTS. THE LINER AND FEMORAL COMPONENTS WERE REVISED. NO MEDICAL OR SURGICAL INTERVENTION WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Aseptic loosening; Discomfort; Failure of Implant; Osteolysis; Pain; Synovitis

    MAUDE report 20008433 ↗
  • InjuryAug 13, 2024Fracture; Unstable; Naturally Worn; Fracture; Unstable; Naturally Worn

    IT WAS REPORTED THAT APPROXIMATELY 137 MONTHS AFTER A TOTAL LEFT KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, CHRONIC SYNOVITIS, INSTABILITY. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant; Joint Laxity; Joint effusion; Synovitis

    MAUDE report 19977185 ↗
  • InjuryAug 8, 2024Mechanical Problem; Naturally Worn; Mechanical Problem; Naturally Worn

    LEGAL CASE ¿ USA (MDL NO. 3044) (NGG) (MMH). PATIENT ID: IRIS SHOOTS RK 1ST REV. REFER TO (B)(4) RK 2ND REV. 8. PLAINTIFF WAS IMPLANTED WITH THE FOLLOWING EXACTECH DEVICE: KNEE. 9. LEG IN WHICH THE EXACTECH DEVICE WAS IMPLANTED: RIGHT. 10. DATE THE EXACTECH DEVICE WAS IMPLANTED: (B)(6) 2016. 11. STATE IN WHICH THE EXACTECH DEVICE WAS IMPLANTED: (B)(6). 12. DATE THE EXACTECH DEVICE WAS SURGICALLY REMOVED/REVISED: 2018. 13. ON (B)(6) 2016, PLAINTIFF UNDERWENT A RIGHT TOTAL KNEE ARTHROPLASTY IN WHICH AN EXACTECH OPTETREK LOGIC DEVICE WAS IMPLANTED. IN 2018, PLAINTIFF UNDERWENT A RIGHT TOTAL KNEE ARTHROPLASTY REVISION. PLAINTIFF SUFFERED FROM PAIN, SWELLING, AND DECREASED RANGE OF MOTION. THE IMPLANTATION OF THE EXACTECH DEVICE AND SUBSEQUENT REVISIONS HAVE CAUSED PERSONAL INJURIES, SEVERE PAIN AND SUFFERING, MENTAL ANGUISH, EMOTIONAL DISTRESS, FEAR, LOSS OF ENJOYMENT OF LIFE, MEDICAL EXPENSES, AND FINANCIAL LOSSES, AND SHE WILL SUSTAIN FUTURE DAMAGES INCLUDING BUT NOT LIMITED TO COST OF MEDICAL CARE, REHABILITATION, MENTAL AND EMOTIONAL DISTRESS, AND PAIN AND SUFFERING. ACCORDINGLY, PLAINTIFF IS SEEKING COMPENSATORY AND SPECIAL DAMAGES, INCLUDING ATTORNEYS' FEES AND COSTS, AND ALL OTH

    Patient problems: Distress; Emotional Changes; Loss of Range of Motion; Pain; Swelling/ Edema

    MAUDE report 19938813 ↗
  • InjuryJun 18, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 148 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS POLYETHYLENE WEAR AND LOOSENING. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. POST OPERATIVE DIAGNOSIS NOTED OSTEOLYSIS OF BOTH MEDIAL AND LATERAL POSTERIOR CONDYLES AS WELL AS THE LATERAL DISTAL CONDYLE. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Joint effusion; Osteolysis; Pain

    MAUDE report 19561247 ↗
  • InjuryJun 6, 2024Unstable; Naturally Worn; Unstable; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 122 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, BONE LOSS, PATELLAR COMPONENT WEAR, PAIN, SWELLING, STIFFNESS, TISSUE DAMAGE, INSTABILITY, OSTEOLYSIS, FEMORAL LOOSENING, AND SYNOVITIS. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure of Implant; Implant Pain; Joint Laxity; Osteolysis; Swelling/ Edema; Synovitis

    MAUDE report 19476444 ↗
  • InjuryMay 21, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 100 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, JOINT PAIN, JOINT SWELLING AND STIFFNESS. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure of Implant; Implant Pain; Swelling/ Edema

    MAUDE report 19360026 ↗
  • InjuryMay 20, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT A PATIENT HAD A LEFT KNEE ARTHROPLASTY ON (B)(6) 2014, AND THEN EXPERIENCED A REVISION SURGICAL PROCEDURE ON (B)(6) 2022, APPROXIMATELY 7 YEARS, 11 MONTHS AFTER INITIAL IMPLANT. THERE WAS NO OTHER PATIENT/MEDICAL INFORMATION PROVIDED. NO X-RAYS OR IMAGES WERE PROVIDED. THE DEVICE WILL NOT BE RETURNED. THERE IS NO OTHER INFORMATION AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Discomfort; Failure of Implant; Pain

    MAUDE report 19353229 ↗
  • InjuryMay 16, 2024Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth; Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 148 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS POLYETHYLENE WEAR AND LOOSENING. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. POST OPERATIVE DIAGNOSIS NOTED OSTEOLYSIS OF BOTH MEDIAL AND LATERAL POSTERIOR CONDYLES AS WELL AS THE LATERAL DISTAL CONDYLE. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant; Joint effusion; Osteolysis; Pain

    MAUDE report 19332529 ↗
  • InjuryMay 16, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT A PATIENT HAD A RIGHT KNEE ARTHROPLASTY ON (B(6) 2016, WITH NO REVISION SURGERY REPORTED. THERE WAS NO OTHER PATIENT/MEDICAL INFORMATION PROVIDED. NO X-RAYS OR IMAGES WERE PROVIDED. THERE IS NO OTHER INFORMATION AVAILABLE.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 19331288 ↗
  • InjuryMay 15, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT A PATIENT HAD A LEFT KNEE ARTHROPLASTY ON (B)(6) 2016, WITH NO REVISION SURGERY REPORTED. THERE WAS NO OTHER PATIENT/MEDICAL INFORMATION PROVIDED. NO X-RAYS OR IMAGES WERE PROVIDED. THERE IS NO OTHER INFORMATION AVAILABLE.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 19327194 ↗
  • InjuryMay 15, 2024Corroded; Fracture; Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth; Corroded; Fracture; Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 126 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, PROGRESSIVE PAIN AND SWELLING, JOINT DAMAGE, OSTEOLYSIS UNDER FEMORAL COMPONENT, OSTEOLYSIS WITH LYTIC LESIONS ON TIBIA, POLY OXIDATION AND BREAKAGE. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure of Implant; Implant Pain; Osteolysis; Swelling/ Edema

    MAUDE report 19325765 ↗
  • InjuryMay 15, 2024Crack; Naturally Worn; Crack; Naturally Worn

    IT WAS REPORTED THAT APPROXIMATELY 147 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS POLYETHYLENE WEAR AND SYNOVITIS. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant; Synovitis

    MAUDE report 19324685 ↗
  • InjuryMay 10, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 78 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS POLYETHYLENE WEAR, PAIN, JOINT LAXITY, PITTING OF THE TIBIAL INSERT, POLYETHYLENE WEAR OF THE PATELLAR COMPONENT, OSTEOLYSIS OF THE TIBIA AND FEMUR, LOOSE FEMORAL COMPONENT, AND LIMITED RANGE OF MOTION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant; Implant Pain; Joint Laxity; Loss of Range of Motion; Osteolysis

    MAUDE report 19285820 ↗
  • InjuryMay 7, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 87 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT MAY REQUIRE A REVISION SURGERY FOR POLYETHYLENE WEAR AND OTHER UNSPECIFIED ISSUES. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 19260020 ↗
  • InjuryApr 12, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT A PATIENT HAD A LEFT TOTAL KNEE ARTHROPLASTY ON (B)(6) 2014 AND THEN EXPERIENCED A REVISION SURGICAL PROCEDURE ON (B)(6) 2023 APPROXIMATELY 8 YEARS AND 8 MONTHS AFTER INITIAL IMPLANT. NO OTHER PATIENT INFORMATION / MEDICAL HISTORY REPORTED. NO IMAGES OF THE DEVICES ARE PROVIDED. THE DEVICE WILL NOT BE RETURNED. THERE IS NO OTHER INFORMATION AVAILABLE.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 19096791 ↗
  • InjuryDec 6, 2023Naturally Worn; Naturally Worn

    AS REPORTED, THE PATIENT HAD AN INITIAL RIGHT TKA ON (B)(6) 2019. THE PATIENT WAS REVISED ON (B)(6) 2023 DUE TO POLY WEAR AND THE LINER WAS EXCHANGED. THERE WAS NO REPORTED BREAKAGE OF A DEVICE OR SURGICAL DELAY/PROLONGATION. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 18280074 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K093360. One submission often covers many models, so check the product and lot details in each record.

  • Z-2310-2012 · Aug 19, 2011 · Terminated
    OPTETRAK LOGIC TIBIAL INSERT POSTERIOR STABILIZED***Match Component Colors 6 YELLOW, 11mm, Use with LOGIC Femoral and Tibial Components***REF 02-012-35-6011. There are three produ

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page