Optetrak Logic 02-012-45-5040Recalls & reported problems

Exactech, Inc. · Class 2 · Implant

All models of Optetrak Logic (408)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10885862004048), received between Aug 23, 2024 and Aug 15, 2025 — classified as 6 injuries.

The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (6 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Loosening of Implant Not Related to Bone-Ingrowth6
  • Unstable2

Patient problems

  • Osteolysis5
  • Joint Laxity3
  • Pain3
  • Synovitis3
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • Inadequate Osseointegration2
  • Swelling/ Edema2
  • Emotional Changes1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 15, 2025Unstable; Loosening of Implant Not Related to Bone-Ingrowth; Unstable; Loosening of Implant Not Related to Bone-Ingrowth

    THE PATIENT WHO IS 10 YEARS POST RIGHT TOTAL KNEE REPLACEMENT. OVER THE PAST SEVERAL MONTHS HE IS DEVELOPING INCREASING PAIN AND SWELLING. HIS X-RAYS SHOWED OSTEOLYSIS UNDER THE TIBIAL PLATEAU WITHOUT ANY OTHER OBVIOUS SIGNS OF LOOSENING IN THE FEMUR OR PATELLA. THERE DID APPEAR TO BE MODERATE WEAR IN THE POLYETHYLENE LINER. DUE TO THE PATIENT¿S INCREASING PAIN AND DEBILITATION, ALONG WITH THE X-RAY SHOWING SIGNS OF OSTEOLYSIS, THEY WERE DETERMINED TO BE A CANDIDATE FOR RIGHT TOTAL KNEE REVISION. INTRAOPERATIVE FINDINGS REVEALED SIGNIFICANT WEAR OF THE MEDIAL ASPECT OF THE POLY LINER WITH SOME PITTING POSTERIOR MEDIALLY. THE TIBIAL COMPONENT WAS NOT OBVIOUSLY LOOSE BUT WAS EASILY REMOVED. THERE WERE ALSO SIGNIFICANT OSTEOLYTIC LESIONS PRIMARILY LATERALLY WHICH WERE CONTAINED AND DID NOT SIGNIFICANTLY AFFECT THE SUPPORTIVE NATURE OF THE BONE. THE FEMORAL COMPONENT WAS WELL FIXED BUT WAS REVISED TO ALLOW FOR CONSTRAINED CONDYLAR IMPLANT. THE PATELLAR COMPONENT WAS WELL FIXED AND LEFT IN PLACE. THE PATIENT WAS TAKEN TO THE RECOVERY ROOM IN SATISFACTORY CONDITION. THERE WERE NO COMPLICATIONS.

    Patient problems: Joint Laxity; Osteolysis; Pain; Swelling/ Edema

    MAUDE report 22814144 ↗
  • InjuryJun 23, 2025Unstable; Loosening of Implant Not Related to Bone-Ingrowth; Unstable; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED THAT A MALE PATIENT, WHO HAD A RIGHT KNEE IMPLANTED, UNDERWENT A REVISION PROCEDURE APPROXIMATELY 5 YEARS POST THE INITIAL PROCEDURE. THE PATIENT PRESENTED WITH INCREASING KNEE PAIN, A SENSE OF INSTABILITY, AND SWELLING. PLAIN RADIOGRAPHS REVEALED POLYETHYLENE WEAR, LOOSENING OF BOTH THE FEMORAL AND TIBIAL COMPONENTS, AND OSTEOLYSIS AFFECTING BOTH BONES. DURING THE REVISION PROCEDURE, A FEMORAL AND TIBIAL BONE DEFECT WAS IDENTIFIED CONSISTENT WITH AORI TYPE 2B. THERE WAS ALSO A SIGNIFICANT PATELLAR DEFECT AND A NOTABLE SOFT-TISSUE REACTION CHARACTERIZED BY SYNOVITIS. THE PATIENT WAS REVISED TO A COMPETITOR¿S DEVICES. THERE WERE NO DEVICE BREAKAGES OR SURGICAL DELAYS DURING THE PROCEDURE. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT. X-RAYS WERE PROVIDED. THE EXPLANTED DEVICES ARE NOT AVAILABLE FOR RETURN. NO FURTHER INFORMATION. THIS IS 1 OF 4 REPORTS.

    Patient problems: Inadequate Osseointegration; Joint Laxity; Osteolysis; Pain; Swelling/ Edema; Synovitis

    MAUDE report 22312117 ↗
  • InjuryNov 25, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT APPROXIMATELY 72 MONTHS AFTER A RIGHT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS LOOSENING, SYNOVITIS, AND OSTEOLYSIS. INITIAL SURGERY AND REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. THERE IS NO OTHER PATIENT DEMOGRAPHIC OR MEDICAL HISTORY AVAILABLE. THERE IS NO INFORMATION ON THE SURGICAL PROCEDURE OR PATIENT OUTCOME. THERE IS NO DEVICE RETURN. THERE ARE NO PHOTOS OR OTHER IMAGES OF THE DEVICE PROVIDED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Osteolysis; Synovitis

    MAUDE report 20775996 ↗
  • InjurySep 6, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED VIA LEGAL DOCUMENTATION THAT A PATIENT HAD A LEFT TOTAL KNEE ARTHROPLASTY ON (B)(6) 2016, AND THEN EXPERIENCED A REVISION SURGICAL PROCEDURE ON (B)(6) 2020, APPROXIMATELY 4 YEARS, 6 MONTHS AFTER INITIAL IMPLANT. REVISION OPERATIVE REPORT OF (B)(6) 2020- POSTOPERATIVE DIAGNOSIS: LOOSE LEFT TOTAL KNEE REPLACEMENT. THE POLYETHYLENE WAS REMOVED, THE FEMORAL COMPONENT WAS NOTED TO BE GROSSLY LOOSE. IT¿S APPEARANCE LOOKED GOOD, HOWEVER, IT WAS NOTED TO BE LOOSE. THE TIBIAL COMPONENT WAS ALSO FOUND TO BE LOOSE AND REMOVED. THE PATIENT TOLERATED THE PROCEDURE WELL. THE DEVICES WERE NOT RETURNED, THERE ARE NO IMAGES PROVIDED. THERE IS NO OTHER INFORMATION AVAILABLE.

    Patient problems: Inadequate Osseointegration

    MAUDE report 20156449 ↗
  • InjuryAug 23, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    LEGAL CASE ¿ USA (B)(4). PATIENT ID: (B)(6). 8/19/24: ADDITIONAL INFORMATION RECEIVED IN IN-BOX ON 8/8/24. ATTACHED EMAIL AND REVISION OP REPORT. UPDATED AWARE DATE ON TIBIAL INSERT AND ADDED DEVICE INFORMATION FOR FEMUR, TIBIAL TRAY, AND PATELLA BASED ON NEW INFORMATION RECEIVED. PREOPERATIVE DIAGNOSIS INDICATED FAILED LEFT TOTAL KNEE ARTHROPLASTY DUE TO POLYETHYLENE WEAR. POSTOPERATIVE DIAGNOSIS INDICATED LEFT TOTAL KNEE ARTHROPLASTY DUE TO POLYETHYLENE WEAR AND EARLY LOOSENING. INDICATION FOR PROCEDURE: (B)(6) IS A 79-YEAR-OLD WHO UNDERWENT A PRIOR TOTAL KNEE ARTHROPLASTY WITH A RECALLED PROSTHETIC DEVICE. DUE TO CONCERNS FOR EARLY LOOSENING AND EARLY POLYETHYLENE WEAR, THE PATIENT WAS OFFERED REVISION TOTAL KNEE ARTHROPLASTY IN ITS ENTIRETY. INTRAOPERATIVE FINDINGS: SIGNIFICANT OSTEOLYSIS, EARLY LOOSENING OF COMPONENTS AND SIGNIFICANT POLYETHYLENE WEAR OF THE PATELLA COMPONENT AND TIBIAL LINER. BRIEF DESCRIPTION OF OPERATIVE PROCEDURE: A PARAPATELLAR ARTHROTOMY WAS PERFORMED AND IMMEDIATELY UPON MAKING THIS, A LARGE EFFUSION WAS ENCOUNTERED. THERE WERE POLYETHYLENE FRAGMENTS FLOATING WITHIN THE KNEE AND POLYETHYLENE WEAR NOTED OF THE PATELLAR COMPONENT AND TIBIAL LINER. SOME EA

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Emotional Changes; Osteolysis; Pain

    MAUDE report 20054105 ↗
  • InjuryAug 23, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth

    LEGAL CASE ¿ UNITED STATES (MDL NO. 3044) RELATED CASE FOR REVISION OF RK DOCUMENTED ON (B)(4). IT WAS REPORTED THAT APPROXIMATELY 119 MONTHS AFTER A LEFT TOTAL KNEE REPLACEMENT PROCEDURE, THE PATIENT UNDERWENT A REVISION PROCEDURE TO ADDRESS LOOSENING, INSTABILITY, RECURRENT EFFUSIONS, EXTENSIVE SYNOVITIS, OSTEOLYSIS, ASEPTIC LOOSENING OF FEMUR AND TIBIA, AND GROSSLY LOOSE PATELLA; DELAMINATED POLYETHYLENE DEBRIS. INITIAL SURGERY AND REVISION SURGERY OPERATIVE NOTES WERE PROVIDED. PATIENT LEFT THE OPERATING ROOM IN STABLE CONDITION. POST OPERATIVE DIAGNOSIS NOTED WEAR OF THE IMPLANT. NO FURTHER ISSUES OR COMPLICATIONS WERE REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Joint Laxity; Joint effusion; Osteolysis; Synovitis

    MAUDE report 20052762 ↗

Other Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices

All Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer devices →