Optetrak Logic 02-012-35-4009Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Optetrak Logic (408)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Exactech Optetrak Logic knee replacement has 106 reports, no deaths and 106 injuries. Most reports describe wear in the plastic knee insert, instability, or loosening of the implant, leading to pain, bone loss around the joint, and revision surgery. There is 1 recall, still open, started in August 2021 because the plastic inserts were packaged without an extra layer that blocks oxygen.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
106
Deaths reported
0
Injuries reported
106
Recalls naming it
1

Summary

The FDA's MAUDE database contains 106 adverse event reports that list this exact device (DI 10885862001887), received between Aug 8, 2020 and Aug 18, 2026 — classified as 106 injuries.

The most frequently reported specific device problem is “Naturally Worn” (86 reports).

1 FDA recall record name this device's identifier, 1 still open.

2 other recalls cover devices cleared under the same FDA submission (K033883); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 2 reportsJan 2024: 1 reportsFeb 2024: 1 reportsMar 2024: 2 reportsApr 2024: 1 reportsMay 2024: 3 reportsJun 2024: 3 reportsJul 2024: 6 reportsAug 2024: 5 reportsSep 2024: 4 reportsOct 2024: 3 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 3 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520266
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 30, 2021Class IIOpen, Classified

    Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

    Z-0021-2022 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Naturally Worn86
  • Unstable25
  • Adverse Event Without Identified Device or Use Problem20
  • Insufficient Device Problem Information9
  • Loosening of Implant Not Related to Bone-Ingrowth7
  • Mechanical Problem3
  • Patient Device Interaction Problem2
  • Accessory Incompatible1

Patient problems

  • Failure of Implant50
  • Pain44
  • Osteolysis42
  • Synovitis39
  • Insufficient Information24
  • Joint Laxity23
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available22
  • Discomfort20

Recent adverse event reports

The 25 most recent of 106 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 18, 2026Naturally Worn

    It was reported via legal notification that a patient underwent initial left total knee arthroplasty. The patient alleges having sustained injuries because of the recalled product have resulted in ongoing treatment. There is no other patient demographic or medical history available. Further patient impact is unknown. No additional information is available.

    Patient problems: Insufficient Information

    MAUDE report 50302279 ↗
  • InjuryMay 12, 2026Naturally Worn

    AS REPORTED, THE PATIENT RECEIVED A BILATERAL TOTAL KNEE ARTHROPLASTY. APPROXIMATELY 8 YEARS AFTER THE INDEX SURGERY, THE PATIENT¿S RIGHT KNEE WAS REVISED DUE TO REPORTED PAIN, FEMORAL LOOSENING, AND POLYETHYLENE WEAR. PER THE OPERATIVE NOTES, THE PATIENT HAS INCREASING PAIN AND RECURRENT EFFUSIONS WITH X-RAY EVIDENCE OF OSTEOLYSIS BEHIND THE FEMORAL COMPONENT. THE SURGEON ELECTED TO REVISE THE PATIENT WITH A CC FEMORAL COMPONENT, A FEMORAL CONE, ONE POSTERIOR AUGMENT BLOCK, TWO DISTAL FEMORAL AUGMENT BLOCKS, A STEM EXTENSION, AND A PS TIBIAL INSERT. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE REVISION SURGERY.

    Patient problems: Aseptic loosening; Joint effusion; Osteolysis; Pain

    MAUDE report 25153018 ↗
  • InjuryNov 17, 2025Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL TKA (TOTAL KNEE ARTHROPLASTY) ON THE LEFT SIDE. APPROXIMATELY 9 YEARS POST INITIAL OPERATION, THEY NEEDED A SYNOVECTOMY AND A REVISION OF THE PATELLA AND TIBIAL INSERT. THE PATELLA EXHIBITED WEAR ALONG THE LATERAL AND MEDIAL TRACKING PATH WHILE THE TIBIAL INSERT HAD WEAR ON POSTEROLATERAL CORNER OF THE LATERAL COMPARTMENT. THE FEMORAL AND TIBIAL COMPONENTS WERE WELL FIXED AND NOT REVISED.

    Patient problems: Osteolysis; Synovitis

    MAUDE report 23570315 ↗
  • InjuryApr 17, 2025Unstable; Naturally Worn; Unstable; Naturally Worn

    IT WAS REPORTED APPROXIMATELY 86 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT SURGERY; A MALE PATIENT UNDERWENT A LEFT KNEE REVISION PROCEDURE TO ADDRESS PAIN AND SYMPTOMS OF INSTABILITY. A REVISION OPERATIVE REPORT WAS PROVIDED. OPERATIVE FINDINGS NOTED THE PATELLA COMPONENT WAS WELL FIXED, FEMORAL COMPONENT LOOSE AND DE-BONDED FROM THE CEMENT AND THE TIBIAL WELL FIXED. IT WAS NOTED THE POLYETHYLENE INSERT SHOWED EVIDENCE OF WEAR OF THE MEDIAL TIBIAL TRAY. INFECTIOUS WORKUP WAS PERFORMED; INFLAMMATORY MARKERS WERE NEGATIVE. THE PATIENT WAS REVISED TO A COMPETITOR'S CONSTRUCT. THE PATIENT WAS NOTED TO BE IN STABLE CONDITION FOLLOWING THE PROCEDURE. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Aseptic loosening; Joint Laxity; Osteolysis; Pain; Swelling/ Edema; Synovitis; Unspecified Tissue Injury

    MAUDE report 21859019 ↗
  • InjuryFeb 18, 2025Peeled/Delaminated; Naturally Worn; Peeled/Delaminated; Naturally Worn

    IT WAS REPORTED THAT APPROXIMATELY 91 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION SURGERY TO ADDRESS PROSTHESIS WEAR AND LOOSENING. A REVISION OPERATIVE REPORT WAS PROVIDED. POST OPERATIVE DIAGNOSIS NOTED ASEPTIC LOOSENING AND SIGNIFICANT POLYETHYLENE WEAR, LEFT KNEE. INTRAOPERATIVELY THE SURGEON OBSERVED SYNOVITIS, OSTEOLYSIS AND NOTED THE POLYETHYLENE HAD EVIDENCE OF WEAR AND DELAMINATION. THE FEMORAL COMPONENT WAS NOTED TO BE SIGNIFICANTLY LOOSE AND DEBONED FROM UNDERLYING CEMENT. POST OPERATIVE NOTES INDICATED AN UNINFECTED OPERATIVE WOUND IN WHICH NO INFLAMMATION WAS ENCOUNTERED. NO SURGICAL OR MEDICAL INTERVENTIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Discomfort; Joint effusion; Osteolysis; Pain; Synovitis

    MAUDE report 21401839 ↗
  • InjuryJan 17, 2025Unstable; Naturally Worn; Unstable; Naturally Worn

    IT WAS REPORTED THAT APPROXIMATELY 144 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, DISCOMFORT, JOINT LAXITY, PAIN, OSTEOLYSIS, SCAR TISSUE, AND SYNOVITIS. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Discomfort; Failure of Implant; Joint Laxity; Osteolysis; Pain; Scar Tissue; Synovitis

    MAUDE report 21181058 ↗
  • InjuryJan 8, 2025Naturally Worn; Naturally Worn

    AS REPORTED, THE PATIENT UNDERWENT AN INITIAL RIGHT TOTAL KNEE ARTHROPLASTY DUE TO GONARTHROSIS. THE PATIENT REPORTED THE ONSET OF PAINFUL SYMPTOMS WITH GONALGIA AND ASSOCIATED FUNCTIONAL LIMITATION, FOR WHICH THEY PRESENTED TO THE EMERGENCY ROOM. ON THIS OCCASION X-RAYS WERE TAKEN THAT SHOWED SIGNS OF PROSTHETIC MOBILIZATION. SAMPLES WERE TAKEN FOR CULTURE TESTS AND A CHEMICAL-PHYSICAL EXAMINATION, AND THE PATIENT WAS PLACED ON THE WAITING LIST FOR A TWO-STAGE SURGICAL REVISION. UPON ADMISSION FOR PROSTHETIC REMOVAL, NEW X-RAYS CONFIRMED THE MOBILIZATION. APPROXIMATELY SIX YEARS POST THE INITIAL TOTAL KNEE ARTHROPLASTY, AND FIVE DAYS AFTER THE PATIENT VISIT TO THE EMERGENCY ROOM, THE PATIENT UNDERWENT SURGERY FOR THE EXPLANTATION OF THE KNEE PROSTHESIS AND THE PLACEMENT OF AN ARTICULATED CEMENTED SPACER. DURING THE SURGERY, SAMPLES WERE TAKEN FOR CULTURE TESTS, WHICH RETURNED NEGATIVE RESULTS; THE PROSTHETIC COMPONENTS WERE SENT FOR SONICATION AND CULTURE EXAMINATION. INTRAOPERATIVELY, THERE WAS A CLEAR MOBILIZATION OF THE FEMORAL COMPONENT AND A SIGNIFICANT VILLOUS SYNOVIAL INFILTRATE. AN ARTICULATED CEMENTED SPACER WAS IMPLANTED, AND THE PATIENT IS CURRENTLY AWAITING SURGERY FOR

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Inflammation; Pain; Synovitis

    MAUDE report 21102007 ↗
  • InjuryJan 7, 2025Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT APPROXIMATELY 69 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS ASEPTIC LOOSENING OF THE FEMORAL COMPONENT, OSTEOLYSIS, PAIN, AND INSTABILITY. INITIAL SURGERY AND REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. POSTOPERATIVE DIAGNOSIS INDICATED ASEPTIC LOOSENING OF THE FEMORAL COMPONENT, SYNOVITIS, AND OSTEOLYSIS SECONDARY TO POLYETHYLENE WEAR OF THE TIBIAL AND PATELLAR COMPONENTS. TISSUE SPECIMENS SENT TO PATHOLOGY REVEALED EVIDENCE OF POLYETHYLENE PARTICULATE DEBRIS, INCLUDING SMALL AND LARGE POLYETHYLENE PARTICLES RESULTING IN TISSUE REACTION TO PARTICULATE IMPLANT MATERIAL AND TISSUE NECROSIS. SURGEON PERFORMED A REVISION TOTAL KNEE REPLACEMENT. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Ambulation or Postural Difficulties; Failure of Implant; Implant Pain; Joint Laxity; Osteolysis; Synovitis

    MAUDE report 21096776 ↗
  • InjuryOct 25, 2024Unstable; Naturally Worn; Unstable; Naturally Worn

    IT WAS REPORTED THAT APPROXIMATELY 67 MONTHS AFTER A TOTAL REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, INSTABILITY, SCARRING, EFFUSION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Ambulation or Postural Difficulties; Joint Laxity; Joint effusion; Pain; Swelling/ Edema

    MAUDE report 20532066 ↗
  • InjuryOct 22, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 100 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT HAS EXPERIENCED PROSTHESIS WEAR. THERE IS NO REPORT OF A PLANNED OR SCHEDULED REVISION. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20505403 ↗
  • InjuryOct 21, 2024Naturally Worn; Naturally Worn

    LEGAL CASE ¿ USA (MDL NO. 3044) CASE-(B)(4) RK REV CASE-(B)(4) RK CURRENT IS ASSOCIATED WITH THIS CASE. IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 100 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR. PLAINTIFF WAS REQUIRED TO UNDERGO A REVISION SURGERY TO REMOVE THE DEFECTIVE DEVICE. PLAINTIFF DOES NOT RECALL EXACT DATE, BUT DISCOVERY AND INVESTIGATION CONTINUE. ADDITIONALLY, PLAINTIFF HAS SUFFERED THE FOLLOWING INJURIES INCLUDING BUT NOT LIMITED TO REHABILITATION, PAIN AND DISCOMFORT, SWELLING, POOR BALANCE, WHICH ALL REQUIRE MEDICAL CARE, REHABILITATION, HOME HEALTH CARE, MENTAL AND EMOTIONAL DISTRESS AND PAIN AND SUFFERING. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE. THE SERIAL NUMBER (B)(6) IS CONFIRMED TO BE A PART OF RECALL NUMBER: Z-0021-2022.

    Patient problems: Balance Problems; Discomfort; Emotional Changes; Pain; Swelling/ Edema

    MAUDE report 20500592 ↗
  • InjurySep 20, 2024Naturally Worn; Naturally Worn

    LEGAL CASE ¿ (B)(4). IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 129 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE. THE SERIAL NUMBER (B)(6) IS CONFIRMED TO BE A PART OF RECALL NUMBER: Z-0021-2022.

    Patient problems: Discomfort; Emotional Changes; Osteolysis; Pain; Swelling/ Edema

    MAUDE report 20272685 ↗
  • InjurySep 11, 2024Unstable; Naturally Worn; Unstable; Naturally Worn

    IT WAS REPORTED THAT APPROXIMATELY 126 MONTHS AFTER A TOTAL REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Implant Pain; Insufficient Information; Osteolysis

    MAUDE report 20192846 ↗
  • InjurySep 11, 2024Naturally Worn; Naturally Worn

    LEGAL CASE ¿ USA PATIENT ID: (B)(6). IMPLANT DATE: (B)(6) 2012, DR. (B)(6) EXPLANT DATE: (B)(6) 2023, DR. (B)(6), OP REPORT ATTACHED REVISION OP REPORT POSTOPERATIVE DIAGNOSIS: FAILED LEFT TOTAL KNEE REPLACEMENT DUE TO POLY WEAR ISSUES WITH OSTEOLYSIS. S. THE FEMORAL COMPONENT WAS SLIGHTLY LOOSE BUT THE SURGEON WAS ABLE TO REMOVE THE FEMORAL IMPLANT WITH VERY MINIMAL BONE LOSS. THERE WAS MASSIVE WEAR OFF OF THE POSTERIOR LIPS BOTH MEDIALLY AND LATERALLY ON THE POLYETHYLENE INSERT. NO DEVICE RETURN ANTICIPATED DUE TO THIS BEING A LEGAL CASE. UNABLE TO LOCATE SURGEON/PATIENT IN EBI. SERIAL NUMBER PROVIDED. HISTORICAL RECORD AND SMARTSOLVE SEARCHED FOR SN/PATIENT NAME WITH NO RESULTS. NO FURTHER INFORMATION. (B)(6) 02-012-35-4009 - LOGIC TIBIA PS MOD INSRT ***SERIAL NUMBER (B)(6) IS CONFIRMED TO HAVE BEEN PACKAGED IN A VACUUM BAG THAT DOES NOT CONTAIN EVOH. 510K: K033883.

    Patient problems: Osteolysis; Pain; Scar Tissue; Synovitis

    MAUDE report 20192703 ↗
  • InjurySep 9, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED THAT APPROXIMATELY 60 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT HAS EXPERIENCED LOOSENING. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Pain; Swelling/ Edema; Synovitis

    MAUDE report 20168905 ↗
  • InjuryAug 22, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT APPROXIMATELY 161 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT HAS EXPERIENCED PROSTHESIS WEAR. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Insufficient Information

    MAUDE report 20033126 ↗
  • InjuryAug 22, 2024Naturally Worn; Naturally Worn

    LEGAL CASE - USA (B)(6) IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 97 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT HAS EXPERIENCED PROSTHESIS WEAR. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Insufficient Information

    MAUDE report 20023971 ↗
  • InjuryAug 15, 2024Naturally Worn; Naturally Worn

    LEGAL CASE - USA IT WAS REPORTED THAT APPROXIMATELY 19 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT HAS EXPERIENCED PROSTHESIS WEAR. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Pain; Scar Tissue; Swelling/ Edema; Synovitis

    MAUDE report 19994098 ↗
  • InjuryAug 6, 2024Naturally Worn; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 153 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, OSTEOLYSIS, AND POPITEAL CYST. INITIAL SURGERY AND REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Cyst(s); Synovitis

    MAUDE report 19922505 ↗
  • InjuryAug 5, 2024Unstable; Naturally Worn; Unstable; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 153 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, OSTEOLYSIS, AND INSTABILITY. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Joint Laxity; Osteolysis

    MAUDE report 19914468 ↗
  • InjuryJul 30, 2024Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth; Naturally Worn; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT APPROXIMATELY 119 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, SIGNIFICANT POLY WEAR, LOOSENING OF THE FEMUR. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Insufficient Information; Synovitis

    MAUDE report 19860135 ↗
  • InjuryJul 17, 2024Unstable; Naturally Worn; Unstable; Naturally Worn

    IT WAS REPORTED THAT APPROXIMATELY 60 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, RIGHT KNEE PAIN, SWELLING OF RIGHT KNEE, SWOLLEN KNEE WITH COMPLICATION OF INTERNAL PROSTHETIC DEVICE AND POSTERIOR INSTABILITY OF RIGHT KNEE. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Implant Pain; Joint Laxity; Swelling/ Edema

    MAUDE report 19763649 ↗
  • InjuryJul 16, 2024Unstable; Naturally Worn; Unstable; Naturally Worn

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 155 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS PROSTHESIS WEAR, PAIN, SWELLING, INSTABILITY AND SYNOVITIS. INITIAL SURGERY AND REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Cyst(s); Discomfort; Failure of Implant; Joint Laxity; Pain; Scar Tissue; Swelling/ Edema; Synovitis

    MAUDE report 19752297 ↗
  • InjuryJul 15, 2024Naturally Worn; Naturally Worn

    APPROXIMATELY 5.5 YEARS AFTER A TOTAL RIGHT KNEE REPLACEMENT, THE PATIENT WAS REVISED DUE TO SUSPECTED EARLY POLY WEAR. THE PATIENT WAS REVISED TO AN ACTIVE-E POLY. THE EVENT IS NOT RELATED TO A BREAKAGE OF THE DEVICE. THE EVENT DID NOT LEAD TO A SURGICAL DELAY/PROLONGATION. PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.

    Patient problems: Insufficient Information

    MAUDE report 19740950 ↗
  • InjuryJul 10, 2024Detachment of Device or Device Component; Naturally Worn; Detachment of Device or Device Component; Naturally Worn

    LEGAL CASE ¿ USA PATIENT ID: (B)(6) + LEFT KNEE. IT WAS REPORTED THAT APPROXIMATELY 119 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE WITH DR. (B)(6) MD, AT (B)(6) IN (B)(6), TO ADDRESS CLINICAL AND RADIOGRAPHIC SIGNS OF PREMATURE POLYETHYLENE WEAR. REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PREOPERATIVE AND POSTOPERATIVE DIAGNOSES INDICATED FAILED LEFT TOTAL KNEE TIBIAL POLYETHYLENE DUE TO PREMATURE POLY WEAR AND POLY RECALL. OPERATIVE FINDINGS INDICATED THERE WAS FOUND TO BE A CLEAN WOUND WITH CLEAR SYNOVIAL JOINT FLUID. THERE WERE WELL-FIXED FEMORAL, TIBIAL, AND PATELLAR COMPONENTS WITH NO OSTEOLYSIS. THERE WAS A VISIBLE AMOUNT AND PALPABLE AMOUNT OF POLYETHYLENE WEAR WITHIN THE WEIGHTBEARING PORTIONS OF THE TIBIAL POLYETHYLENE. THERE WAS ENCAPSULATED POLYETHYLENE DEBRIS WITHIN THE SYNOVIUM WITH A SECONDARY THICKENED SYNOVIUM AND A COMPLETE SYNOVECTOMY WAS PERFORMED. ATTENTION WAS THEN TURNED TO THE TIBIAL POLYETHYLENE INSERT, WHICH WAS DISLODGED FROM THE TIBIAL COMPONENT. A TRIAL REDUCTION WITH A 10MM INSERT WAS TRIALED. THE KNEE WAS TAKEN INTO FULL EXTENSION WITH EXCELLENT RANGE OF MOTION TO 130 DEGREES WITH GOOD PATELLAR

    Patient problems: Insufficient Information; Synovitis

    MAUDE report 19711474 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K033883. One submission often covers many models, so check the product and lot details in each record.

  • Z-0584-2024 · Nov 21, 2023 · Open, Classified
    Exactech Optetrak Logic PS Knee System, Catalog Numbers: 02-012-35-5009
  • Z-0906-2015 · Oct 9, 2014 · Terminated
    Optetrak Asymmetric Hi-Flex Posterior Stabilized Cemented Femorals, size 3, Catalog #244-02-03, Catalog #244-03-03. Usage:Femoral implants

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →